In 1976, Dr. Leon Frazin published the results of studies using an esophageal Mmode transducer Matsumoto and associates used M-mode TEE to study left ventricular (LV) function In the early 1980s, investigators led by Hanrath introduced a two-dimensional (2D), phased-array transducer mounted on the tip of a flexible gastroscope
2. Monitor the heart over time, such as during cardiac surgeries 3. Extremely safe & well tolerated so that it can be performed in critically ill patients & very small infants
Mitral valve
MR (very precisely) function of prosthetic valves refined suitability for valvotomy in severe MS
Cont.
Aorta Detection of dissection Detection of atheroma Chambers Patent foramen ovale Online monitoring Monitoring interventional procedures o atrial septostomy o baloon valvotomy o pulmonary vein interventions
Properties of Ultrasound
dual process called compression and
rarefaction
sequence of compression and rarefaction is described by sine waves characterized in terms of Wavelength distance between two peaks of the sine wave Frequency number of cycles that occur in 1 second Amplitude measure of tissue compression Propagation velocity speed of an ultrasound wave
traveling through tissue
Echocardiography uses frequencies of 2.5 to 7.5 million cycles/sec (MHz) Important relationships between transducer frequency, depth, and resolution include (1) the slower the frequency, the greater the depth of penetration at the expense of resolution, and (2) the faster the frequency, the greater the resolution at the expense of depth of penetration
Cont.
Image resolution is characterized in terms of
Axial resolution Elevational resolution Lateral resolution Temporal resolution
Cont.
Axial resolution is resolution along the length of an ultrasound beam A function of transducer frequency and pulse width Elevational resolution refers to the thickness of the ultrasound image: typically 8 to 10 mm
Cont.
Lateral resolution is a function of ultrasound beam formation The initial segment (near field) of an ultrasound beam is columnar Beyond the near field, the beam diverges (far field) Divergence is also a function of transducer frequency. As the frequency rises, the near field gets longer and the divergence angle decreases.
Cont.
Cont.
Temporal resolution is the ability to accurately locate moving structures at a particular instant in time
Attenuation :- a function of tissue absorption , divergence of ultrasound energy as it moves away from the transducer, reflection, and scattering
Absorption is dependent on tissue type. For example, air absorbs more ultrasound energy than bone does, but bone absorbs more than blood or water does. Acoustic impedance :- refers to the resistance that an ultrasound wave meets when traveling though tissue
The greater the tissue density, the faster the ultrasound beam moves through the tissue. Extent of ultrasound beam reflection is a function of the difference in acoustic impedance between two adjacent tissues
larger the difference - the more ultrasound energy will be reflected
Mismatches in acoustic impedance and attenuation are important to consider in imaging the heart For example, the upper aortic arch is difficult to visualize from the esophagus because of interposition of the air-filled trachea. Most of the ultrasound energy that does make it to the trachea-tissue interface is reflected as a result of the large difference in acoustic impedance between air and tissue.
Cont.
Doppler shifts are presented in three ways: 1. Pulsed Wave Doppler (PWD), 2. Continuous Wave Doppler (CWD), and
One limitation of PWD is that it may be too slow to capture the velocity of fast-moving blood cells. This phenomenon is known as aliasing. The limit at which the sampling rate fails to accurately capture the true velocity is called the Nyquist limit Aliasing of PWD occurs at blood flow velocities greater than 0.8 to 1.0 m/sec. Normal flow within the heart may reach 1.4 m/sec and pathologic flow up to 6 m/sec.
CFD is based on PWD and uses multiple sample volumes along a scan line A color code is used to depict flow toward (red) and away (blue) from the transducer
Contraindications
1. 2. 3. 4. Absolute Previous esophagectomy, Severe esophageal obstruction, Esophageal perforation, and Ongoing esophageal hemorrhage
Cont.
1. 2. 3. 4. Relative Esophageal diverticulum, Varices, Fistula, and Previous esophageal surgery, as well as a history of gastric surgery, mediastinal irradiation, unexplained swallowing difficulties
Patient preparation
Informed consent Pt. should fast for at least 4 6 hrs Thorough history should be taken any dysphagia i.v. access Pre oxygenation Suction should be available
Transesophageal Echocardiography
Risks of TEE
Topical anesthesia Allergic reactions Toxic methemoglobinemia Conscious sedation Hypoxia Hypotention Agitation Idiosyncratic reactions
Cont.
Probe insertion : immediate Oral trauma Dental trauma Esophageal trauma Vagal reaction
TEE views
Upper oesophageal (UE) level 20-25cm Mid Esophageal (ME) level 30-40cm
Different Views
Assessment of Hemodynamics
Evaluation of Ventricular Filling o TEE reveals changes in LV preload more reliably than filling pressure o TEE demonstrates a significant decrease in cross-sectional area at end diastole (EDA) o An EDA of less than 12 cm2 indicates hypovolemia
Cont.
Estimation of Cardiac Output o TEE can quantify cardiac output more precisely o By measuring both the velocity and the cross-sectional area of blood flow at appropriate locations in the heart or great vessels
Cont.
Assessment of Ventricular Systolic Function Fractional area change (FAC) during systole is a commonly used measure of global LV function.
Assessment of Ventricular Diastolic Function Heart failure is due to diastolic dysfunction: abnormal diastolic relaxation and filling
Cont.
Cont.
TEE During Life-Threatening Hypotension Hypotension has only two possible causes: inadequate cardiac output or inappropriately low systemic vascular resistance Qualitative TEE estimates of ventricular filling and function serve as practical guides for the administration of fluids, inotropes, and vasopressors
Cont.
Myocarditis, myocardial infarction, and myocardial stunning also cause SWMAs. However,a sudden, severe decrease in segmental contraction is almost certainly due to myocardial ischemia.