Anda di halaman 1dari 3

SYARAT PENGAMBILAN SPERMA

1. PASIEN PUASA MEROKOK SELAM 3 HARI


2. TIDAK BERHUBUNGAN BADAN SELAMA 3 HARI SEBELUM PENGAMBILAN SAMPEL
3. TIDAK MINUM ALKOHOL
4. PASIEN DATENG JAM 8 PAGI PADA WAKTU YG DITENTUKAN

NOTE :

KALO PASIEN DATENG JGN LUPA MINTA NO TELPON SAMA KTP PASIEN, TRUS DAFTARIN , SURUH
BAYAR KEKASIR , PASIEN DIKASI NO KAK AGUS KALO MAU NANYAIN HASIL

ANALISIS SPERMA

LIQUIFAKSI : STOPWATCH DARI SPERMA DATANG SAMPAI ENCER


VOLUME : UKUR SPERMA DENGAN GELAS UKUR
PH : BACA DENGAN KERTAS PH KETIKA SPERMA DATANG
BAU : KHAS /TIDAK KHAS
JUMLAH :Nacl 100 Ul + 5 ul sperma -> homogenkan -> teteskan dikamar hitung ->
tutup dengan deck glass ( dr zul yg baca)
WARNA : ABU-ABU/ KEKUNINGAN/MERAH / KECOKLATAN
MOTILITAS : TETESKAN SPERMA DI OBJEK GLASS -> TUTUP DENGAN DECK GLASS -> BACA
DIPERBESARAN 40X10 ( DR ZUL YG BACA)
MORFOLOGI :BUAT APUSAN SPERMA SEPERTI APUSAN DARAH -> FIKSASI -> JIKA SUDAH
KERING TETESKAN GIEMSA KE SELURUH APUSAN 15 MENIT -> CUCI AIR
MENGALIR
ALAT KIMIA ANALYZER

PASSWORD = #BS8A#SEU (CAPSLOCK TULISANNYA)

PEMBACAAN SAMPEL DIMULAI DARI NO 3, NO 1 KHUSUS CALIBRATOR, NO 2 KHUSUS


QC/KONTROL

CARA KALIBRASI SEKALIGUS QC

1. KLIK REAGEN YANG GA MASUK ( NO NYA ) SAMPE BIRU


2. MASUKAN REAGEN KALIBRASI DI NO 1 , QC NO 2
3. PILIH TULISAN CAL F5 , KLIK KALIBRASI, KLIK OK
4. KLIK PROGRAM -> QUALITY CONTROL -> KLIK REAGEN YANG DI KALIBRASI -> KLIK SAVE ->
RUNNING
5. LIHAT SUDAH MASUK ATAU BELUM REAGEN YG DIKALIBRASI DAN QC -> RESULT-> CURRENT

MEMASUKKAN REAGEN BARU

1. MENU REAGEN, LIHAT REAGEN YG SUDAH ABIS DI NO BERAPA


2. MASUKKAN REAGEN BARU SESUAI NOMOR
3. KLIK END LOAD, KLIK INVENTORY
4. KLIK FOLLOWING POSITION, ISI DIDALAM KOTAK REAGEN ADA DIPOSISI BERAPA ( MISAL NO
18 UREA ), KLIK CHEK -> LIHAT REAGEN SUDAH TERISI ATAU BELUM
5. JIKA REAGEN YG DIMASUKKAN LEBIH DARI SATU TULIS DI FOLLOWING POSITION 1,2 ( PAKE
KOMA YA BUKAN TITIK)

JIKA REAGEN BERWARNA MERAH DI MENU SAMPLE/ REAGEN TIDAK MASUK

1. BUKA MENU REAGEN


2. LAP BARCODE TEMPAT REAGEN DIDALAM ALAT DENGAN TISU SAMPAI KERING->LETAKKAN
KEMBALI REAGEN KEDALAM ALAT
3. KLIK END LOAD , JIKA TULISAN SUDAH HITAM BERARTI SUDAH MASUK

JIKA KALIBRASI ABIS

1. QC-> QC SET UP (LIAT NILAI MEAN)


2. REAGEN -> CAL SET UP -> UBAH NILAI CONC SAMA KAYA MEAN DI QC -> SAVE
3. LANGSUNG KALIB DAN QC PAKE REAGEN QC

JIKA REAGEN DI NO 39 HABIS

1. AMBIL TEMPAT REAGEN DI NO 39 DALAM ALAT , ISI DENGAN DTERGEN D 80 SAMPAI FULL
2. KLIK REAGEN
3. KLIK WASH D SAMPAI BIRU
4. KLIK LOAD F1
5. DIDALAM KOTAK , ISI VOLUME 100
6. KLIK LOAD F3
7. KLIK EXIT
8. KLIK ENDLOAD
TAMBAH PEMERIKSAAN

1. KLIK PROGRAM , SAMPEL


2. DEMOG, TULIS POSISI SAMPEL DI NO BERAPA
3. KLIK ENTER , SAVE
4. TAMBAH PEMERIKSAAN YG DIINGINKAN
5. SAVE , RUNNING

STATUS SLEEP PADA ALAT ( DIPOJOK KIRI ATAS)

1. KLIK UTILITY , COMMANDS


2. KLIK WAKE UP , STATUS AKAN BERTULISKAN INCUBATION , TUNGGU ALAT SAMPAI TULISAN
READY/ STANDBY

MEN DUPLO PEMERIKSAAN

1. KLIK RESULT , PILIH SAMPEL,PILIH PEMERIKSAAN


2. KLIK RERUN PADA BAGIAN BAWAH , KLIK PLAY/ RUNNING

MENAMBAH PEMERIKSAAN

1. KLIK PROGRAM
2. PADA TULISAN POS DIISI PEMERIKSAAN NO BERAPA YANG AKAN DITAMBAH (MISAL NO 6)
3. KLIK ENTER PADA KOMPUTER
4. TAMBAH PEMERIKSAAN , KLIK SAVE , KEMUDIAN RUNNING SAMPEL

NOTE:

CALIBRASI SEMUA PEMERIKSAAN PAKE REAGENT CALIB KECUALI UNTUK REAGENT CKMB
CALIBRASINYA KHUSUS PAKE QC