Budi Rochmanto
BT2MP-BPPT 1
Kalibrasi Parameter Aliran-Volume Untuk aplikasi
Peralatan Monitoring Udara Lingkungan
Materi :
2
Flowmeter pada Air Sampling
3
Flowmeter ( Air Sampling )
4
Flowmeter ( Air Sampling )
5
Skema Umum Kalibrasi Flowmeter Gas
Persamaan umum : →
6
Mengapa harus dikalibrasi ?
7
Hirarki Standard Kalibrasi
8
Metode Kalibrasi Flowmeter Gas
Kalibrasi debit meter gas umumnya dilakukan dengan
metode :
1. Gravimetric Methods (weighing methods)
2. Volumetric Methods
9
Metode Kalibrasi Flowmeter Gas
10
Hasil Kalibrasi Flowmeter Gas
2. Ketidakpastian Pengukuran
Besaran terukur (aliran atau volume)
Prosen ( % of read atau % FS )
11
Akurasi / Kemampuan Ukur Metode/ Flowmeter STD
I. Standar Utama :
- Gravimetric Methods (weighing methods)
- Piston Prover / Bell Prover (volumetric)
- Sonic Nozzle
14
Kalibrasi Low Flowmeter Gas / Udara
15
Skema Instalasi Kalibrasi Rotameter
16
Low Flowmeter Udara (aplikasi rotameter)
17
Beberapa Instalasi Kalibrasi Low Flowmeter Udara (LVS)
18
Prosedur Kalibrasi Rotameter Aliran Rendah ( hembus)
Pastikan :
Sumber udara/gas harus stabil
Gunakan regulator dengan kepresisian yang cukup
Bila menggunakan flowmeter seting, gunakan flowmeter dengan akurasi
yang cukup
Pastikan tidak terjadi kebocoran, khususnya di dearah kritis atau seksi
uji
Pastikan pemasangan sesuai dengan manual peralatan yang digunakan
untuk meter uji dan meter referensi
Patikan media kerja kalibrasi sama dengan media desain rotameter, jika
tidak lakukan perhitungan koreksi
Pengambilan data
Tentukan berapa titik ukur yang mengcover semua rentang kerja atau
sesuai kebutuhan
Lakukan pengambilan / pembacaan data pada saat kondisi aliran benar-
benar stabil
Lakukan pembacaan sesuai dengan petunjuk rotameter ( tengah / atas
floater)
Sebaiknya lakukan pengulangan data jika memungkinkan untuk
mendapatkan data yang representatif
Lakukan pencatatan : Qrota, Oact-master, P & T pada setiap pengukuran
19
Perhitungan Pada Rotameter / Variable Area Flowmeter
21
Contoh Perhitungan ketidakpastian Kalibrasi Rotameter vs Piston Prover
Set :
50
Cov.fact
Unit / Expanded
Uncertainty source / or / Deg.of Std.uncertai Sens.coeff /
No Satua Distribusi Type Symbol uncertanty ci.ui (ci.ui)2 (ci.ui)4/vi
Komponen Pembagi, freedom / vi nty / ui ci.
n / U
k
Repeatability of student't
1 measurements l/min (esdm) A urep 0,0059 Ö10 9 0,0019 1,00 0,0019 3,5E-06 1,3E-12
Piston prover
2 reference l/min normal B ustd1 0,0188 2 ~ 0,00942 1,00 0,0094 8,9E-05 0,0E+00
3 Piston Prover drift l/min rectangular B udrift std 0,00188 Ö3 ~ 0,0010873 1,00 0,00109 1,2E-06 0,0E+00
4 Actual pressure kPa normal B ustd2 0,021 2 ~ 0,0105 0,000189 0,0000020 3,9E-12 0,0E+00
5 Actual temperature oC normal B ustd3 0,04 2 ~ 0,02 -0,00011 -0,0000022 4,8E-12 0,0E+00
Division of meter
6 reading grade rectangular B ures 0,0049 Ö3 ~ 0,00285 0,99 0,0028 7,9E-06 0,0E+00
1,343E-
Sums 1,013E-04 12
Combined Uncertaity, uc 0,01006
Effective. Deg Of Freedom, Veff 7633,512
Cov. Factor for 95% CL 1,96
Expanded uncertainty, U95 0,020 l/min
Expanded uncertainty, U95 0,17 %
22
2. Kalibrasi High Volume Air Sampler (HVAS) - Orifice
I. Standar Utama :
- Roots Meter (volumetric)
- Piston Prover / Bell Prover (volumetric)
- Differential Pressure Device ( ΔP methods)
- Sonic Nozzle & Master Flowmeter (High Precision)
24
Skema & Hasil Kalibrasi Orifice vs Root’s meter
25
Sumber : Andersen Instr., Inc. Orifice Transfer Standard Certificate Worsheet
Contoh-1 Hasil Kalibrasi Orifice Transfer std. dengan Rootsmeter
26
Sumber : Andersen Instr., Inc. Orifice Transfer Standard Certificate Worsheet
Contoh Model Grafik Hasil Kalibrasi Orifice vs Aliran Referensi.
27
2.B Metode Kalibrasi Aliran Udara Pada HVAS
28
Prosedur khusus Kalibrasi HVAS
29
Pemasangan Orifice kalibrator dengan Penunjukan
Manometer pada HVAS
30
Pemasangan Kalibrator Orifice (manometer) pada HVAS
Aliran dari kalibrator/ref. diperoleh dari pembacaan sertifikat orifice
Manometer harus terpasang tegak lurus
Pembcaan manometer diperoleh dari jarak ketinggian dua sisi tubing
31
Pemasangan Digital Calibrator Pada HVAS
32
Contoh2.Data Hasil Kalibrasi HVAS - indikator Flow Record.
33
Sumber : ENSR Asia(HK) Ltd. TSP HVAS Field Cal.Report
Contoh-3 Data Hasil Kalibrasi HVAS Penunjukan Chart Record.
dengan formula Setting Flowrate
34
Sumber : PM 10 HVAS cal.date sheet AQMP yerington, Nevada
Contoh Grafik Hasil Kalibrasi HVAS vs Orifice Transfer Std.
35
Contoh Data dan Perhitungan Kalibrasi HVAS → Regresi
Site Conditions
Barometric Pressure (in Hg):29,50 Corrected Pressure (mm Hg):749
Temperature (deg F):68 Temperature (deg K):293
Average Press. (in Hg):29,65 Corrected Average (mm Hg):753
Average Temp. (deg F):70 Average Temp. (deg K):294
Calibration Orifice
Make: Qstd Slope:1,58304
Model: Qstd Intercept:-0,01520
Serial#: Date Certified:
Calibration Information
Plate or H2O Qstd I IC
Test # (in) (m3/min) (chart) (corrected) Linear Regression
1 7,25 1,713 52,0 52,07 Slope:28,7714
2 5,75 1,526 47,0 47,06 Intercept:2,9463
3 4,55 1,359 42,0 42,06 Corr. Coeff:0,9998
4 3,45 1,185 37,0 37,05
5 2,70 1,049 33,0 33,05 # of Observations:5
36
Contoh Data dan Perhitungan Kalibrasi HVAS → Koreksi
Site Conditions
Temp (deg F): 68,0
Ta (deg K): 293 Barometric Press (in Hg): 29,50
Ta (deg C): 20 Pa (mm Hg): 749
Calibration Orifice
Make: Qa Slope:0,99128
Model: Qa Intercept:-0,00941
Serial#: Calibration
Due Date:
Calibration Data
Run Orifice Qa Sampler Pf Look Up
Number "H2O (m3/min) "H2O (mm Hg) Po/Pa (m3/min) % Diff
1 3,80 1,239 6,40 11,944 0,984 1,288 3,95
2 3,80 1,239 6,80 12,691 0,983 1,287 3,87
3 3,80 1,239 7,20 13,437 0,982 1,286 3,79
4 3,75 1,231 9,25 17,263 0,977 1,278 3,82
5 3,75 1,231 10,20 19,036 0,975 1,275 3,57
37
Keterangan : Kalkulasi Kalibrasi HVAS → Regresi & Koreksi
Calculations (regresi)
Qstd = 1/m[Sqrt(H2O(Pa/Pstd)(Tstd/Ta))-b] m = sampler slope
IC = I[Sqrt(Pa/Pstd)(Tstd/Ta)] b = sampler intercept
I = chart response
Qstd = standard flow rate Tav = daily average temperature
Calculations ( koreksi)
kPa o
C m3/min m3/min % m3/min %
100,50 24,0 1,09 1,144 5,2 0,03 3,0
39
Ketidakpastian aliran pada kalibrasi HVAS
Perhitungan Ketidakpastian
40
Tabel Perhitungan Ketidakpastian Dasar kalibrasi HVAS vs Orifice Cal.
Set : 20 %
Cov.facto
Expanded
Uncertainty source / Unit / r/ Deg.of Std.uncertai Sens.coef
No Distribusi Type Symbol uncertanty ci.ui (ci.ui)2 (ci.ui)4/vi
Komponen Satuan Pembagi, freedom / vi nty / ui f / ci.
/ U
k
Repeatability of student't urep
1 m3/min A 0,025 Ö5 4 0,0113 1,00 0,0113 1,3E-04 4,1E-09
measurements (esdm)
2 Calibrator flowrate (act) m3/min normal B uflow 0,019 2 ~ 0,0093 0,994 0,0092 8,5E-05 0,0E+00
3 Pressure Calibrator kpa normal B uP1 0,0100 2 ~ 0,0050 0,011 0,0001 3,2E-09 0,0E+00
4 Presuure HVAS kPa normal B uP2 0,0100 2 ~ 0,0050 0,011 0,00005 2,9E-09 0,0E+00
5 Temperature Calibrator o
C normal B uT1 0,050 2 ~ 0,0250 -0,0038 -0,00010 9,2E-09 0,0E+00
6 Temperature HVAS o
C normal B uT2 0,050 2 ~ 0,0250 -0,0037 -0,00009 8,4E-09 0,0E+00
Division of meter ures
9
reading
m3/min rectangular B 0,0050 Ö3 ~ 0,00289 1,00 0,00289 8,3E-06 0,0E+00
44
Kalibrasi Gas Meter dengan WGM standard
45
(Q ,P ,T ) ( Q ,P ,T )
B. Kalibrasi Gas Meter dengan Flowmeter standard
Skema Kalibrasi Gas Meter Dengan Standar Flowmeter
(JIS B 7556 : 2008)
Pressure System
Suction System
Kalibrasi Gas Meter
48
Ketidakpastian Kalibrasi Gas Meter dengan WGM standard ( Q ref )
Set : 25 % ; 2,5 l/min
Cov.facto
Expanded
Uncertainty source / Unit / r/ Deg.of Std.uncertai Sens.coef
No Distribusi Type Symbol uncertanty ci.ui (ci.ui)2 (ci.ui)4/vi
Komponen Satuan Pembagi, freedom / vi nty / ui f / ci.
/ U
k
Repeatability of student't urep
1 l A 0,033 Ö5 4 0,0148 1,00 0,0148 2,2E-04 1,2E-08
measurements (esdm)
2 Volume WGM reference l normal B uvol-wgm 0,106 2 ~ 0,0529 0,94 0,0499 2,5E-03 0,0E+00
3 Timer WGM reference min normal B ut1 0,00417 2 ~ 0,00208 2,40 0,0050 2,5E-05 0,0E+00
4 Timer UUT min normal B ut2 0,00450 2 ~ 0,00225 2,55 0,00573 3,3E-05 0,0E+00
Pressure WGM uP1
5 kPa normal B 0,010 2 ~ 0,0050 0,09 0,00046 2,1E-07 0,0E+00
reference
6 Pressure UUT kPa normal B uP2 0,010 2 ~ 0,0050 0,09 0,00046 2,1E-07 0,0E+00
Temperature WGM uT1
7 o
C normal B 0,050 2 ~ 0,025 0,03 0,00078 6,1E-07 0,0E+00
reference
8 Temperature UUT o
C normal B uT2 0,050 2 ~ 0,025 0,03 0,00078 6,1E-07 0,0E+00
Division of meter ures
9 l rectangular B 0,0025 Ö3 ~ 0,00144 1,00 0,00144 2,1E-06 0,0E+00
reading
Temperature Sensor
50
Kalibrasi Isokinetik
51
KRITERIA ALAT UKUR TERSEBUT MASIH DAPAT DIGUNAKAN
TOLERANSI (+)
U
U
NILAI
U
U
TOLERANSI (-)
TITIK UKUR
52
53