SERTIFIKAT KALIBRASI
CALIBRATION CERTIFICATE

Nomor
: 3-02-22-00204
Number
ALAT
Equipment
1. Nama : Pressure Gauge 2. Kapasitas : 69 Mpa
Name Capacity 1 divisi 1 Mpa
3. Tipe/Model : ANALOG 4. Nomor Seri : 286.60.1 / 2
Type/Model Serial
5. Merek/Buatan : MTS / USA - 6. Ukuran : -
Manufacture Internal
7. Pengontrol Suhu : - 8. Lain-lain :
Temperature Control Other
PEMILIK
Owner
1. Nama : PT MINTEC ABADI
Name
2. Alamat : Tambang PT Kideco Jaya Agung, Desa Batukajang Kec. Batu Sopang
Kab. Paser Regency, East Kalimantan 76252
Address

STANDAR
Standard
1. Nama : DWT No. Seri : 25/213
Name DWT No. Seri : 2533
2. Ketelusuran : Hasil kalibrasi yang dilaporkan tertelusur ke satuan pengukuran SI melalui
Traceability LK-022-IDN

TANGGAL DITERIMA : 27 September 2022 TANGGAL KALIBRASI : 3 Oktober 2022


Date of acceptance Date of Calibration

KONDISI LINGKUNGAN : Suhu udara : Suhu 19,8 ± 0,4 °C


Environment condition of Calibration Kelembaban 55 ± 2 % RH

LOKASI KALIBRASI : Lab. Kalibrasi Tekanan B4T, Bandung.


Location of Calibration

METODA KALIBRASI : PC-302-01


Calibration method

ACUAN : BS EN 837-1: 1998


Refference - Bourdon tube pressure gauges, Dimension, Metrology, Requirement and Testing

HASIL KALIBRASI DAN KETIDAKPASTIAN KALIBRASI ( Terlampir)


Result of Calibration and uncertainity of Calibration ( Attached )

DITERBITKAN TANGGAL : 5 Oktober 2022

Halaman 1 dari 2

Keterangan : 1.Laboratorium ini di akreditasi oleh Komite Akreditasi Nasional (KAN) No. LK-022-IDN.
2.Dilarang memproduksi sertifikat ini tanpa izin tertulis dari B4T kecuali secara keseluruhan
3.Hasil kalibrasi ini tidak untuk diumumkan dan hanya berlaku untuk alat yang bersangkutan.
- Dokumen ini ditandatangani secara elektronik menggunakan sertifikat elektronik yang di terbitkan oleh Balai Sertifikasi Elektronik, Badan Saiber dan Sandi Negara
- Berdasarkan Undang- Undang Informasi dan transaksi elektronik pasal 11 tahun 2008, tandatangan elektronik memiliki kekuatan hukum dan akibat hukum yang sah
- Dokumen dapat diverifikasi menggunakan dokumen aplikasi VeryDS yang tersedia di Playstore atau dengan mengakses http:/bsre.bssn.go.id/verifikasi
F. 50/01/Rev.0/04122014
BADAN STANDARDISASI DAN KEBIJAKAN JASA INDUSTRI
BALAI BESAR STANDARDISASI DAN PELAYANAN JASA INDUSTRI
INDUSTRI BAHAN DAN BARANG TEKNIK
Jl. Sangkuriang No. 14 Bandung 40135 JAWA BARAT - INDONESIA
Telp. 022 - 2504088, 2510682, 2504828, 2507626 Fax. 022 - 2502027
E-mail : b4t@b4t.go.id Website : http://www.b4t.go.id

LAMPIRAN SERTIFIKAT KALIBRASI


ATTACHMENT OF CALIBRATION CERTIFICATE

Nomor
: 3-02-22-00204
Number
1. HASIL KALIBRASI
Calibration of Result

Overload Test : 10 %
Residual of zerro : 0 %
No. Seri 286.60.1
Penunjukan
Pressure Penunjukan Standar Kesalahan Ketidakpastian
Gauge Indicated by standard Error Uncertaynty
Indicated by
Pressure Gauge Naik Turun Naik Turun Naik Turun
Increase Dicrease Increase Dicrease Increase Dicrease

Mpa Mpa Mpa % % % %

10 10,018 10,028 - 0,03 - 0,05 ± 0,34 ± 0,34


20 19,993 20,012 + 0,01 - 0,03 ± 0,34 ± 0,34
30 29,871 29,891 + 0,19 + 0,14 ± 0,34 ± 0,34
40 39,700 39,729 + 0,43 + 0,39 ± 0,34 ± 0,34
50 49,622 49,642 + 0,55 + 0,51 ± 0,34 ± 0,34
60 59,565 59,585 + 0,63 + 0,59 ± 0,34 ± 0,34
69 68,746 68,746 + 0,37 + 0,37 ± 0,34 ± 0,34

No. Seri 286.60.2


Penunjukan
Pressure Penunjukan Standar Kesalahan Ketidakpastian
Gauge Indicated by standard Error Uncertaynty
Indicated by
Pressure Gauge Naik Turun Naik Turun Naik Turun
Increase Dicrease Increase Dicrease Increase Dicrease

Mpa Mpa Mpa % % % %

10 10,194 10,204 - 0,30 - 0,30 ± 0,34 ± 0,34


20 20,091 20,110 - 0,13 - 0,16 ± 0,34 ± 0,34
30 29,930 29,911 + 0,10 + 0,13 ± 0,34 ± 0,34
40 39,798 39,828 + 0,28 + 0,25 ± 0,34 ± 0,34
50 49,819 49,838 + 0,25 + 0,23 ± 0,34 ± 0,34
60 59,762 59,782 + 0,33 + 0,32 ± 0,34 ± 0,34
69 68,765 68,943 + 0,34 + 0,08 ± 0,34 ± 0,34

*) Ketidakpastian yang dilaporkan adalah ketidakpastian bentangan pada tingkat kepercayaan 95% dengan faktor cakupan k=2
The reported uncertainty is the uncertainty of a stretch at 95% confidence level with coverage factor k = 2

-Menurut BS EN 837-1: 1998, kelas akurasi Pressure Gauge ditentukan oleh kesalahan maksimum yang diijinkan,
setelah ditambahkan nilai ketidakpastian maksimum.
-According to BS EN 837-1: 1998, Pressure Gauge accuracy class is determined by maximum allowable error,
after the added value of the maximum uncertainty

Halaman 2 dari 2

Keterangan : 1.Laboratorium ini di akreditasi oleh Komite Akreditasi Nasional (KAN) No. LK-022-IDN.
2.Dilarang memproduksi sertifikat ini tanpa izin tertulis dari B4T kecuali secara keseluruhan
3.Hasil kalibrasi ini tidak untuk diumumkan dan hanya berlaku untuk alat yang bersangkutan.