Anda di halaman 1dari 60

2014

Service Manual
Smart Steam Autoclave
SSA 11/12-006
Smart Steam
Autoclave SSA 11/12-006
Smart Steam Autoclave SSA
11/12-006

Oleh :
Gregorius Pasca Christi Kaisar(121313006)
Leander Adik Tri Sulistiono (111313006)
Smart Steam Autoclave SSA 11/12-006

DAFTAR ISI

1. Pendahuluan ....................................................................................................1
1.1. Penjelasan Alat ...........................................................................................1
1.2. Fungsi Alat .................................................................................................1
1.3. Cara Kerja Alat ..........................................................................................2
1.4. Komponen Utama ......................................................................................3
1.4.1 Sensor Suhu Thermocouple ............................................................3
1.4.2 Sensor Tekanan MPX-2100-DP .....................................................3
1.4.3 Sensor Capacitive Autonics LR18-8D............................................3
1.4.4 Sensor Level Air .............................................................................4
2. Bagian-Bagian Alat .........................................................................................5
2.1. Bagian Fisik dan Mekanik .........................................................................5
2.2. Bagian Elektronik ....................................................................................12
2.1.1 Bagian-bagian PCB......................................................................15
2.1.2 Rangkaian Elektronik...................................................................21
3. Penjelasan Penting ........................................................................................35
3.1. Cara Instalasi ............................................................................................35
3.2. Cara Memilih Menu .................................................................................36
3.3. Cara Pemeliharaan Alat ...........................................................................37
3.3.1 Pemeliharaan Rutin ........................................................................37
3.3.2 Service Ringan ...............................................................................39
3.4. Tindakan Darurat dan Troubleshooting ...................................................43
4. Pengoperasian Alat .......................................................................................48
4.1. Persiapan Alat dan Bahan ........................................................................48
4.2. Aturan Seting Pengoperasian ...................................................................48
4.3. Cara Melakukan Proses Sterilisasi ...........................................................49
4.4. Diagram Alir Proses Sterilisasi ................................................................50
5. Spesifikasi Alat ..............................................................................................51

Instrumentasi Medis Sanata Dharma Halaman ii


Smart Steam Autoclave SSA 11/12-006

DAFTAR GAMBAR

Gambar 1. Bagian chamber (bak sterilisasi) ............................................................4


Gambar 2. Bagian saklar power samping ................................................................6
Gambar 3. Tombol emergency stop .........................................................................6
Gambar 4. Bagian panel kontrol ..............................................................................7
Gambar 5. Kran dan saluran drainase .....................................................................8
Gambar 6. Saluran pengisian air ..............................................................................8
Gambar 7. Bagian tutup chamber ............................................................................9
Gambar 8. Object basket (keranjang objek).............................................................9
Gambar 9. Bagian water tank (bak penampung air) ..............................................10
Gambar 10. Elemen pemanas dalam chamber.......................................................11
Gambar 11. Sensor capacitive proximity ...............................................................11
Gambar 12. Rangkaian pengendali utama .............................................................12
Gambar 13. Bagian driver sensor suhu dan tekanan..............................................13
Gambar 14. Bagian solenoid valve (kran elektrik).................................................14
Gambar 15. Board catu daya..................................................................................15
Gambar 16. Board output.......................................................................................16
Gambar 17. Board sistem minimal ........................................................................17
Gambar 18. Board driver .......................................................................................18
Gambar 19. Board indikator proses diagram .........................................................19
Gambar 20. Board amplifier MPX2100DP ...........................................................20
Gambar 21. Board amplifier thermocouple ...........................................................20
Gambar 22. Rangkaian PCB board catu daya ........................................................21
Gambar 23. Rangkaian catu daya sederhana..........................................................22
Gambar 24. Rangkaian catu daya dengan regulator dan penguat arus ..................22
Gambar 25. Rangkaian catu daya dengan regulator untuk output tegangan
positif dan nagatif .............................................................................22
Gambar 26. Rangkaian PCB driver output ............................................................23
Gambar 27. Rangkaian driver output dengan pemisah sumber tegangan..............24
Gambar 28. Rangkaian PCB microcontroller unit.................................................25

Instrumentasi Medis Sanata Dharma Halaman iii


Smart Steam Autoclave SSA 11/12-006

Gambar 29. Rangkaian Sistem Minimum Mikrokontroler ATmega328P ............26


Gambar 30. Rangkaian Downloader dengan Komunikasi Serial RS-232 .............26
Gambar 31. Rangkaian konektor penghubung pada board mikrokontroler ..........27
Gambar 32. Rangkaian PCB driver input ..............................................................28
Gambar 33. Rangkaian skematik tombol (kiri) dan rangkaian PCB
tombol (Kanan) .................................................................................29
Gambar 34. Rangkaian skematik konfigurasi LCD (kiri) dan rangkaian
PCB LCD (kanan).............................................................................29
Gambar 35. Rangkaian driver sensor capacitive untuk mikrokontroler ................30
Gambar 36. Rangkaian pengendali otomatis dengan input photosensor ...............30
Gambar 37. Rangkaian pengendali otomatis water level.......................................31
Gambar 38. Rangkaian PCB LED indikator proses...............................................32
Gambar 39. Rangkaian skematik pengkondisi sinyal thermocouple .....................33
Gambar 40. Rangkaian PCB pengkondisi sinyal thermocouple ............................33
Gambar 41. Rangkaian skematik pengkondisi sinyal MPX2100DP .....................34
Gambar 42. Rangkaian PCB pengkondisi sinyal MPX2100DP ............................34
Gambar 43. Bagian panel kontrol ..........................................................................36
Gambar 44. Board indikator ..................................................................................39
Gambar 45. Diagram alir cara melakukan sterilisasi .............................................50

Instrumentasi Medis Sanata Dharma Halaman iv


Smart Steam Autoclave SSA 11/12-006

DAFTAR TABEL

Tabel 1. Daftar menu, nilai seting, beserta keterangan ..............................................37


Tabel 2. Troubleshooting ...........................................................................................46
Tabel 3. Pengaturan batas waktu, tekanan, dan temperatur untuk
berbagai macam hal yang akan disterilkan ..................................................48
Tabel 4. Spesifikasi alat .............................................................................................51

Instrumentasi Medis Sanata Dharma Halaman v


Smart Steam Autoclave SSA 11/12-006

1. Pendahuluan

1.1. Penjelasan :

SSA(Smart Steam Autoclave) adalah sebuah perangkat sterilisasi yang

menggunakan uap bertekanan tinggi dalam ruang tertutup untuk

mendekontaminasi atau mensterilkan peralatan.

SSA merupakan alat khusus yang digunakan untuk mensterilkan peralatan

medis dan alat kesehatan lainnya. Dengan SSA diharapkan beberapa jenis

bakteri, virus, jamur, dan mikroorganisme lainnya menjadi tidak aktif (steril).

Namun tetap saja ada beberapa mikroorganisme yang tetap dapat bertahan

pada suhu diatas 121°C. Autoclave merupakan salah satu sarana sterilisasi

alat medis.

1.2. Fungsi :

SSA ini dapat digunakan untuk mensterilkan peralatan kesehatan maupun

peralatan labolatorium pasca penggunaan terhadap pasien/objek yang dapat

menimbulkan kontaminasi pada peralatan tersebut.

Bahan-bahan yang dapat disterilkan oleh autoclave ini antara lain : karet,

kain, perabot, instrumen, dan botol cairan.

Instrumentasi Medis Sanata Dharma Halaman - 1 -


Smart Steam Autoclave SSA 11/12-006

1.3. Cara Kerja :

SSA ini bekerja dengan mengendalikan elemen pemanas untuk

memanaskan air di ruang tertutup yang kedap udara. Air dalam chamber ini

berasal dari penampungan air yang terdapat dalam alat yang akan mengisi

secara otomatis ke dalam chamber. Banyaknya air yang dibutuhkan telah di

set secara bawaan saat proses mulai dijalankan. Setelah air mencapai batas

pengisian air dalam chamber maka pengisian terhenti dan dilanjutkan dengan

pemanasan air oleh heater untuk mendapatkan suhu dan tekanan yang

diminta user. Pemantauan suhu dan tekanan dilakukan dengan sensor suhu

Thermocouple tipe-K dan sensor tekanan MPX-2100-DP. Pemantauan ini

bertujuan untuk memonitor secara langsung dan menjaga kestabilan suhu

maupun tekanannya. Saat suhu mencapai batas yang diminta, pengitung waktu

mundur mulai berjalan, lamanya proses sterilisasi tergantung dari bahan yang

akan disterilkan sesuai yang diatur oleh user.

Setelah proses sterilisasi selesai dengan otomatis alat akan membuang air

dan tekanan di dalam chamber. Setelah air dan tekanan kosong alat akan

langsung memberi peringatan bahwa proses telah selesai dan akan

menampilkan total waktu berjalan terhitung dari setelah tombol start/oke di

tekan.

Instrumentasi Medis Sanata Dharma Halaman - 2 -


Smart Steam Autoclave SSA 11/12-006

1.4. Komponen Utama:

1.4.1 Sensor Suhu Thermocouple

Sensor suhu thermocouple yang dipakai disini menggunakan

thermocouple tipe K, yaitu thermocouple yang disusun dari nikel dan

kromoium pada sisi positif dan pada sisi negative menggunakan nikel dan

aluminium. Pemakaian sensor suhu ini berdasar pada harga yang relatif lebih

murah dibanding thermocouple tipe lain dan rentang suhu yang lebar antara 0

– 1100oC. Karena keluaran thermocouple relative kecil dan belum dapat

dibaca pada mikrocontroler maka diperlukan penguat instrumentasi ad 620.

(Gbr. 1-E).

1.4.2 Sensor Tekanan MPX-2100-DP

Sensor tekanan MPX 2100 DP adalah sensor tekanan yang relatif dapat

digunakan pada tekanan antara 0 – 100kPa. Sensor ini digunakan untuk

mengukur besarnya tekanan di dalam tabung sterilisasi pada saat sterilisasi

dijalankan atau digunakan. Sensor tekanan keluaran Motorola ini dapat

dipasangkan dengan arduino dengan penambahan penguat instrumentasi yang

disusun dengan menggunakan IC Opamp op07(Gbr. 13-B).

1.4.3 Sensor Capacitive Autonics LR18-8D

Sensor capasitive ini digunakan untuk mensensor banyaknya air yang

dialirkan apakah sudah sesuai dengan yang diinginkan atau tidak. Sensor ini

akan mendeteksi air di dalam wadah penampungan air di dalam modul alat

dan bila melewati sensor ini maka sensor capasitive akan mati. Sensor yang

Instrumentasi Medis Sanata Dharma Halaman - 3 -


Smart Steam Autoclave SSA 11/12-006

digunakan adalah tipe PNP yaitu bila mensensor adanya air maka akan

mengeluarkan tegangan demikian pula berlaku sebaliknya (Gbr. 11).

1.4.4 Sensor Level Air

Sensor level air digunakan untuk memantau ketinggian air di dalam bak

penampung agar sewaktu autoclave dijalankan tidak kekurangan air. Sensor

level air ini dibangun dengan dua buah komponen aktif yaitu transistor

BC557 dan TIP31. Untuk mengetahui ketinggian air cukup atau kurang pada

bak penampung dipasang 3 transduser yang berupa kabel yang dipasang

berurutan. Ketika proses pengisian air akan tetap mengalir sampai batas

tranduser atas ditandai dengan nyala LED berwarna hijau menyala. Ketika

terjadi pemakaian/ pembuangan berlebih dan melewati tranduser ke dua maka

dinyatakan air kosong dan ditandai dengan nyala LED merah. Bersamaan

dengan nyala LED merah tersebut maka solenoid valve pengisi akan

membuka secara otomatis dan bak penampung akan terisi kembali sampai

level batas atas. Untuk menjamin ketersediaan air dalam bak penampung

dalam maka setiap pertama kali alat dihidupkan maka secara otomatis bak

penampung akan terisi sampai batas atas atau sampai dengan lampu LED

hijau menyala (Gbr. 9-B).

Instrumentasi Medis Sanata Dharma Halaman - 4 -


Smart Steam Autoclave SSA 11/12-006

2. Bagian-BagianAlat

2.1. Bagian Fisik dan Mekanik

E A

B
F

Gambar 1. Bagian chamber (bak sterilisasi)

KETERANGAN :

A. Handle pintu chamber


B. Safety valve
C. Pressure gauge
D. Baut pengunci (6 buah)
E. Sensor suhu thermocouple
F. Nepel tubing penghubung sensor tekanan

Instrumentasi Medis Sanata Dharma Halaman - 5 -


Smart Steam Autoclave SSA 11/12-006

A B C D

Gambar.2. Bagian saklar power samping

KETERANGAN :

A. Kabel power
B. Saklar power
C. Rumah sekring
D. Port komunikasi serial DB9

Gambar.3. Tombol Emergency Stop

Instrumentasi Medis Sanata Dharma Halaman - 6 -


Smart Steam Autoclave SSA 11/12-006

F
E

A B C D

Gambar.4. Bagian panel kontrol

KETERANGAN :

A. Tombol Naik
B. Tombol Turun
C. Tombol Geser
D. Tombol Oke
E. LCD Monitor
F. Indikator bak penampungan air

Instrumentasi Medis Sanata Dharma Halaman - 7 -


Smart Steam Autoclave SSA 11/12-006

Gambar.5. Kran dan saluran drainase

Gambar.6. Saluran pengisian air

Instrumentasi Medis Sanata Dharma Halaman - 8 -


Smart Steam Autoclave SSA 11/12-006

A B C D E F

Gambar.7. Bagian tutup chamber

KETERANGAN :

A. Seal pintu chamber


B. Lubang saluran pressure gauge
C. Lubang saluran buang safety valve
D. Lubang saluran sensor tekanan
E. Sensor suhu Thermocouple
F. Engsel pintu

Gambar.8. Object basket (Keranjang objek)

Instrumentasi Medis Sanata Dharma Halaman - 9 -


Smart Steam Autoclave SSA 11/12-006

A
B

D
E

Gambar.9. Bagian water tank (bak penampung air)

KETERANGAN :

A. Lubang pembuangan uap panas

B. Sensor level air

C. Tubing saluran drainase chamber

D. Tubing saluran pembuang tekanan berlebih

E. Tubing saluran pengisian bak penampung air

Instrumentasi Medis Sanata Dharma Halaman - 10 -


Smart Steam Autoclave SSA 11/12-006

Gambar.10. Elemen pemanas dalam chamber

Gambar.11. Sensor capacitive proximity

Instrumentasi Medis Sanata Dharma Halaman - 11 -


Smart Steam Autoclave SSA 11/12-006

2.2. Bagian Elektronik

F
L
E

D
G
M
C

A
H I J K N

Gambar.12. Rangkaian pengendali utama

KETERANGAN :

A. Konektor elemen heater


B. Papan lapisan pertama : rangkaian catu daya
C. Papan lapisan kedua : rangkaian driver output
D. Papan lapisan ketiga : rangkaian sistem minimal MCU
(Microcontroller Unit)
E. Papan lapisan keempat : rangkaian driver input
F. Papan lapisan kelima : rangkaian indikator proses input/output
G. Tombol reset mikrokontroler
H. Tombol By Pass solenoid valve pembuang tekanan
I. Tombol By Pass solenoid valve drainase chamber
J. Tombol By Pass solenoid valve pengisian air pada chamber
K. Tombol By Pass elemen heater
L. SSR (Solid State Relay)
M. Transistor penguat arus rangkaian MCU
N. Transformator CT 3A dan 500mA

Instrumentasi Medis Sanata Dharma Halaman - 12 -


Smart Steam Autoclave SSA 11/12-006

A B C D E

Gambar.13. Bagian Driver sensor suhu dan tekanan

KETERANGAN :

A. Konektor tubing sensor tekanan

B. Sensor tekanan MPX-2100DP

C. Konektor sensor suhu Thermocouple

D. Rangkaian pengkondisi sinyal suhu

E. Rangkaian pengkondisi sinyal tekanan

F. Box tempat bagian pengkondisi sinyal

Instrumentasi Medis Sanata Dharma Halaman - 13 -


Smart Steam Autoclave SSA 11/12-006

Gambar.14. Bagian solenoid valve (kran elektrik)

KETERANGAN :

A. Solenoid valve pengisian bak penampung air

B. Solenoid valve drainase chamber

C. Solenoid valve pembuang tekanan berlebih

D. Solenoid valve pengisian chamber

E. Filter air

Instrumentasi Medis Sanata Dharma Halaman - 14 -


Smart Steam Autoclave SSA 11/12-006

2.2.1 Bagian-bagian PCB

A B

C D E F G

Gambar 15. Board catu daya

Keterangan:

A. AC input ( 0, 9, 15, NC, 28)

B. AC input 12V

C. Output +5VDC

D. Output +12VDC

E. Output +24VDC

F. Output -9VDC

G. Output +9VDC

Instrumentasi Medis Sanata Dharma Halaman - 15 -


Smart Steam Autoclave SSA 11/12-006

A B C

D E F G

Gambar 16. Board Output

Keterangan:

A. Input data (sistem minimal C)

B. Output indikator proses diagram (Gbr. 19-D)

C. Input 12VDC

D. Output Heater

E. Output valve pengisian air pada chamber

F. Output valve drainase

G. Output valve pembuang tekanan

Instrumentasi Medis Sanata Dharma Halaman - 16 -


Smart Steam Autoclave SSA 11/12-006

A B C D E F G

N J K L M

Gambar 17. Board sistem minimal

Keterangan :

A. Input tombol

B. Water level

C. Keluaran / output

D. Proses diagram

E. Sensor MPX2100DP

F. Sensor thermocouple

G. Sensor kapasitif

H. Downloader serial

I. Tampilan LCD

J. Catu daya +5VDC

K. Catu daya +9VDC

L. Catu daya -9VDC

M. Catu daya +12VDC

N. Tombol reset

Instrumentasi Medis Sanata Dharma Halaman - 17 -


Smart Steam Autoclave SSA 11/12-006

A B D E F G H I J
C

K L M N O P Q R

Gambar 18. Board Driver

Keterangan :

A. Catu daya

B. LED indikator level air

C. Saklar pressure control

D. Rangkaian sistem minimal (E)

E. Solenoid valve pembuang tekanan berlebih

F. Sensor kapasitif

G. Rangkaian sistem minimal (G)

H. Tombol

I. Rangkaian sistem minimal (A)

J. Indikator

K. Solenoid valve pengisi air

L. Tombol by pass solenoid valve pengisi air

M. Panel indikator water level

Instrumentasi Medis Sanata Dharma Halaman - 18 -


Smart Steam Autoclave SSA 11/12-006

N. Sensor water level

O. Pemilih jenis air (RO/Aquadest)

P. LCD monitor

Q. LCD monitor

R. Rangkaian sistem minimal (I)

A B C D E

Gambar 19. Board indikator proses diagram

Keterangan :

A. Indikator tombol

B. Indikator sensor

C. Indikator keluaran (SSR)

D. Indikator keluaran (solenoid valve)

E. Catu daya utama

Instrumentasi Medis Sanata Dharma Halaman - 19 -


Smart Steam Autoclave SSA 11/12-006

B
A

Gambar 20. Board amplifier MPX2100DP

Keterangan :

A. Sensor MPX2100DP

B. Sistem minimal (E)

A B

Gambar 21. Board amplifier thermocouple

Keterangan :

A. Sensor suhu thermocouple tipe K

B. Sistem minimal (F)

Instrumentasi Medis Sanata Dharma Halaman - 20 -


Smart Steam Autoclave SSA 11/12-006

2.2.2 Rangkaian Elektronik

1. Board Catu Daya

Terdiri dari 5 rangkaian catu daya yang menghasilkan output berbeda-

beda yaitu : +5V, +9V, -9V, +12V, dan +24V. setiap rangkaian dalam board

catu daya akan dijelaskan dalam gambar-gambar di bawah ini.

Gambar 22. Rangkaian PCB board catu daya

Instrumentasi Medis Sanata Dharma Halaman - 21 -


Smart Steam Autoclave SSA 11/12-006

Gambar 23. Rangkaian catu daya sederhana

Gambar 24. Rangkaian catu daya dengan regulator dan penguat arus

Gambar 25. Rangkaian catu daya dengan regulator untuk output tegangan positif

dan nagatif

Instrumentasi Medis Sanata Dharma Halaman - 22 -


Smart Steam Autoclave SSA 11/12-006

2. Board Driver Output

Terdiri dari 4 buah rangkaian driver yang sama yang dilengkapi dengan

tombol by pass untuk menguji rangkaian driver apakah berfungsi dengan baik

atau tidak.

Gambar 26. Rangkaian PCB driver output

Instrumentasi Medis Sanata Dharma Halaman - 23 -


Smart Steam Autoclave SSA 11/12-006

Gambar 27. Rangkaian driver output dengan pemisah sumber tegangan

Instrumentasi Medis Sanata Dharma Halaman - 24 -


Smart Steam Autoclave SSA 11/12-006

3. Board Microcontroller Unit (MCU)

Terdiri dari rangkaian sistem minimum mikrokontroler, rangkaian

downloader dan konektor-konektor penghubung ke rangkaian lain.

Gambar 28. Rangkaian PCB microcontroller unit

Instrumentasi Medis Sanata Dharma Halaman - 25 -


Smart Steam Autoclave SSA 11/12-006

Gambar 29. Rangkaian Sistem Minimum Mikrokontroler ATmega328P

Gambar 30. Rangkaian Downloader dengan Komunikasi Serial RS-232

Instrumentasi Medis Sanata Dharma Halaman - 26 -


Smart Steam Autoclave SSA 11/12-006

Gambar 31. Rangkaian konektor penghubung pada board mikrokontroler

Instrumentasi Medis Sanata Dharma Halaman - 27 -


Smart Steam Autoclave SSA 11/12-006

4. Board Input

Terdiri dari beberapa rangkaian berbeda yaitu : rangkaian tombol,

rangkaian LCD, rangkaian pengendali tekanan otomatis, rangkaian driver

sensor capacitive, rangkaian pengendali water level otomatis. Masing-masing

rangkaian akan dijelaskan secara terpisah secara sederhana.

Gambar 32. Rangkaian PCB driver input

Instrumentasi Medis Sanata Dharma Halaman - 28 -


Smart Steam Autoclave SSA 11/12-006

Gambar 33. Rangkaian skematik tombol (kiri) dan rangkaian PCB tombol (Kanan)

Gambar 34. Rangkaian skematik konfigurasi LCD (kiri) dan rangkaian PCB LCD

(kanan)

Instrumentasi Medis Sanata Dharma Halaman - 29 -


Smart Steam Autoclave SSA 11/12-006

Gambar 35. Rangkaian driver sensor capacitive untuk mikrokontroler

Gambar 36. Rangkaian pengendali otomatis dengan input photosensor

Instrumentasi Medis Sanata Dharma Halaman - 30 -


Smart Steam Autoclave SSA 11/12-006

Gambar 37. Rangkaian pengendali otomatis water level

Instrumentasi Medis Sanata Dharma Halaman - 31 -


Smart Steam Autoclave SSA 11/12-006

5. Board Indikator Proses

Terdiri dari rangkaian LED yang disusun pada gambar dan

dihubungkan pada input dan output dari alat untuk mengetahui suatu input /

output sedang aktif atau tidak.

Gambar 38. Rangkaian PCB LED indikator proses

Instrumentasi Medis Sanata Dharma Halaman - 32 -


Smart Steam Autoclave SSA 11/12-006

6. Board Pengkondisi Sinyal Thermocouple

Merupakan rangkaian yang berfungsi menguatkan sinyal tegangan dari

thermocouple sebelum masuk ke mikrokontroler.

Gambar 39. rangkaian skematik pengkondisi sinyal thermocouple

Gambar 40. Rangkaian PCB pengkondisi sinyal thermocouple

Instrumentasi Medis Sanata Dharma Halaman - 33 -


Smart Steam Autoclave SSA 11/12-006

7. Board Pengkondisi Sinyal MPX2100DP

Merupakan rangkaian yang berfungsi menguatkan sinyal tegangan dari

sensor tekanan MPX2100DP sebelum masuk ke mikrokontroler.

Gambar 41. Rangkaian skematik pengkondisi sinyal MPX2100DP

Gambar 42. Rangkaian PCB pengkondisi sinyal MPX2100DP

Instrumentasi Medis Sanata Dharma Halaman - 34 -


Smart Steam Autoclave SSA 11/12-006

3. Penjelasan Penting

3.1. Cara Instalasi

Langkah yang perlu diperhatikan dalam proses instalasi, yaitu :

a) Install alat pada meja atau lantai dengan ketinggian alat maksimal

setinggi dada. Tempatkan alat didalam ruangan yang bersirkulasi

dengan suhu ruangan maksimal 28°C. Hindarkan alat dari cahaya

matahari langsung.

b) Install alat dekat dengan sumber air / kran air dan hubungkan bagian

saluran pengisian dengan sumber air menggunakan selang. Juga

hubungkan saluran pembuangan air dengan bagian pembuangan.

c) Hubungakan kabel power pada sumber listrik AC 220V. Untuk

menyalakannya tekan saklar power pada posisi I hingga terlihat

saklar power menyala berwarna orange diikuti dengan LCD display

menyala. Jika terjadi masalah pada proses menyalakan baca pada

bagian Troubleshooting.

Instrumentasi Medis Sanata Dharma Halaman - 35 -


Smart Steam Autoclave SSA 11/12-006

3.2. Cara Memilih Menu

Gambar 43. Bagian panel kontrol

Di dalam menu pada alat Autoclave SSA ini terdiri dari 5 sub

menu yang dapat diseting atau dirubah nilainya, antara lain pengaturan batas

suhu (Temp Set), pengaturan waktu sterilisasi (Time Set), pengaturan batas

tekanan (Press Set), pemilihan jenis mode basah / kering (tidak kompatibel

dengan versi saat ini), dan water cleaning / mode untuk mengganti air dan

membersihkan tangki air, Berikut ini adalah cara-cara untuk memilih menu :

a) Untuk merubah / memilih nilai setting tekan tombol naik / turun

b) Untuk mengganti atau menggeser menu tekan tombol geser

c) Untuk memverifikasi pengaturan dan melanjutkan ke proses sterilisasi

tekan tombol oke

Instrumentasi Medis Sanata Dharma Halaman - 36 -


Smart Steam Autoclave SSA 11/12-006

Saat alat pertama kali dinyalakan default setting-nya adalah 121 derajat

celcius untuk batas suhu, 15 menit untuk batas penghitung waktu mundur, dan

105 Kpa untuk batas tekanan. Jadi alat dapat langsung dioperasikan dengan

setingan awal tersebut dengan menekan tombol oke langsung.

Untuk lebih jelasnya isi menu dan sub menu dapat dilihat dalam tabel

berikut ini :

Tabel 1. Daftar menu, nilai seting, beserta keterangan

Default Perintah
Menu Nilai seting Satuan Keterangan
seting Eksekusi

Temp Set 100 – 134 121 °C ˄ atau ˅ Seting batas suhu

Time Set 0 – 60 15 Menit ˄ atau ˅ Seting batas waktu

Press Set 0 – 110 105 Kpa ˄ atau ˅ Seting batas tekanan

Metode sterilisasi yang


Mode Basah / Kering Basah ˄ atau ˅
digunakan

Water
 Wt. Tank > Chamber
Mati ˄ atau ˅ Modus penggantian air
 Chamber > Wt. Tank
Cleaning
 Mati

3.3. Cara Pemeliharaan Alat

3.3.1. Pemeliharaan rutin:

a. Kebersihan seal chamber

Proses sterilisasi memerlukan tekanan yang tinggi, oleh karena itu

kerapatan chamber penting untuk mencapai tekanan tersebut. Karet /

seal pada tutup chamber harus dijaga dari hal yang mengotori atau

dari hal yang berbau tajam. Bila karet/seal kotor atau rusak maka akan

Instrumentasi Medis Sanata Dharma Halaman - 37 -


Smart Steam Autoclave SSA 11/12-006

terjadi kebocoran dan proses sterilisasi tidak akan tercapai dengan

maksimal.

b. Kebersihan safety valve

Safety valve digunakan untuk menjamin tekanan dalam chamber pada

kondisi / tekanan maksimal pada nilai tertentu, bila tekanan

melampaui nilai tersebut akan sangat membahayakan dan bisa

berakibat tabung meledak. Pengecekan safety valve menyangkut

kebersihan dan fungsinya sangat diperlukan untuk mencegah

kecelakaan akibat pemakaian.

c. Kekencangan baut pengunci

Pada saat proses sterilisasi, penutup chamber harus benar-benar rapat

dan seimbang untuk menghindari kebocoran chamber dan mencegah

dari kemungkinan chamber meledak.

d. Tersedianya aquadest atau air yang cukup

Alat akan bekerja secara normal jika aquadest/air di bak

penampungan pada alat mencapai batas maksimal ditunjukan dengan

nyala LED berwarna hijau, bila nyala LED berwarna merah maka

aquadest masih kurang memenuhi. Bila selang masukan aquadest

dipasang maka secara otomatis bak penampungan akan terisi, bila

tidak dapat diisi secara manual lewat lubang bagian atas bak

penampungan aquadest/air.

e. Sesuaikan tombol seleksi dengan air yang digunakan

Instrumentasi Medis Sanata Dharma Halaman - 38 -


Smart Steam Autoclave SSA 11/12-006

Tombol seleksi masukan air di bagian belakang harus sesuai dengan

air yang digunakan, memakai aquadest atau air RO. Kesalahan

pemilihan pada tombol seleksi dapat menimbulkan kerusakan.

f. Kebersihan air aquadest / air yang digunakan

Aquadest atau air dalam tampungan air dalam harus dijaga

kebersihannya, bila dilihat banyak endapan kotoran harus segera

lakukan pengurasan untuk mencegah terjadinya kebuntuan saluran air.

3.3.2. Servis Ringan:

a) Cek kerusakan bagian elektronik

Autoclave SSA ini dilengkapi dengan self test, yaitu sesi dimana

alat ini akan menjalankan tes pada input dan output yang digunakan

pada alat, hasil dari test ini dapat dilihat pada papan indikator proses

yang terdapat didalam kotak pengendali, tes ini hanya akan berjalan

satu kali setiap alat baru dinyalakan. Jika dilihat pada papan lapisan

kelima bagian indikator proses, semua LED akan menyala selama

satu detik, sehingga jika terjadi malfungsi pada input/output LED

pada bagian tersebut tidak akan menyala, kemudian setelah tes

selesai input dan output akan menjalankan perannya masing-masing.

Instrumentasi Medis Sanata Dharma Halaman - 39 -


Smart Steam Autoclave SSA 11/12-006

Gambar 44. Board Indikator

Self test dan indikator ini memudahkan bagi teknisi untuk melihat

proses alat selama bekerja, jika terjadi malfungsi dalam sistem atau

terdapat input / output yang rusak bisa langsung diketahui melalui

indikator dan dapat langsung ditindak lanjuti oleh teknisi.

b) Water Cleaning, Membersihkan bak penampung dan Chamber

Lakukan proses ini secara berkala paling lama seminggu sekali,

autoclave ini didesain untuk dapat melakukan penggantian air sendiri

tanpa harus membongkar alat atau mengisi dan membuang air secara

manual. Dimana fitur ini dapat ditemui dalam sub menu “Water

Cleaning”. Didalamnya terdapat dua pilihan yaitu yang pertama “Wt

tank > Chamber” dan yang kedua “Chamber > Wt tank”. Untuk

pilihan yang kedua dapat bekerja hanya saat di dalam chamber

terdapat tekanan.

Dalam proses sterilisasi air yang diolah terus-menerus akan

membawa kotoran dan lendir / lumut yang dapat menyebabkan

terganggunya kinerja dari solenoid valve. Selain itu kebersihan

Instrumentasi Medis Sanata Dharma Halaman - 40 -


Smart Steam Autoclave SSA 11/12-006

terhadap air juga mengurangi dampak negatif lain terhadap air bekas

sterilisasi.

Saat dilakukan proses water cleaning disarankan untuk melakukan

pembersihan bagian dalam chamber menggunakan sikat dan cairan

pembersih seperti detergen.

c) Membersihkan Solenoid valve

Solenoid valve merupakan bagian yang sangat penting dalam

sistem pengendalian air dan tekanan pada alat. Oleh sebab itu jika

terjadi kerusakan pada solenoid valve maka sangat besar

pengaruhnya terhadap alat,

Bersihkan seluruh bagian pada solenoid valve dengan cara berikut

ini :

1. Buka body bagian panel : bagian depan, kanan, dan kiri.

2. Lepaskan selang-selang yang terhubung dengan solenoid valve

dengan cara tekan pangkal nepel dan tarik selang.

3. Solenoid valve menempel pada body dalam yang dikunci

dengan dua buah baut. Buka baut dari arah depan

menggunakan obeng panjang.

4. Tarik keluar solenoid valve ke daerah bebas dan lepaskan

bagian solenoid / kumparan dari valve bila perlu beri tanda

setiap bagian sehingga dapat mempermudah untuk

mengembalikannya seperti semula.

Instrumentasi Medis Sanata Dharma Halaman - 41 -


Smart Steam Autoclave SSA 11/12-006

5. Buka valve satu persatu dan bersihkan seluruh bagian

menggunakan kuas / sikat dan kain. Bersihkan hingga tidak

ada kerak yang menempel pada valve.

6. Setelah selesai kembalikan semuanya pada keadaan semula

kemudian jalankan test untuk menguji apakah solenoid sudah

bekerja normal atau tidak, yaitu dengan menjalankan secara

default sampai selesai, jika terjadi kesalahan pada solenoid

valve atau yang lain segera hentikan proses atau tekan tombol

emergency stop dan pastikan kembali urutan pemasangan

solenoid terhadap valve, ataupun selang terhadap valve tidak

ada yang terbalik.

d) Membersihkan Body Alat

Lakukan pembersihan terhadap body alat selama minimal satu

bulan sekali sehingga kebersihan dan kualitas body alat dapat tetap

terjaga.

Bersihkan body alat menggunakan silicon cair. Jangan

menggunakan air sabun, karena bagian dari body alat ini terbuat dari

bahan triplek berlapis melamin, akan dengan mudah rusak jika

kondisi body lembab terlalu lama, jaga kondisi body alat tetap

kering.

e) Mengganti Tubing untuk Sensor Tekanan

Ganti tubing untuk sensor secara berkala setiap 6 bulan sekali

(untuk kinerja pembacaan sensor tekanan yang optimal), untuk

Instrumentasi Medis Sanata Dharma Halaman - 42 -


Smart Steam Autoclave SSA 11/12-006

melepas tubing tekan bagian pangkal nepel dan tarik tubing dari

nepel.

Pilih spesifikasi tubing untuk sensor yang sesuai jangan

sembarangan menggunakan tubing walaupun ukurannya sama.

Karena kualitas dari bahan tubing akan berpengaruh terhadap umur

pakai, akurasi data sensor, dan ketahanan terhadap suhu dan tekanan,

Pakailah tubing pneumatic berbahan nylon dengan ukuran 4mm

dengan panjang yang disesuaikan kebutuhan pada alat.

f) Mengganti Sensor Suhu

Untuk pemeliharaan sensor suhu sebenarnya tidak diperlukan

pemeliharaan rutin, jika sensor suhu mengalami kerusakan maka

perlu dilakukan penggantian sensor.

Sensor suhu yang digunakan pada autoclave ini adalah sensor

Thermocouple baut pendek Tipe – K. Pada saat penggantian sangat

penting dilakukan yaitu mengisolasi bagian sensor terhadap body

chamber menggunakan seal ring ataupun sealer, sehingga dapat

dipastikan tidak terjadi kebocoran tekanan.

3.4. Tindakan Darurat dan Troubeshooting

a) Alat Tidak Mau Menyala

1. Jika alat tidak menyala pastikan tombol emergency stop dalam

keadaan tidak ditekan atau tombol seharusnya dalam keadaan

menonjol keluar. Jika tombol posisi ditekan maka putar tombol

Instrumentasi Medis Sanata Dharma Halaman - 43 -


Smart Steam Autoclave SSA 11/12-006

searah jarum jam / putar ke arah kanan sehingga tombol

menonjol keluar.

2. Jika alat masih tidak menyala periksa bagian sekring apakah

masih baik atau putus. Buka bagian sekring dengan memutar

knob berlawanan arah jarum jam atau ke arah kiri dan lihat

sekring. Jika kawat yang terdapat dalam tabung sekring putus

maka ganti dengan sekring yang serupa dengan spesifikasi yang

sama.

3. Jika masih tidak dapat menyala segera hubungi teknisi yang

bersangkutan.

b) Alat Tidak Mau Dijalankan

Periksa peringatan yang ada :

1. “Water Tank KOSONG”

jika tangki air kosong alat tidak akan dapat bekerja. Pengisian

air akan berlangsung otomatis jika saluran pengisian

dihubungkan dengan sumber air. Jika saluran pengisian tidak

terhubung maka pengisian air harus dilakukan secara manual

melalui bagian atas box alat pada bagian lubang pembuang uap

panas.

2. “MODE TIDAK TERSEDIA”

Alat tidak kompatibel dengan mode yang dipilih. Beberapa versi

masih dalam tahap pengembangan

Instrumentasi Medis Sanata Dharma Halaman - 44 -


Smart Steam Autoclave SSA 11/12-006

3. “PROGRAM PROTECTED”

Masa berlaku alat adalah 200 kali reset atau alat dinyalakan

sampai 200 kali, lebih dari itu alat tidak dapat digunakan, alat

membutuhkan maintenance dari teknisi. Hubungi teknisi terkait

untuk memperbaikinya.

c) Alat Error / Mengindikasikan Bahaya

Dalam pengoprasian alat sesuatunya tidak selalu dapat berjalan

dengan lancar, dalam situasi atau jangka waktu tertentu saat alat

digunakan alat mengalami gangguan teknis atau sesuatu yang tak

diharapkan seperti : tekanan dalam chamber terlalu tinggi, atau

tekanan bocor melalui selang, suhu tidak membaca normal sehingga

mengakibatkan heater menyala terus tak terkendali, air dalam

chamber tidak sesuai dengan batas minimalnya/ chamber terisi

penuh dengan air, dll.

Instrumentasi Medis Sanata Dharma Halaman - 45 -


Smart Steam Autoclave SSA 11/12-006

Tabel 2. Troubelshooting

Kerusakan Penyebab Tindak lanjut

 Alat tidak nyala  Tombol  Putar tombol


Emergency
tertekan
 Sumber listrik
belum
tersambung  Sambung ke
 Sekring putus jaringan listrik
 Ganti sekring
 Alat tidak mau  Water tank  Isi water tank
dijalankan kosong dengan aquadest
 Hubungi teknisi
untuk servis
 PROGRAM total
PROTECTED
 Sensor water  Tidak sesuai air  Pastikan pilihan
level eror yang digunakan air yang masuk
sesuai tombol
selektor bagian
belakang.
 Tekanan udara  Tutup chamber  Kencangkan
tidak tercapai kurang rapat baut tutup
 Safety valve  Tutup safety
terbuka valve
 Kebocoran seal
pada tutup
tabung  Ganti seal
 Kebocoran pada dengan yang
solenoid valve baru
 Bersihkan
solenoid valve
(hubungi
teknisi)
 Suhu tidak  Heater rusak  Hubungi teknisi
tercapai  SSR rusak
 Pengosongan  Saluran  Bersihkan
tidak maksimal pengosongan saluran
buntu pengosongan
 Valve macet  Cek solenoid
valve

Instrumentasi Medis Sanata Dharma Halaman - 46 -


Smart Steam Autoclave SSA 11/12-006

PENTING…!!!

Untuk tindakan darurat yang perlu dilakukan adalah

Tekan Tombol Emergency Stop,

sehingga alat akan dimatikan secara paksa, jika keadaan pintu masih terkunci

maka

Buka Katup Pada Safety Valve,

sehingga tekanan uap dalam chamber dapat keluar, `

Setelah Tekanan Pada Pressure Gauge Menunjuk Ke Angka Nol

Maka Pintu Sudah Dalam Keadaan Aman Untuk Dibuka.

Setelah semua tindakan pengamanan dilakukan maka yang perlu dilakukan

berikutnya adalah memanggil teknisi terkait.

Instrumentasi Medis Sanata Dharma Halaman - 47 -


Smart Steam Autoclave SSA 11/12-006

4. Pengoperasian Alat

4.1. Persiapan Alat dan Bahan

Untuk melakukan sterilisasi siapkan alat-alat berikut :

a) Objek yang akan disterilkan,

b) Indicator Tape (plester indikasi steril),

c) Autoclave SSA, (pastikan alat dalam kondisi siap digunakan).

Bungkus objek yang telah dicuci bersih dengan kain atau pembungkus

sehingga objek terbebas dari udara luar. Kemudian pasangkan plester

indicator tape pada bagian pembungkus objek (tidak perlu panjang,

cukup 2 atau 3 garis). Lalu masukan objek dalam chamber autoclave.

4.2. Aturan Seting Pengoprasian

Tabel 3. pengaturan seting batas waktu, tekanan, dan temperatur untuk

berbagai macam hal yang akan disterilkan

Waktu Tekanan Uap Suhu

HAL Sterilisasi (Pressure Gauge) Relatif

(Menit) Mpa Kpa Lbf/in2 (°C)

Karet 15 0.11 110 16 121 – 126

Kain 30 – 45 0.105 - 0.140 105 - 140 15 – 20 121 – 126

Perabot 15 0.105 - 0.140 105 - 140 15 – 20 121 – 126

Instrumen 10 0.105 - 0.140 105 - 140 15 – 20 121 – 126

Botol Cairan 20 – 40 0.105 - 0.140 105 - 140 15 – 20 121 – 126

Instrumentasi Medis Sanata Dharma Halaman - 48 -


Smart Steam Autoclave SSA 11/12-006

4.3. Cara Melakukan Proses Sterilisasi

Langkah yang perlu diperhatikan untuk melakukan proses sterilisasi :

1. Pastikan alat menyala. Lihat cara menyalakan alat pada bagian cara

instalasi.

2. Masukan objek ke dalam chamber.

3. Lihat pada tabel dan sesuaikan pengaturan pada alat.

4. Masukan pengaturan suhu, tekan tombol “UP” / ”DOWN” untuk

merubah nilai, setelah selesai tekan tombol geser “>”

5. Masukan pengaturan waktu, tekan tombol “UP” / ”DOWN” untuk

merubah nilai, setelah selesai tekan tombol geser “>”

6. Masukan pengaturan tekanan, tekan tombol “UP” / ”DOWN”

untuk merubah nilai, setelah selesai tekan tombol geser “>”

7. pilih pengaturan mode dengan menekan tombol “UP” / ”DOWN”

setelah selesai tekan tombol geser “>”

8. Setelah selesai pastikan lagi pengaturan yang dipilih sudah benar

dengan cara menekan tombol geser “>” dan periksa batas

pengaturan, jika sudah benar maka tekan tombol “OK”

9. Kemudian tutup chamber dan kencangkan baut pengunci. Selama

alat melakukan proses sterilisasi user dapat melihat total waktu

yang berjalan dengan menekan dan menahan tombol “OK” selama

kurang lebih 2 detik.

10. Tunggu hingga proses selesai dan tekan tombol “OK” untuk

memverifikasi bahwa proses telah selesai.

Instrumentasi Medis Sanata Dharma Halaman - 49 -


Smart Steam Autoclave SSA 11/12-006

4.4. Diagram Alir Proses Sterilisasi

Gambar 45. Diagram alir cara melakukan sterilisasi

Instrumentasi Medis Sanata Dharma Halaman - 50 -


Smart Steam Autoclave SSA 11/12-006

5. Spesifikasi Alat

Tabel 4. Spesifikasi alat

Nama Smart Steam Autoclave

Tipe SSA - 11/12-006

Power usage 220V AC 50 Hz

Current usage 7,60 A

Heater power 2000 Watt

Sekring 220V 8.0A

Chamber Volume Capacity 24 Liter / 0,024 m3

Water tank Capacity 6 Liter

Max. Noise 76,4 dBA

Temperature operating range 100 - 134 °C

Temperature ambient 25 - 28 °C

Max. Safe Pressure 165 Kpa

Temperature Sensor Thermocouple K-Type

Pressure Sensor MPX-2100DP

Serial number SSA - 001-2014

Contact Person :

+628976882570

Phone Number +628179442722

g.pasca_christ@yahoo.com

Email Address leander@pmsd.ac.id

Instrumentasi Medis Sanata Dharma Halaman - 51 -


Smart Steam Autoclave SSA 11/12-006

CATATAN :

Instrumentasi Medis Sanata Dharma Halaman - 52 -


Smart Steam Autoclave SSA 11/12-006

Instrumentasi Medis Sanata Dharma Halaman - 53 -


Smart Steam Autoclave SSA 11/12-006

Politeknik Mekatronika Sanata Dharma


Kampus paingan Maguwoharjo, sleman Yogyakarta
Telp. (0274) 883037 , 886530 ext. 2272 / 2263 | Fax. (0274) 886529
Email : admin@pmsd.ac.id | http://www.pmsd.ac.id

Instrumentasi Medis Sanata Dharma Halaman - 1 -