Konsep Dasar Ventilasi Mekanis Dan Aplikasi Mode
Konsep Dasar Ventilasi Mekanis Dan Aplikasi Mode
Outline
■ SEJARAH VENTILASI MEKANIK
■ DEFINISI VENTILASI MEKANIK
■ FUNGSI VENTILASI MEKANIK
■ PEMBAGIAN DASAR VM
■ INDIKASI PEMAKAIAN VM
■ VENTILATOR PARAMETER
■ MODE VM
■ SETING DASAR VM
■ PEMANTAUAN VM
■ VENTILATOR WEANING
FUNGSI RESPIRASI
MEMENUHI KEBUTUHAN
METABOLISME SEL AKAN O2 DAN
MENGELUARKAN CO2 SEBAGAI
SISA METABOLISME SEL
Ventilasi
Proses transport gas antara alveoli dan udara
Terbatas pada :
Kelainan paru Obstruktif
Restriktif
Kombinasi
Hipoventilasi dapat disebabkan karena
Perfusi
Adalah aliran darah kapiler yang mengambil
O2 dan mengeluarkan CO2
Kelainan perfusi :
ARDS
VENTILASI PARU
AIRWAY COMPLIANCE
RESISTANCE (RAW) (COMPL)
RAW
AIRWAY
CL
LUNG
FAKTOR-FAKTOR YANG MEMPENGARUHI DIFUSI
UDARA BEBAS:
PiO2 : 20.9 % x 760 = 159 mmHg
PiCO2 : 0.04 % x 760 = 0.3 mmHg
PiN2 : 78.6 % x 760 = 597mmHg ALVEOLUS
PiH2O : 0.46 % x 760 = 3.5 mmHg
N2 H2O
KAPILER PARU
PAN2: PAH2O:
PROSES DIFUSI 573 mmHg 47 mmHg
PAO2: PACO2:
104 mmHg 40 mmHg
O2 O2 CO2 O2
PaO2:
PaO2: 104 mmHg
40 mmHg CO2
CO2 PaCO2: PaCO2:
45 mmHg 40 mmHg
ANATOMICAL
NOSE TRACHEA
BRONCHIOLI DEAD SPACE
V ; minute V/Q = ∝
ventilation 5 L/
ALVEOLAR
minute. DEAD SPACE
V/Q > 1
Q ; cardiac output
5 L/minute/m2 MECHANICAL DEAD
V SPACE:
■TUBE
Q V/Q = 1 NORMAL ■CONNECTOR
PULMONARY CAPILLARY ■ET CO2
■BREATHING CIRCUIT
V/Q < 1
VENOUS ADMIXTURE
(SHUNT)
V/Q = 0
SEJARAH VENTILATOR
“And he went up, and lay upon the child,
and put his mouth upon his mouth, and
his eyes upon his eyes, and his hands
upon his hands; And he stretched himself
upon the child and the flesh waxed warm”
(II Kings 4:34)
SEJARAH VENTILATOR
Sebelum 1900: Penggunaaan respirator u/ tujuan penelitian.
~1940 Penemuan Intermitten Positive Pressure Breath (IPPB) untuk “lung inflation
therapy” dan short term ventilation
SEJARAH VENTILATOR
Thailand tahun 2009
SEJARAH
1960’S
■ TIDAL VOLUME KONSISTEN
■ HANYA MONITORING AIRWAY PRESSURE
■ NAFAS MEKANIK BERDASARKAN UPAYA NAFAS
PASIEN
1970’S
■ INTERMITENT MANDATORY VENTILATION
1980’S
■ MICROPROCESSOR, ALARM SISTIM + PRESSURE LIMIT
FUNGSI VENTILATOR
■ DIRANCANG U/ MEMBANTU INSPIRASI DAN
EKSPIRASI PASIEN
■ BISA SECARA ELEKTRIK ATAU PNEUMATIK
■ HIGHLY RELIABLE, DURABLE AND PRECISE
■ PERKEMBANGAN MICROPROCESSOR AND
ELECTRONIC CONTROL MAKIN KOMPLEKS
■ HARGA RATA-RATA $ 30.000 – 40.000
■ SAAT INI TERSEDIA VENTILATOR KECIL DAN
MURAH UNTUK DIGUNAKAN DI RUMAH (HOME
VENTILATOR)
■ Memperbaiki oksigenasi
■ Membantu eliminasi carbondioksida
■ Membantu kerja otot pernafasan
VENTILATOR PARAMETER
■ Tidal volume (VT): jumlah udara
yang diberikan pada pasien tiap napas
(satuan: mL)
■ Respiratory rate/frequency(f):
jumlah napas (pasien/mesin/keduanya
dalam 1 menit (satuan: napas/menit)
■ Minute ventilation(MVE):jumlah
udara yang diberikan pada pasien
dalam 1 menit (satuan: L/menit).
Merupakan hasil perkalian tidal volum
dan respiratory rate.
VENTILATOR PARAMETER
TEKANAN POSITIVE
PRESSURE
TEKANAN NEGATIVE
TIME
0
Inspirasi Ekspirasi
Inspirasi adalah tekanan POSITIVE
Ekspirasi tekanan LEBIH NEGATIVE
TIME
0
Inspirasi Ekspirasi
KURVA NAFAS SPONTAN
DISTENDING PRESSURES
Plat P - Base P (PEEP) = pressure to distend resp system (lung + chest wall)
Plat P = peak alveolar pressure
Transpulmonary pressure = Plat P – Pleural pressure
Peak P - PlatP = pressure for flow
R = Flow/(PeakP - Plat P)
35 cm H2O 35 cm H2O
5 cm H2O 10 cm H2O
30 cm H2O 25 cm H2O
MODE VENTILATOR
KATEGORI MODUS VENTILASI MEKANIS
T
0
6 DETIK 6 DETIK 6 DETIK
T
0
4 DETIK 3 DETIK 5 DETIK
SIMV mode
Nafas Ventilator
P
Nafas Spontan Pasien
0 T
1. Contoh, Jika setting SIMV rate = 6. Berarti siklus SIMV = 60/6 = 10 detik
2. Jika RR pasien 20; berarti siklus respirasi pasien 60/20=3 detik, maka
periode SIMV (nafas ventilator) harus dibuat supaya mengikuti siklus
resp pasien yaitu 3 detik, bisa dengan menaikkan flow rate antara 60-80
L/menit atau setting T inspirasi
• BREATH –
Spontaneous
Assisted
MANDATORY
Triger : Patient/ Time
Limit : Volume
Cycle : Volume/Time
Breath – SPONTANEOU
Trigger – PATIENT
Limit - PRESSURE
Cycle – FLOW ( 5-25% OF
PIFR
Setelah ter-trigger, ventilator memberi suatu flow yg cukup utk menaikkan dan
mempertahankan tekanan jalan napas pd level yang preset selama waktu usaha napas
spontan pasien
Mode
Trigger Limit Cycling
Pressure Control Ventilation (PCV) Ventilator or Patient Pressure Time (Time also controls pause)
volume-cycled Flow
Ventilator or Patient Volume (mandatory breath)
Synchronised-Intermittent-Mandatory Ventilation (mandatory
(SIMV) breath)
Pressure
pressure-limited SIMV Ventilator or Patient Time (mandatory breath)
(mandatory
breath)
Note that where CPAP is combined with ventilator triggered modes, confusing terminology kicks in again - CPAP is then called
"PEEP" (Positive End-Expiratory Pressure).
POLA KURVA VENTILASI
■ PRESSURE-TIME DIAGRAM
■ FLOW-TIME DIAGRAM
■ VOLUME-TIME DIAGRAM
■ INTERPRETATION OF CURVE
PATTERNS
Pressure-time diagram
Volume-controlled, constant flow
Interpretation of PV loops in
ventilation.
P-V LOOP
EKSPIRASI
Vol
LOW HIGH
NORMAL COMPLIANCE COMPLIANCE
500 500 500
0 15 30 15 30 15 30
PEEP 5
INSPIRASI
NAFAS
SPONTAN
Pemantauan Ventilasi Mekanik
■ Penilaian Ventilasi dan Oksigenasi
■ Alarm
■ Klinis : inspeksi, palpasi, perkusi, auskultasi
– Analisis gas darah: pCO2, pO2, SaO2
– Foto toraks: posisi ujung ETT, paru, pleura
■ Monitor: RR, SpO2
■ Ventilator: tidal volume, respiratory rate
(RR), Pinsp, grafik flow, pressure, volume
Penilaian Ventilasi
■ Ventilasi: proses keluar masuknya udara
dari atmosfer ke dalam sistem respirasi
– Inspirasi
– Ekspirasi
■ Parameter:
– Tidal volume
– Respiration Rate
– Minute volume
Alarm
■ Pplateau < 30 cmH2O
■ RR < 6 x/m > 30 x/m
■ MV < 25 mL/kg > 125 mL/kg
■ PEEP < 3 cmH20
Ventilator Weaning
■ Penyapihan ventilator adalah proses
pelepasan dukungan mesin ventilator dan
mengembalikan kerja pernapasan dari
ventilator ke pasien
■ Penyapihan ventilator terdiri dari 2 tahap:
Penghentian dukungan
Pelepasan
mesin ventilator
endotracheal tube (ETT)
Indikasi Weaning
■ Penyakit dasar telah diobati dan membaik
■ Fungsi respirasi
– RR < 35 x/m
– FiO2 < 0.5, SaO2 > 90%, PEEP <10 cmH2O
– Tidal volume > 5ml/kg
– Minute volume < 10 l/min
■ Kardiovaskular stabil
■ Cairan dan elektrolit cukup
Penyapihan
Tiba-tiba Bertahap
■Pasiendengan ■Berlangsungharian,
ventilator untuk waktu mingguan bahkan
singkat bulanan
■+70%kasus ■+30%kasus
Parameter Penyapihan
Kriteria Pasien
■Sadar, kooperatif
■Hemodinamik stabil
■Frekuensi napas <30 kali/menit
■Tanpa efek sedasi/blok neuromuskuler ▪
Sekresi minimal
■Status nutrisi baik
Berhasil Gagal
■Pasienharus kembali
■Pasien mampu
mendapat dukungan mesin
mempertahankan napas
ventilator setelah periode waktu
spontan dalam periode tertentu dengan napas spontan
waktu tertentu yang tidak terus-menerus
■Kebanyakan penelitian ■Sulit ditentukan, kebanyakan
menetapkan waktu yaitu penelitian berdasarkan hasil
24 jam kondisi klinis dan AGD
Kegagalan Penyapihan
Tanda awal pada pasien:
■Takipneu
■Nyeri dada
■Dada-perut asinkron – diaphoresis
Parameter Penyapihan
Indikator Kegagalan Penyapihan
Penyebab Kegagalan Penyapihan
1. Peningkatan resistensi aliran udara Evaluasi
penggunaan ETT yang sesuai dan benar
2. Berkurangnya compliance paru Evaluasi
berbagai kondisi yang bisa mengurangi compliance
3. Kelelahan otot napas Bisa akibat peningkatan
resistensi aliran udara, penurunan compliance, DO2
atau CO rendah, nutrisi kurang, gangguan
keseimbangan elektrolit, dll
Metode Penyapihan
Panduan Penyapihan Ventilasi Mekanik
CONCLUSION
CONCLUSION
MODE KEUNTUNGAN KERUGIAN
PSV • Pasien lebih nyaman • Bila apneu bisa berbahaya (meskipun alarm
akan berbunyi)
TERIMA KASIH