Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

CALIBRATION OF LABORATORY GLASSWARE AND PUMPKIN


MEASURE IN MAKING STOCKS STERILE

Cahyani Putri Ngulwiyah, Ina Suci Pratiwi, Iwan Santoso

of Pharmacy Faculty of Mathematics and Natural Sciences,


University of Sriwijaya
Email: farmasiunsri2015@gmail.com

ABSTRACT

Kalibrasi mempunyai pengertian sebagai rangkaian kegiatan membandingkan


hasil pengukuran suatu alat dengan alat standard yang sesuai untuk menentukan besarnya
koreksi pengukuran alat serta ketidak pastian nya. Alat ukur yang telah dikalibrasi tidak
akan secara terus- menerus berlaku masa kalibrasi nya, karena peralatan tersebut selama
masa penggunaan nya pasti mengalami perubahan spesifikasi akibat pengaruh frekuensi
pemakaian, lingkungan penyimpanan, cara pemakaian, dan sebagainya. Hasil pengukuran
yang diberikan oleh beberapa alat sejenis tidak selalu menunjukan hasil yang sama,
meskipun alat tersebut mempunyai tipe yang sama. Perbedaan ini diperbesar lagi dengan
adanya pengaruh lingkungan, operator, serta metode pengukuran. Agar setiap alat dapat
mmemberikan hasil ukur dengan keabsahan yang sama alat ukur tersebut perlu
mempunyai ketelusuran kepada standard nasional atau standard internasional. Cara untuk
memberikan jaminan bahwa alat yang digunakan mempunyai ketelusuran kepada
standard nasional adalah dengan melakukan kalibrasi terhadap alat tersebut.

Keywords: Calibration , Method, Autokaf, Oven, Bacteria

I. INTRODUCTION Secara umum kalibrasi


mempunyai pengertian sebagai

1 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

rangkaian kegiatan membandingkan adanya pengaruh lingkungan, operator,


hasil pengukuran suatu alat dengan alat serta metode pengukuran. Padahal dalam
standard yang sesuai untuk menentukan menghasilkan hasil pengukuran tersebut
besarnya koreksi pengukuran alat serta sangat diharapkan bahwa setiap alat ukur
ketidak pastian nya. Dalam pengertian yang digunakan dimanapun memberikan
ini alat standard yang digunakan juga hasil ukur yang samadalam kaitan nya
harus terkalibrasi dibuktikna dengan dengan keperluan keamanan, kesehatan
sertifikat kalibrasi. dengan demikian transaksi, dan keselamatan (Aulton,
maka besarnya koreksi pengukuran alat 1994)
dapat ditelusurkan ke standard nasional
Agar setiap alat dapat
atau standard internasional dengan suatu
mmemberikan hasil ukur dengan
mata rantai kegiatan kalibrasi yang
keabsahan yang sama alat ukur tersebut
tidak terputus (Akers, 2010)
perlu mempunyai ketelusuran kepada
Alat ukur yang telah dikalibrasi
standard nasional atau standard
tidak akan secara terus- menerus
internasional. Cara untuk memberikan
berlaku masa kalibrasi nya, karena
jaminan bahwa alat yang digunakan
peralatan tersebut selama masa
mempunyai ketelusuran kepada standard
penggunaan nya pasti mengalami
nasional adalah dengan melakukan
perubahan spesifikasi akibat pengaruh
kalibrasi terhadap alat tersebut. Lebih
frekuensi pemakaian, lingkungan
dari itu untuk memelihara ketelusuran
penyimpanan, cara pemakaian, dan
tersebut perlu dilakukan perawatan alat
sebagainya. Untuk itulah selama
dalam selang waktu kalibrasi tertentu.
berlaku nya masa kalibrasi alat
Ketelusuran pengukuran tidak hanya
bersangkutan perlu dipelihara
sekedar menjadi penyaratan
ketelusuran nya dengan cara perawatan
administrative, melainkan telah menjadi
dan cek aatara secara periodic (Felton,
kebutuhan teknis yang mendasar
2012)
terutama dengan diwajibkannya
Hasil pengukuran yang diberikan mencantumkan utimasi ketidakpastian
oleh beberapa alat sejenis tidak selalu dalam hasil uji (Martin, 1993)
menunjukan hasil yang sama, meskipun
alat tersebut mempunyai tipe yang sama. Tujuan dilakukannya kalibrasi
Perbedaan ini diperbesar lagi dengan untuk mencapai ketelusuran

2 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

pengukuran. Hasil pengukuran dapat di kedap getaran. Alat yang di kalibrasi


kaitkan atau di telusur sampai kestandar dalam keadaan berfungsi baik atau tidak
yang lebih tinggi atau teliti (standard rusak (Akers, 2010)
primer nasional dan atau internasional).
II. RESEARCH METHODOLOGY
Melalui rangkaian perban dengan yang
tak terputus. Menentukan deriasi a. Place and Time
(penyimpangan) kebenaran nilai
konfesional penunjukkan suatu Activities carried out in the

instrument ukur menjamin hasil-hasil Laboratory of Pharmaceutical

pengukuran sesuai dengan standard Technology Sriwijaya University, on

nasional maupun sesuai dengan Tuesday, October 24, 2017 from 10:00

standard internasional (Martin, 1993) am until 18:00 pm

Manfaat kalibrasi yakni menjaga


b. Equipment and Materials
kondisi imstrumen ukur dan bahan ukur
agar tetap sesuai dengan spesifikasinya. The tools used in this experiment
Untuk mendukung system mutu yang include oven (1 piece), flask (1 piece),
diterapkan di berbagai industri pada measuring cups (1 piece), thermometers
perawatan laboratorium dan produksi (1 piece), analytical balance (1 piece),
yang dimiliki. Bisa mengetahui calculator (1 piece), and weights (1 set).
perbedaan (penyimpangan) antara harga Meanwhile 100ml distilled water
benar dengan harga yang di tunjukkan necessary materials
oleh alat ukur. Persyaratan kalibrasi
yakni standard acuan yang mampu c. Procedure
telusur ke standard nasional atau
1) Calibration of the measuring cup
standard internasional. Metode kalibrasi
measuring cup washed and dried,
yang di akui secara nasional atau
and then weighed measuring cup blank
internasional, personil kalibrasi yang
is filled with distilled water to mark
terlatih, yang dibuktikan dengan
boundaries, kept cool to 20 degrees
sertifikasi laboratorium yang
Celsius, then weigh the measuring cup
terakredetasi. Ruangan atau tempat
berisi distilled water, measure and
kalibrasi yang terkondisi, seperti suhu,
record the weighing results, calculate
kelembapan udara, aliran udara dan juga
the volume of the average and standard

3 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

deviation and tolerance. Repeat these 20 degrees Celsius, then weighed flask
steps 3x which berisi distilled water, measure
and record the results of weighing,
2) Calibration flask calculating the mean volume and
flask washed and dried, and then standard deviations and tolerance.
weighed empty flask filled with distilled Repeat these steps 3x
water to mark boundaries, kept cool to
III. RESULTS AND DISCUSSION
Labu Ukur

Suhu (C ) FK V Pindah V Ssebenarnya % Galat


20 1,0038 38,42 ml 50,19 76,5 %
20 1,0040 38,40 ml 50,20 76,49 %
20 1,0042 38,41 ml 50,21 76,49 %
20 1,0045 38,41 ml 50,22 76,48 %

Volume sebenarnya : massa air x Fk

I 50 ml x 1.0038 = 50.19
II 50 ml x 1.0040 = 50.20
III 50 ml x 1.0042 = 50.21
I 50 ml x 1.0045 = 50.22
V

Value correction = x 100%

I 38.42 / 50.19 x 100% = 76.5%

4 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

II 38.40 / 50.20 x 100% = 76.49%


III 38.41 / 50.21 x 100% = 76.49%
IV 38.41 / 50.22 x 100% = 76.48%

So, calibrated measuring flask error because% > 50% and <100%

Gelas ukur

Suhu (C) FK V Pindah V Sebenarnya % Galat


20 1,0038 0,16 50,19 61 %
20 1,0040 0,65 50,20 65 %
20 1,0042 0,61 50,21 61 %
20 1,0045 0,59 50,22 59 %

volume of water actually = mass x Fk

I 50 ml x 1.0038 = 50.19
II 50 ml x 1.0040 = 50.20
III 50 ml x 1.0042 = 50.21
I 50 ml x 1.0045 = 50.22
V

Value correction = V Pindah – V Sebenarnya

I 50.80 - 50.19 x 100% = 61 %

II 50.85 - 50.20 x 100% = 65 %

5 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

III 50.82 - 50.21 x 100% = 61 %

IV 50.81 - 50.22 x 100% = 59 %

So, calibrated measuring flask error because%> 50% and <100%

In the lab this time doing practical konvensional a measuring instrument.


work regarding the calibration praktikan Guaranteeing things measurement in
glassware. In general, the calibration accordance with national and
instrument calibration glass or glass international standards. Calibration is
have an understanding as a series of done is also useful to maintain the
activities to compare the results of condition of measuring instruments and
measurements of a tool with the measuring materials in keeping with the
appropriate standard tool for specifications. To support the quality
determining the amount of correction system is applied in various industries
and lack kepastiannya measurement in the laboratory and production
tool. Digunakan standard tools must equipment owned. Can estimate the
also be calibrated according to the difference (deviation) between the price
requirements of national and indicated on the measuring instrument.
international standards. Calibrated The calibration method that do must
measuring devices will not continuously conform to established procedures.
apply the calibration period, because at Volume measuring devices are
the time of use is still changing calibrated in this trial include the flask
specifications. and a measuring cup. Flask is a measure
Calibration is done aiming to of the volume containing a quantity of
achieve ketelusuran measurement. The liquid filled to the mark. Measuring cup
measurement results can be attributed or is a tool used to measure how banyak
traced back to a higher standard or substances used or a liquid solution that
thorough (primary standard of national will be used in accordance with the up
and international), through rangkaian to the desired mark.
perbandingan unbroken. Determine Calibration of measuring is done
deviations (storage) truth value to adjust the output or indication of a

6 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

volume measuring device to match the the alcohol quickly evaporates so as to


magnitude of the standards used within speed up drying.
specified accuracy. The principle of Terms of a calibration among the
measuring instrument calibration is reference standard that is capable of
performed by measuring the weight of searching to national or international
the volume of the volume of distilled standards and calibration methods are
water released by the measuring recognized nationally / internationally,
instrument volume. This weight is then personnel calibration trained as
compared with the specific gravity of evidenced by certification from an
water at a temperature of volume accredited laboratory, room calibration
measurement is done so that it can be conditioned such as temperature,
determined the value of accuracy. humidity, air pressure, air flow, which
For the experimental procedure in tool is calibrated in a state of well
a flask used is cleaned and dikeringkan functioning / not damaged.
(30 minutes, 100°C). flask from the From the calibration is done to get
oven allowed to stand for a minute, put the result, among others, object value
in ekrilator. Flask weighed measurement, correction values or
appropriately. flask filled with distilled deviations, the value of the uncertainty
water to the mark, the volume of 1g of of measurement (magnitude of the error
water at various temperatures into which may occur in the measurement),
account. Experiments conducted triplo. evaluated after the work is measured
In a measuring cup, the same and analyzed uncertainty properly
experiment was performed and almost taking into account all available
identical to the flask experiments. They resources in the fancy comparison
differed only in the measurement until method used and the amount of errors
the mark is 50 ml. To not use dryer that occur. The nature of other
drying apparatus for high temperature metrology, such as the calibration
ukur tool that expands and tools are not factor, the calibration curve. To get the
in standard conditions. To speed up the accurate mass measurements, booyancy
drying process tools can use blotting effect (due to the buoyancy of objects in
paper and enough alcohol and allowed the air stream). air at standard
to stand in the air. Due to the nature of conditions with the current situation is

7 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

different weighed. Thus causing a 0.07650. test results flask calibrated


difference booyancy on these tools instrument with good reason percent
(equilibrium level when the two objects error percent error or qualify. Where the
have the same mass) condition that the measuring instrument
to know how big the error of the can be calibrated by either having a rate
measurement results, it can be greater than 50% but not less than
calculated percent error. Where the 100%.
percent error is an absolute error of the In the second test for measuring
actual value multiplied by 100%. in a cup has been cleaned up and put in the
measurement, the most important is the oven, dried, weighed in and filled the
extent to which the error is acceptable. void volume with distilled water or
In a measuring cup and flask method of distilled water. Similarly, the flask, the
calculation made the point where the first glass calibrating the measuring cup
differences occur in the calculation for used flask (50 ml measuring cup is done
the percent error. To flask% error or as much as 4x). volume moved obtained
errors obtained directly from the ranged between 50.80 ml - 50.82 ml.
volume switch and the actual volume by Error percent gain ranging between
means of the division. While measuring 59% to 65% or 0.059 or 0.065. it can be
cup minus. concluded from the results of a
The first test the calibration of the calibrated measuring cup for percent
flask, which had been cleaned and put error of more than 50% and less than
in the oven for the drying process. Do 100%.
as many as four times. Calibrating the Tool to be calibrated must be
first flask 50 ml flask used for the mass absolutely clean and dry so that the
of the empty flask is 56.44 grams and a results obtained are accurate, precise,
mass of flask containing water into and thorough. Measurements made a 3x
55.635 grams. Mass of water obtained so that results are more accurate.
from measurements of 45.53 grams. The results of the calibration
Flask volume is 50.19 ml and the among other objects measuring values,
percent average deviation obtained for correction or deviation value and the
replication 4x range between 76% - value of the uncertainty. Reference
76.5% or a range between 0.076 to standards are traceable to international

8 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

standards or nasiional. The calibration correction factor is useful to know how


method that is recognized nationally or big an error in the calibration of
internationally, calibration trained equipment. The greater the deviation
personnel. then diminishing the accuracy of
Volumetric is a quantitative measurement.
analysis of the amount of a substance
that is sought by reacting a solution REFERENCES
volume of the substance with a solution
Akers, MJ 2010, Sterile Drug Products:
of a standard substance of known
Formulation,Packaging,Manufact
concentration. Key to the success of a
uring, Informa Healthcare, New
measuring instrument is the similarity
York, USA
of the volume obtained. Umtuk
Aulton, ME, 1994, Pharmaceutics, The
decimate the error in the measurement,
Science of Dosage
the tools to be used needs to be
Forms and Dug Delivery
calibrated beforehand. Calibration can
System,William and
be done by comparing two data using
Wilkins, Parkway PA, USA
different measuring devices

Felton , L. 2012, Remington: Essential


IV. CONCLUSION
of Pharmaceutics, Philladelphia,
Factors that contain calibration of
USA
measuring instruments are temperature,
Martin,A.1993.PhysicalChemical
kelmbaban, the density, etc. To
Prinsiplesin
determine a deviation of the measuring
PharmaceuticalScience, Waverly
instrument, must be known in advance
International, Maryland, USA
and the water mass correction factor. A

9 Faculty of Mathematics and Natural Sciences, University of Sriwijaya


Journal of Pharmacy Technology preparations Sterile Sriwijaya University 2017

10 Faculty of Mathematics and Natural Sciences, University of Sriwijaya