Anda di halaman 1dari 16

SALINAN HASIL KALIBRASI ALAT STANDAR dan MEDIA KALIBRASI TEMPERATUR

Standar : Termometer Digital + PT 100


Merk : Graphtec
Tipe : GL 840
No. Seri : 05660
Resolusi 0.01
Ketertelusuran : LK-133-IDN

DATA SERTIFIKAT
Set Poin Standar Value Uut Value Error U95
y x
-30 -29.936 -31.3 1.364 0.11
0 -0.038 -0.79 0.752 0.11
150 150.014 152.15 -2.136 0.3
300 299.963 305.02 -5.057 0.3
450 449.935 458.05 -8.115 0.3
600 600 611.3 -11.3 0.3

SET POINT UKUR


Rage -20~100
Set Poin Standar Value Error U95
0 0.80 0.80 0.11
37 37.07 0.07 0.11
10 10.60 0.60 0.11
55 54.72 -0.28 0.3
100 98.84 -1.16 0.3

Media temperatur : Metrology Well Dry Block


Merk : Fluke
Tipe : 9171
No. Seri : B86948
Resolusi 0.001
Ketertelusuran : NVLAP

Data sertifikat media


Set Actual axial differential U95 stabilitas
-30 0.003 0.02 0.0019
155 -0.001 0.025 0.0018
Data sertifikat media interpolasi
Set Actual axial differential U95 stabilitas
0 0.002 0.021 0.0019
37 0.002 0.022 0.0019
10 0.002 0.021 0.0019
10 0.002 0.021 0.0019
100 0.000 0.024 0.0018
SUMMARY OUTPUT
mi ( g ) = a + bxi
a 0.798103655 Regression Statistics
b 0.980406829 Multiple R 1
R Square 1
Adjusted R Square 1
Standard Error 0.101632
Observations 6

ANOVA
df SS MS F
Regression 1 315700 315700 30564300
Residual 4 0.041316 0.010329
Total 5 315700.1

Coefficients
Standard Error t Stat P-value
Intercept 0.798104 0.060601 13.16985 0.000192
X Variable 1 0.980407 0.000177 5528.499 6.423E-15
Significance F
6.423E-15

Lower 95%Upper 95%Lower 95,0%


Upper 95,0%
0.629849 0.966359 0.629849 0.966359
0.979914 0.980899 0.979914 0.980899
Lab Kalibrasi UPT BPI - LIPI

DATA KALIBRASI
Dept :
Nama Alat : Thermometer Glass 2 Standar : Termometer Digital + PT 100 Media temperatur
Merk Isolab Merk : Graphtec Merk
Tipe : Tipe : GL 840 Tipe
No. Seri :1 No. Seri : 05660 No. Seri
Rentang Ukur : 25 °C Resolusi 0.01 °C Resolusi
Resolusi 1 °C Ketertelusuran : LK-133-IDN Ketertelusuran

Kondisi Ruang Tanggal Suhu (oC) RH(%)


Sebelum Kalibrasi
Selama Kalibrasi 12-Jun-20
Sesudah Kalibrasi

Penunjukan Penunjukan T Penunjukan


Set Point Standar (TS) Set Point Standar (TS)
(Ttest)
0 -0.72 °C 0.00 °C 55 55.5
-0.72 °C 0.00 °C 55.45
-0.72 °C 0.00 °C 55.45
-0.72 °C 0.00 °C 55.5
-0.72 °C 0.00 °C 55.5
Rata-rata -0.72 °C 0 °C Rata-rata 55.48
Std dev. (σ) 0.000 °C 0.000 °C Std dev. (σ) 0.027
koreksi bulb koreksi bulb
koreksi Standar 0.80 °C koreksi Standar -0.28
Temperatur Aktual 0.08 °C 0 °C Temperatur Aktual 55.20
koreksi, C = ( TS - Ttest) 0.08 °C koreksi, C = ( TS - Ttest) -1.80
37 36.78 °C 37 °C 100 100.65
36.79 °C 37 °C 100.75
36.75 °C 37 °C 100.85
36.74 °C 37 °C 100.85
36.68 °C 37 °C 100.85
Rata-rata 36.748 °C 37.00 °C Rata-rata 100.79
Std dev. (σ) 0.043 °C 0.000 °C Std dev. (σ) 0.089
koreksi bulb koreksi bulb
koreksi Standar 0.07 °C koreksi Standar -1.16
Temperatur Aktual 36.82 °C 37.00 °C Temperatur Aktual 99.63
koreksi, C = ( TS - Ttest) -0.18 °C koreksi, C = ( TS - Ttest) -0.16
10 9.59 °C 10.25 °C
9.59 °C 10.25 °C
9.59 °C 10.25 °C
9.57 °C 10.25 °C
9.59 °C 10.25 °C
Rata-rata 9.586 °C 10.25 °C Rata-rata
Std dev. (σ) 0.009 °C 0.000 °C Std dev. (σ)
koreksi bulb °C °C koreksi bulb
koreksi Standar 0.60 °C °C koreksi Standar
Temperatur Aktual 10.19 °C 10.25 °C Temperatur Aktual
koreksi, C = ( TS - Ttest) -0.06 °C koreksi, C = ( TS - Ttest)
Lab Kalibrasi UPT BPI - LIPI

a temperatur : Metrology Well Dry Block


: Fluke
: 9171
: B86948
0.001 °C
elusuran : NVLAP

enunjukan Penunjukan T
andar (TS) (Ttest)
°C 57 °C
°C 57 °C
°C 57 °C
°C 57 °C
°C 57 °C
°C 57 °C
°C 0.000 °C

°C
°C 57 °C
°C
°C 99.8 °C
°C 99.8 °C
°C 99.8 °C
°C 99.8 °C
°C 99.75 °C
°C 99.79 °C
°C 0.022 °C

°C 99.79 °C
°C
°C °C
UNCERTAINTY BUDGET TERMOMETER DIGITAL

Nama Alat UUT : : Thermometer Glass 2 Alat Standar : Termometer Digital + PT 100
Range : -20~100 Merk : Graphtec
Merk : Isolab Tipe/No. Seri : GL 840 : 05660
Tipe/No. Seri : : :1 Resolusi 0.01 °C
Resolusi 0
: C 1 °C Ketertelusuran : LK-133-IDN

Test Point : 0 °C Media : Metrology Well Dry Block

Komponen Satuan Distribusi U Pembagi νi ui ci ui ci (ui ci)² (ui ci)4/vi

1. Repeatability alat °C T'Student 0 2.23606798 4 0 1 0 0 0


2. Repeatability standar °C T'Student 0 2.23606798 4 0 1 0 0 0
3. Resolusi alat °C Rectangular 0.5 1.73205081 200 0.2886751 1 0.28867513 0.083333333 3.47222E-05
4. Sertifikat standar °C Normal 0.11 2 60 0.055 1 0.055 0.003025 1.5251E-07
5. Resolusi standar °C Rectangular 0.005 1.73205081 200 0.0028868 1 0.00288675 8.33333E-06 3.47222E-13
6. Ketidakstabilan suhu media °C Rectangular 0.00 1.73205081 200 0.0010876 1 0.0010876 1.18288E-06 6.99603E-15
7. Homogenitas suhu media °C Rectangular 0.0023514 1.73205081 200 0.0013576 1 0.00135755 1.84295E-06 1.69823E-14
Jumlah 0.086369692 3.48747E-05
Ketidakpastian baku gabungan, uc uc = √ [Σ(ui ci)²] 0.293887210504598
Derajat kebebasan efektif, νeff νeff = uc4 / [Σ(ui ci) 4/νi] 213.900527355176

6
Faktor cakupan, k-student's untuk derajat kebebasan efektif, ν eff , pada CL = 95% k(νeff) 1.98
Ketidakpastian bentangan, U = k . u c UE = k . u c 0.581896676799104

test point 37 °C Media : Metrology Well Dry Block


Komponen Satuan Distribusi U Pembagi νi ui ci ui ci (ui ci)² (ui ci)4/vi
1. Repeatability alat °C T'Student 0 2.23606798 4 0 1 0 0 0
2. Repeatability standar °C T'Student 0.0432435 2.23606798 4 0.0193391 1 0.01933908 0.000374 3.4969E-08
3. Resolusi alat °C Rectangular 0.5 1.73205081 200 0.2886751 1 0.28867513 0.083333333 3.47222E-05
4. Sertifikat Standar °C Normal 0.11 2 60 0.055 1 0.055 0.003025 1.5251E-07
5. Resolusi standar °C Rectangular 0.005 1.73205081 200 0.0028868 1 0.00288675 8.33333E-06 3.47222E-13
6. Ketidakstabilan suhu media °C Rectangular 0.00 1.73205081 200 0.0010761 1 0.00107606 1.1579E-06 6.70362E-15
7. Homogenitas suhu media °C Rectangular 0.0015514 1.73205081 200 0.0008957 1 0.00089567 8.0223E-07 3.21787E-15
Jumlah 0.086742627 3.49097E-05
Ketidakpastian baku gabungan, uc uc = √ [Σ(ui ci)²] 0.294521012482419
Derajat kebebasan efektif, νeff νeff = uc4 / [Σ(ui ci) 4/νi] 215.535592481452
Faktor cakupan, k-student's untuk derajat kebebasan efektif, ν eff , pada CL = 95% k(νeff) 1.97105912213365
Ketidakpastian bentangan, U = k . u c UE = k . u c 0.580518328

test point 10 °C Media : Metrology Well Dry Block


Komponen Satuan Distribusi U Pembagi νi ui ci ui ci (ui ci)² (ui ci)4/vi
1. Repeatability alat °C T'Student 0 2.23606798 4 0 1 0 0 0
2. Repeatability standar °C T'Student 0.0089443 2.23606798 4 0.004 1 0.004 1.6E-05 6.4E-11

7
3. Resolusi alat °C Rectangular 0.5 1.73205081 200 0.2886751 1 0.28867513 0.083333333 3.47222E-05
4. Sertifikat Standar °C Normal 0.11 2 60 0.055 1 0.055 0.003025 1.5251E-07
5. Resolusi standar °C Rectangular 0.005 1.73205081 200 0.0028868 1 0.00288675 8.33333E-06 3.47222E-13
6. Ketidakstabilan suhu media °C Rectangular 0.00 1.73205081 200 0.0010845 1 0.00108448 1.1761E-06 6.91608E-15
7. Homogenitas suhu media °C Rectangular 0.0021351 1.73205081 200 0.0012327 1 0.00123272 1.5196E-06 1.15459E-14
Jumlah 0.086385362 3.48748E-05
Ketidakpastian baku gabungan, uc uc = √ [Σ(ui ci)²] 0.293913868963555
Derajat kebebasan efektif, νeff νeff = uc4 / [Σ(ui ci) 4/νi] 213.977756792868
Faktor cakupan, k-student's untuk derajat kebebasan efektif, ν eff , pada CL = 95% k(νeff) 1.97116388546526
Ketidakpastian bentangan, U = k . u c UE = k . u c 0.579352404

test point 55 °C Media : Metrology Well Dry Block


Komponen Satuan Distribusi U Pembagi νi ui ci ui ci (ui ci)² (ui ci)4/vi
1. Repeatability alat °C T'Student 0 2.23606798 4 0 1 0 0 0
2. Repeatability standar °C T'Student 0.0273861 2.23606798 4 0.0122474 1 0.01224745 0.00015 5.625E-09
3. Resolusi alat °C Rectangular 0.5 1.73205081 200 0.2886751 1 0.28867513 0.083333333 3.47222E-05
4. Sertifikat Standar °C Normal 0.3 2 60 0.15 1 0.15 0.0225 8.4375E-06
5. Resolusi standar °C Rectangular 0.005 1.73205081 200 0.0028868 1 0.00288675 8.33333E-06 3.47222E-13
6. Ketidakstabilan suhu media °C Rectangular 0.00 1.73205081 200 0.0010845 1 0.00108448 1.1761E-06 6.91608E-15
7. Homogenitas suhu media °C Rectangular 0.0021351 1.73205081 200 0.0012327 1 0.00123272 1.5196E-06 1.15459E-14
Jumlah 0.105994362 4.31653E-05
Ketidakpastian baku gabungan, uc uc = √ [Σ(ui ci)²] 0.325567753884082
4 4
Derajat kebebasan efektif, νeff νeff = uc / [Σ(ui ci) /νi] 260.273702908515

8
Faktor cakupan, k-student's untuk derajat kebebasan efektif, ν eff , pada CL = 95% k(νeff) 1.96913000336018
Ketidakpastian bentangan, U = k . u c UE = k . u c 0.641085232

test point 100 °C Media : Metrology Well Dry Block


Komponen Satuan Distribusi U Pembagi νi ui ci ui ci (ui ci)² (ui ci)4/vi
1. Repeatability alat °C T'Student 0.0223607 2.23606798 4 0.01 1 0.01 1E-04 2.5E-09
2. Repeatability standar °C T'Student 0.0894427 2.23606798 4 0.04 1 0.04 0.0016 6.4E-07
3. Resolusi alat °C Rectangular 0.5 1.73205081 200 0.2886751 1 0.28867513 0.083333333 3.47222E-05
4. Sertifikat Standar °C Normal 0.3 2 60 0.15 1 0.15 0.0225 8.4375E-06
5. Resolusi standar °C Rectangular 0.005 1.73205081 200 0.0028868 1 0.00288675 8.33333E-06 3.47222E-13
6. Ketidakstabilan suhu media °C Rectangular 0.00 1.73205081 200 0.0010564 1 0.00105639 1.11597E-06 6.22695E-15
7. Homogenitas suhu media °C Rectangular 0.0001892 1.73205081 200 0.0001092 1 0.00010923 1.19308E-08 7.11726E-19
Jumlah 0.107542795 4.38022E-05
Ketidakpastian baku gabungan, uc uc = √ [Σ(ui ci)²] 0.327937180825552
Derajat kebebasan efektif, νeff νeff = uc4 / [Σ(ui ci) 4/νi] 264.038032396052
Faktor cakupan, k-student's untuk derajat kebebasan efektif, ν eff , pada CL = 95% k(νeff) 1.9689904969297
Ketidakpastian bentangan, U = k . u c UE = k . u c 0.645705193

9
Digital + PT 100

untuk sertifikat standard derajat effectivitas diambil dari table t-student, karena k=2 maka v = 60

10
Lab Kalibrasi UPT BPI - LIPI

LAPORAN KALIBRASI TERMOMETER DIGITAL

No Order
Nama Alat : Thermometer Glass 2 Tempat kalibrasi :
Merk Isolab Tanggal kalibrasi : 6/12/2020
Tipe :
No. Seri :1
Rentang Ukur : 25 °C Suhu : 0
Resolusi 1 0C Kelembaban : 0
Pemilik :

HASIL KALIBRASI

Set point Temperatur


koreksi Ketidakpastian
Aktual Alat

0 °C 0.00 °C 0.08 °C ± 0.582 °C


37 °C 37.00 °C -0.18 °C ± 0.581 °C
10 °C 10.25 °C -0.06 °C ± 0.579 °C
55 °C 57.00 °C -1.80 °C ± 0.641 °C
100 °C 99.79 °C -0.16 °C ± 0.646 °C

Catatan :
- Alat dikalibrasi menggunakan standar : Termometer Digital + PT 100 merk : Graphtec yang tertelusur
ke satuan SI melalui : LK-133-IDN
- Prosedur kalibrasi yang digunakan IK01/SH-02 mengacu pada ASTM E-77/1998, ASTM E-644/1998
- Perhitungan ketidakpastian hasil kalibrasi mengacu ke ISO Guide to the Expression of Uncertainty
in Measurement 1993
- Ketidakpastian kalibrasi dinyatakan pada tingkat kepercayaan 95% dengan faktor cakupan k = 2
Lab Kalibrasi UPT BPI - LIPI

0
C
%

yang tertelusur
PERHITUNGAN KETIDAKPA

Sumber Ketidakpastian
Standar Tipe A Tipe B
Set Alat Standar Koreksi Sertifikat ResolusiRef ResolusiUUT
+ Koreksi STDV alat STDV std aRef aUUT
Standar
UNGAN KETIDAKPASTIAN

astian Perhitungan Ketidakpastian


Tipe B Tipe A Tipe B
Hmgnbath Instability UB1 UB2 UB3 UB4 UB5 Uc
UA Alat UA std
Hbath Ibath Uref/k (aRef/2)/√3 (aUUT/2)/√3
k U95
Thermometer Glass 4
Thermometer Glass 1
Thermometer Glass 13
Thermometer Glass 1 1.19
Thermometer Glass 12

Anda mungkin juga menyukai