50
100 / 80
150 / 100
200 / 120
250 / 150
KETERANGAN
1. ………………………………………………………………………………………………………………….
2. ………………………………………………………………………………………………………………….
Memeriksa Paraf
Penyelia
Manajer Teknis F-LK-MKT-01
50
100 / 80
150 / 100
200 / 120
250 / 150
KETERANGAN
1. ………………………………………………………………………………………………………………….
2. ………………………………………………………………………………………………………………….
Memeriksa Paraf
Penyelia
Manajer Teknis F-LK-MKT-01
PHYGMOMANOMETER MERCURY / ANEROID
………../……….- 16 / E -….………….Dt/ DL
…..……………………..
1………………………..
2………………………..
……………………………………………………….
……………………………………………………….
……………………………………………………….
……………………………………………………….
ANALISA DATA KALIBRASI SPHYGMOMANOMETER MERCURY
Nomor Label : 72 / 1 / II - 17 / E-037 Dt
Merek Riester
: Alat Standart ( Nr. seri ) :
Model/Tipe Nova
: Presameter 1. Vacuum and Pressure meter , Merek : Crystal, Model : IS 33 Sn.2535-0175
No. Seri 030178304
: 2. Therma-Hygrometer, Merek : Sekonic, Model :ST-50 (Tertelusur ke BPFK
Daya Baca Operator 1: mmHg Tempat Kalibrasi Laboratorium
: Tekanan BPFK S
Rentang Ukur 0-300
: mmHg Nama Ruang - :
Suhu : 21.2 ± 0.14 °C Tanggal Diterima :
11 Januari 2017
Kelembaban : 73.2 ± 0.3712310601 Tanggal Kalibrasi 9 Februari
: 2017
Metode Kerja : MKT-01 Nama Petugas Kalibrasi Egan: Graha Utama, Amd.Tem
Keterangan P.Standar P.Standar Koreksi Koreksi Terkoreksi Terkoreksi Rata-rata Standard Uji Kebocoran
Awal Akhir Awal Akhir Awal Akhir Deviasi Laju Kebocoran
Suhu ( 0C ) 21.3 21.5 -0.15 -0.16 21.15 21.34 21.25 0.14
Kelembaban (% RH) 71.3 71.8 1.61 1.63 72.91 73.43 73.17 0.37
Pembacaan Pembacaan Standar
Pada Naik I Turun I Naik II Turun II Naik III
Alat Terbaca Terkoreksi Terbaca Terkoreksi Terbaca Terkoreksi Terbaca Terkoreksi Terbaca Terkoreksi
(mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg) (mmHg)
0 0.00 0.00 0 0.00 0 0.00 0 0.00 0 0.00
50 52.47 52.55 52.42 52.61 52.33 52.41 52.31 52.50 52.39 52.47
100 102.42 102.38 102.29 102.35 102.21 102.18 102.19 102.25 102.36 102.32
150 152.00 151.86 151.98 151.91 151.87 151.73 151.77 151.71 151.95 151.81
200 201.58 201.32 201.50 201.31 201.48 201.22 201.35 201.16 201.61 201.35
250 251.01 250.64 250.89 250.57 251.19 250.82 251.12 250.80 251.11 250.74
HASIL INTERPOLASI KOREKSI
Naik I Turun I Naik II Turun II Naik III
0 0.00 0.00 0.00 0.00 0.00
50 0.08 0.19 0.08 0.19 0.08
100 -0.04 0.06 -0.03 0.06 -0.04
150 -0.14 -0.07 -0.14 -0.06 -0.14
200 -0.26 -0.19 -0.26 -0.19 -0.26
250 -0.37 -0.32 -0.37 -0.32 -0.37
SI SPHYGMOMANOMETER MERCURY
: 72 / 1 / II - 17 / E-037 Dt
m and Pressure meter , Merek : Crystal, Model : IS 33 Sn.2535-017553 (Tertelusur ke PT. KALIMAN)
-Hygrometer, Merek : Sekonic, Model :ST-50 (Tertelusur ke BPFK Surabaya )
Laboratorium Tekanan BPFK Surabaya
1 mBar =
350 mBar =
Ketidakpastian naik/turun terbesar standar : ± 0.24 mmHg
0 0.00 0 0.00
0.00 0.00 0.00 0.00
50 0.09 50 0.09
50 0.09 50 0.09
52.47 0.08 52.33 0.08
100 -0.03 100 -0.03
NAIK
0 0.00 0
0.00 0.00 0.00
50 0.09 50
50 0.09 50
52.39 0.08 52.42
100 -0.03 100
0.00 0 0.00
0.00 0.00 0.00 0.00
0.20 50 0.20
0.20 50 0.20
0.19 52.31 0.19 52.22
0.07 100 0.07
0 0.00
0.00
50 0.20
50 0.20
0.19
100 0.07
100 0.07
0.06
150 -0.06
150 -0.06
-0.06
200 -0.19
200 -0.19
-0.19
250 -0.32
250 -0.32
-0.32
300 -0.45
INTERPOLASI SUHU
19.90 -0.125
-0.15
25.00 -0.223
19.90 -0.125
-0.16
25.00 -0.223
INTERPOLASI KELEMBABAN
63.40 1.21
1.61
72.60 1.67
63.40 1.21
1.63
72.60 1.67
U N C E R T A
Nomor L
No. Pembacaan Komponen Ketidakpastian Satuan Distribusi U Divisor Reliability vi Ui c (Ui.c)
Alat
( mmHg)
1. Daya Baca Operator mmHg Rectangular 0.500 √3 10 50 0.289 1 0.289
2. Zero Error mmHg Rectangular 0.000 √3 10 50 0.000 1 0.000
3. Histerisis mmHg Rectangular 0.000 √3 10 50 0.000 1 0.000
4. Naik mmHg Rectangular 0.000 √3 3 2 0.000 1 0.000
Repeatability
5. 0 Turun mmHg Rectangular 0.000 √3 3 2 0.000 1 0.000
6. Naik mmHg Rectangular 0.100 √3 10 50 0.029 1 0.029
Drift Standar
7. Turun mmHg Rectangular 0.110 √3 10 50 0.032 1 0.032
8. Naik mmHg Normal 0.240 2 5 200 0.120 1 0.120
Sertifikat Standar
9. Turun mmHg Normal 0.240 2 5 200 0.120 1 0.120
Naik
Jumlah / Sum
Turun
Naik
Ketidakpastian Gabungan (Uncertainty Combinance / Uc)
Turun
Naik
Derajat kebebasan efektif / Effective degrees of freedom, u eff
Turun
Naik
Faktor cakupan / coverage factor, k=Student's for u eff and CL 95%
Turun
Naik
Ketidakpastian bentangan / Expanded uncertainty, U = k uc , mmHg
Turun
(Ui.c)2 (uici)4/vi No. Pembacaan Komponen Ketidakpastian Satuan Distribusi U Divisor Reliability vi
Alat
( mmHg)
0.083 0.0001 1. Daya Baca Operator mmHg Rectangular 0.500 √3 10 50
0.000 0.0000 2. Zero Error mmHg Rectangular 0.000 √3 10 50
0.002 0.0000 3. Histerisis mmHg Rectangular 0.047 √3 10 50
0.007 0.0000 4. Naik mmHg Rectangular 0.130 √3 3 2
Repeatability
0.013 0.0001 5. 200 Turun mmHg Rectangular 0.150 √3 3 2
200
0.001 0.0000 6. Naik mmHg Rectangular 0.850 √3 10 50
Drift Standar
0.002 0.0000 7. Turun mmHg Rectangular 0.860 √3 10 50
0.014 0.0000 8. Naik mmHg Normal 0.240 2 5 200
Sertifikat Standar
0.014 0.0000 9. Turun mmHg Normal 0.240 2 5 200
0.1073 0.0002
Jumlah / Sum
0.1150 0.0002
0.33
Ketidakpastian Gabungan (Uncertainty Combinance / Uc)
0.34
71.40
Derajat kebebasan efektif / Effective degrees of freedom, u eff
58.02
1.99
Faktor cakupan / coverage factor, k=Student's for u eff and CL 95%
2.00
0.65
Ketidakpastian bentangan / Expanded uncertainty, U = k uc , mmHg
0.68
(Ui.c)2 (uici)4/vi No. Pembacaan Komponen Ketidakpastian Satuan Distribusi U Divisor Reliability vi
Alat
( mmHg)
0.083 0.0001 1. Daya Baca Operator mmHg Rectangular 0.500 √3 10 50
0.000 0.0000 2. Zero Error mmHg Rectangular 0.000 √3 10 50
0.001 0.0000 3. Histerisis mmHg Rectangular 0.043 √3 10 50
0.015 0.0001 4. Naik mmHg Rectangular 0.180 √3 3 2
Repeatability
0.004 0.0000 5. 250 Turun mmHg Rectangular 0.229 √3 3 2
0.009 0.0000 6. Naik mmHg Rectangular 0.990 √3 10 50
Drift Standar
0.010 0.0000 7. Turun mmHg Rectangular 1.000 √3 10 50
0.014 0.0000 8. Naik mmHg Normal 0.240 2 5 200
Sertifikat Standar
0.014 0.0000 9. Turun mmHg Normal 0.240 2 5 200
0.1223 0.0002
Jumlah / Sum
0.1122 0.0001
0.35
Ketidakpastian Gabungan (Uncertainty Combinance / Uc)
0.33
60.11
Derajat kebebasan efektif / Effective degrees of freedom, u eff
84.01
2.00
Faktor cakupan / coverage factor, k=Student's for u eff and CL 95%
1.99
0.70
Ketidakpastian bentangan / Expanded uncertainty, U = k uc , mmHg
0.67
T
Ui c (Ui.c) (Ui.c)2 (uici)4/vi
Merek : Riester
Model/Tipe : Nova Presameter
No. Seri : 030178304
Tanggal Diterima : 11 Januari 2017
Tanggal Kalibrasi : 9 Februari 2017
Tempat Kalibrasi : Laboratorium Tekanan BPFK Surabaya
Nama Ruang :-
Metode Kerja : MKT-01
I. Kondisi Ruang
1. Suhu : 21.2 ± 0.14 °C
2. Kelembaban : 73.2 ± 0.3712310601
IV. Keterangan
1. Kesalahan maksimal yang diijinkan + 3 mmHg sesuai dengan OIML R 16-1, Edition 2002 (E)
2. Penentuan Kesalahan Maksimal dengan menjumlahkan nilai Kesalahan & nilai Ketidakpastian Pengukuran
3. Ketidakpastian pengukuran diestimasi pada tingkat kepercayaan 95%, k=2
4. Hasil pemeriksaan kondisi fisik & fungsi alat baik.
5. Konversi satuan; 1 mmHg = 133.3 Pa.
6. Dari hasil pengukuran uji kebocoran alat dinyatakan aman.
7. Ganti bola tensi dan perbaiki engsel pada sphygmomanometer.
Menyetujui,
Kepala Instalasi Kalibrasi
Wiji, ST.
NIP. 196608081995031001
F-PMT-MKT-01
Dilarang keras mengutip / memperbanyak dan / atau mempublikasikan sebagian isi sertifikat ini tanpa ijin tertulis dari BPFK Surabaya
Sertifikat ini sah bila dibubuhi cap BPFK Surabaya dan ditandatangani oleh pejabat yang berwenang
Tanggal Paraf
Kesimpulan
Baik
Ketidakpastian Pengukuran
Naik Turun
(mmHg) (mmHg)
± 0.63 ± 0.63
± 0.65 ± 0.68
± 0.70 ± 0.67
± 0.68 ± 0.71
± 0.80 ± 0.81
± 0.87 ± 0.89
Merek : Riester
Model/Tipe : Nova Presameter
No. Seri : 030178304
Tanggal Diterima : 11 Januari 2017
Tanggal Kalibrasi : 9 Februari 2017
Tempat Kalibrasi : Laboratorium Tekanan BPFK Surabaya
Nama Ruang : -
Metode Kerja : MKT-01
I. Kondisi Ruang
1. Suhu : 21.2 ± 0.14 °C
2. Kelembaban : 73.2 ± 0.3712310601
IV. Keterangan
1. Kesalahan maksimal yang diijinkan + 3 mmHg sesuai dengan OIML R 16-1, Edition 2002 (E)
2. Penentuan Kesalahan Maksimal dengan menjumlahkan nilai Kesalahan & nilai Ketidakpastian Pengukuran
3. Ketidakpastian pengukuran dilaporkan pada tingkat kepercayaan 95 % dengan faktor cakupan k = 2
4. Hasil pemeriksaan kondisi fisik & fungsi alat baik.
5. Konversi satuan; 1 mmHg = 133.3 Pa.
Menyetujui,
Kepala Instalasi Kalibrasi
Wiji, ST.
NIP. 196608081995031001
F-PMT-MKT-01
Dilarang keras mengutip / memperbanyak dan / atau mempublikasikan sebagian isi sertifikat ini tanpa ijin tertulis dari BPFK Surabaya
Sertifikat ini sah bila dibubuhi cap BPFK Surabaya dan ditandatangani oleh pejabat yang berwenang
asi Sphygmomanometer Mercury
Label : 72 / 1 / II - 17 / E-037 Dt
Kesimpulan
Baik
Ketidakpastian Pengukuran
Naik Turun
(mmHg) (mmHg)
± 0.63 ± 0.63
± 0.65 ± 0.68
± 0.70 ± 0.67
± 0.68 ± 0.71
± 0.80 ± 0.81
± 0.87 ± 0.89
Wiji, ST.
NIP. 196608081995031001
Halaman 2 dari 2 halaman
atau mempublikasikan sebagian isi sertifikat ini tanpa ijin tertulis dari BPFK Surabaya
cap BPFK Surabaya dan ditandatangani oleh pejabat yang berwenang