Anda di halaman 1dari 5

PT.

CALIBRAMED
hBM&L@dH@,tuffiWidtgvb
ERAMED

SERTIFIKAT KALIBRASI
C A L I B M TION CERTI FICATE

Norctor&t: Crr 22 / U / 29t-t


Nort9ttifr&t.: S 1',{'/f,/ U /22
Iffiottb

:tt r i Bcd Sid. MoDlaor

tlt Ltut'tif- : Inflium

rvc .' Omni lI

No*rS.i i Omnn-A-140+35t73

Itftrtiut e',i&f.

Na r RSUD. Dr. H. BOB BAZA"R- SKM

tb.t . JL L.ttu Robrai No. l4B, Krlirndr, L.mpu||g S.lrtsr.

S,'tiflt init ,ei[ui 5 fratmt


.'iir .ntfut .a/i*
"f
Oe.inlte t4wdf 2grtttltt 2022

Mrbti,Ua 2gttrrrtt 2t23

aj''ll1'r ,tmitte&*
n
P[. CflIIERAIIED
StonnfKdlq SE Sugat, gr

Tid.t ditql^e mtg'nip / lMrFbrly.t de o|tu e!pubri!.;r6 ebqse ti stift!1 ini t&r! sizi! pT.CALIBRAMED
Ilti'../6L& t llndt tu ns.dtub'fitll u!.s'l3'it7,frr t,in*odudd\tuthshaobbi,Mthditineh,n CAUTtUMED
s6[6trr ili neys dibci]o hruk .l.r tjrg dik lib.si
1ri. d6Jr.t b h. 4ppX.d qbld ofi 6A- e

Address Phone Far E-Mail W€bsft€


ll. Dr. sahado No. 41 E lakana S€latan Indonesia (021)28542554 (021128543637 calibGmed@smail-.om wwwcatibramed co.id
No.Ordc. : cN, 22 / 08 t29t-5

Laob.r ke 2 drri S lembsr


!s:. - -- - -. - - -- - -. - -..J.

Data Alat P€langgan :

l. Ntmt Al^t I NdE ofEquipnhl : Bed Side Monitor


2. M.tkPrbrik I L!tuJ@rtw I Infnillm
3. TiW I r,p, : Omni
4. No.S.ri / .'.'dl x!'r.' : Omni2-A.1404-355?3

Pelakranaan Krlibrasi :
tmplen htatio" ol(:dlibradm
l. Temptt I pt.e
RSUD. Dr. H. BOB EAZAR SKM
2. Rurtrgm /xah OK
3. 'Irrggsl Pcnarimrrn I ne..ipt ate 22 Agustus 2022
4. T.nggrl Krlibr.si / cattbrdio, Dak 22 ASustus 2022
5. T.knisi I rechnicial Yusril, A.MdTEM

Stsodrr Peralatao lftlibrssi ;


t;tot lar.t oJcdibrdrid tquipna|

No Diskripsi Tip" No. Seri No. SertiliLst Tertelusur


ECG Simulator F'luke Prosim 2 ItE4009 YK.ot .o3f,{J-vm 2EKt2 LK-l rO-IDN
022 4082
NIBP Simulator rnd BP-PUMP 2L 99440t9 s-2t2lo1D2 LK 128 rDN
Tester
3. SPO2 Simulator tndex 2nFE 9676008 cAU065/9676008- PT Quantun
0s/|3/22
Electical Safety ESA 620 2t490.25 YK.0t.03r(Lutr.?/PK./2 LK -1IG TDN
Analyzer o22 4081
5. Thermohygrometer CAIj-AB/I03 rNV/THGM-022 (M) S.21 0t5174 LK-O32.IDN

Koodici R[Nng8n ;
EwiM.tt l t:ot'aitint

No Paromeaer / ptru,ao Terukur I M*.d


Akhir /
("c) 23,1
%RU )

Kondisi Fisikdan Fun$i Alst *) :


Pklatcal l:onditid a1d Furctid ol Eqtip,ne,n

No Dirkripsi / o.,c,aao Kondisi Fbik Furgsi Alat

l. Badrn drn Permulran I Erctns.e @'t it, st!t@, Baik llArk


2. Kontrof drn irdikttor I co'ttol ort tvlt.atot Baik Baik
3. Kt}ej ii.r.n s*t.,,otil wb. bdae*o ., Brik Baik
No.Ot{cr r cl|I n, I 0t I 29r-s

Lcrb|rL3d.d5hDb|f
f.r..................,t

HASIL KALIBRASI I ca.wuornwtr

No Prrrne{er Ter||hr Tolennd $tu|r


l*8 Ltu
T€Dgr! Cdl Drya z2l,l Zm +t0

Rcoietmsi Pembulnie Prot ktif 0,088 <0,2


Proucda Foth Rettate
botari Rogi*i >', MO

Aros Boc6 P@hraie 36,5 s 500


MtL.dq.Cn dN@dlblaitr
rA
Arl|8 Bocor Bagie ye8 Dibt$fe Ir3 s 500 FA
Ap?A.d P@r Eeh L.tq. Ctdrl

Dbl.llp.l S.c&E St rd|r Prrhu Alrr KolcLd Kcdd.bsd.tr


(%%Cr.lF.2l

Sy!tol. 100 102,0 0,58


Irirdob 65 66,5 -1,5 0,73
Syrbh r20 122,0 -2,0 q58
80 8.1,0 -1,0 0,58
Stniolc 150 1s2,0 40 0,J8
It lto|. 100 r05p -5,0 05t
202,0 .2,0
154.0 -{,0
No,Ordcr: cAL 22 I @ t 29r-5

L.dbor lI. ,t d|rt 5 hdb.r


Lf ..................,1

S.tlb8 St|Dd|r P.Db.c.D Al Koltld I&dd.lportian


( 9r% cr4 b2)
% tY.
80 8t,0 .1,0 0,58
85 t6,0 -1,0 0,58
q) 91,0 -1,0 058
95 %,0 -1,0 q58
lfl) 1,0 0,58

Scaairg Sbrd.r Pcmbc..rn ALa Korcki KodddDo!ti.n


tutMSefiv
( % c,4'F4
bpo I bFr
60 60,0 q0 0,58
8{) 80,0 0,0 0,5E
lm !20,0 q0 q58
t80 180,0 0,0 q58
2& 240,O 0p 0,5t

S€fiirg Strdrr Pcnbraun Alrt Xot.td


( 9!t% CI,lF2)

0,0 0,58
60 60,0 o,o q58
r20 120,0 0,0 q58
l8{) 180,0 0,0 q58
240 240,0 0,0 0,jS

SGldD8 Sard|' PrnDcIr Aha l(or&l Kcdlrtlpoder


(93U. c|.H)
t
l5 15,0 0,0 q58
20 20,0 0,0 0,58
30,0 0,0 0,58
40 40,0 0,0 0J8
60 60,0 0,0 8
No.Order : cAL 22 I 0a t29t-5

Lember k 5 drri 5 l.mb.r

Keteraogan ;

l. Hasil Kctidakpastian Pengukuratr mempuyai tiDgkat k€percayaan 95olo d€ogan fatto, cakupan (k) 2
=
Re'll of utue ainty M.6u@.nt t@1ae a @,fuj rc. t vet €L) olgt%vih,h. cMt&e t@tot (h) of 2.
IrE
2. Korcklf - Pembacaan Standar - Pernbacaan Al at
@tt :natd@d Red'rA - Fqtpd.nt PIdDE
3 Flasil lklibrasi hanya berlaku untuk alat dengan nomor seri sebagaimam tercantun diatas
Cdibtdto, tuttlt ed,:tlu t@b vih seriat Mb.r d ttrt tt ab@
orc @U
4. Hasil kalibrasi tertelusur ke Sist€m Satuatr htemagiond (SI) melalui LK-{L2-IDN
71c R.flL ofcdlibrotioa it ttddt b Lt ,,diotlot SI,,d of t/tu (SI) throrgh L64J,IDN.

5- Hasil Peo8ulure pem€dksaao fisik da nfigsi serta k€sitnpulu hasil katibrasi dicortumkar sesuai
Perocnkd RI No. 54 Trhu! 2015 drD kcpuau!.tr Dirjcr yrDl.. nomor ts,K- O2,O2fulO4l2nO2O
T€dlang Mrio& Krrjs Pedgujiar ds! KalibEsi Alst Kreehdsn.
lk Aank (4 Pt'6cd dtdM ctE sftsd b *b,rt6 of *.dtbrd,t ,*te *e eordra to p.n ra*a RI
'
N^ tl ot2ott.rt D@ qbc Db.e, o@.tqE..Lh Sdij. r.r',d EA,O2,oz,r/utz2UO @tuDr,avoilj,eM.ho.te
OfT.ttDA a1d Calibratlon oJM.dicdl Dan6
6 Kesimpulan I Berdasarkan hasil pemeriksaan keseluruhan kine{a masih masuk dalam batas tolemnsi.
bat&t M the RsA olhe i8p..ti8 aI p.dontu Ftui.teB @,ti *irhi, tk tot m., Idil

ALAT II\IK
Akhir Sertif*at / eu4c.an a.
PI.