KEBIJAKAN
b. Teknik Pemeriksaan
kV 120
Effective mAs 640
Rotation Time 0,5 sec
Acquisition 64 x 0,6 mm
Slice Colimation 0,6 mm
Slice Width 5,0 mm 1,0 mm
Feed/Rotation 26,9 mm
Pitch Factor 1,40
Increment 5,0 mm 0,7 mm
Kernel B30f B20f
CTDIvol 12,3 mGy
Effective Dose Male : 4,51 mSv
Female : 5.84 mSv
Contras Medium IV Injection untuk CT Abdo 2 Phase
Start delay 20 – 25 sec (ar)
50 – 75 sec (venous phase)
Flow rate 4 – 5 ml/sec
Total amount 100 – 120 ml