7.2-FR-GQA-WQ-085
Thalium, Tl
APHA 3113B, Sampling :SNI 7117.20:2009
Tanggal Analisa : 26-Feb-24
Analis : Yanu
Matriks : Udara Emisi
KURVA KALIBRASI
Identitas Standar Konsentrasi Absorbansi
Blanko 0.0 0.0007 Kurva Kalibrasi
Standar 1 1.0 0.0067
0.06
Standar 2 5.0 0.0169
Standar 3 10.0 0.0278 y = 0.0019x + 0.0051
0.05
Standar 4 15.0 0.0331
R² = 0.9717
Standar 5 20.0 0.0446 0.04
Absorbansi
Standar 6 25.0 0.0485
Standar 7
0.03
Standar 8 0.02
Standar 9
Intercept 0.0051 0.01
Slope 0.002
Regresi R2 0.9717 0.00
Sdxy 0.00 0.00 5.00 10.00 15.00 20.00 25.00 30.00
Rerata Konsentrasi 10.9 Konsentrasi
Σ(xi-x̄)2 550.9
HASIL ANALISIS
Kadar Faktor Kadar EKP U95% Kadar Penimbangan Kadar
Nomor Contoh Uji Absorbansi
(ug/L) Pengenceran (ug/L) ± ug/L (mg/L) (Gram) (mg/Kg)
Blanko Peralatan 0.00000 -2.70 1 0.00 - 0.000 0.0 #DIV/0!
Blanko Lab 0.00000 -2.70 1 0.00 - 0.000 0.0 #DIV/0!
Blanko 0.00000 -2.70 1 0.00 - 0.000 0.0 #DIV/0!
1824099-1 0.00035 -2.51 1 -2.51 -3.92 -0.003 0.0 #DIV/0!
𝑚𝑔 𝐴𝑏𝑠 − 𝐼𝑛𝑡
𝐶𝑜𝑛𝑐( )= × 𝑓𝑝
𝐿 𝑆𝑙𝑜𝑝𝑒
𝑚𝑔 𝑐𝑜𝑛𝑐 𝑚𝑔/𝐿 × 𝑣𝑜𝑙𝑢𝑚𝑒 𝑠𝑎𝑚𝑝𝑒𝑙 (𝑚𝐿)
𝑐𝑜𝑛𝑐 ( ) =
𝐾𝑔 𝐵𝑒𝑟𝑎𝑡 𝑆𝑎𝑚𝑝𝑒𝑙 (𝑔)
PT GLOBAL QUALITY ANALITICAL Diverifikasi Oleh :
7.2-FR-GQA-WQ-085
Thalium, Tl
APHA 3113B, Sampling :SNI 7117.20:2009
Tanggal Analisa : 26-Feb-24
Analis : Yanu
Matriks : Udara Emisi
𝑆𝑖𝑚𝑝𝑙𝑜 − 𝐷𝑢𝑝𝑙𝑜
%𝑅𝑃𝐷 = 100%
𝑥lj 𝑆𝑖𝑚𝑝𝑙𝑜 𝐷𝑢𝑝𝑙𝑜
ESTIMASI KETIDAKPASTIAN
Ketidakpastian
SD Validasi %Rvalidasi
Akurasi
3% 2.7 95
Ketidakpastian
Presisi
5%
KONTRIBUSI KETIDAKPASTIAN
Sumber K. Sensitivitas
Linieritas 70% K. Sensitivitas
Presisi 5%
Akurasi 3%
Akurasi 3%
Presisi 5%
Linieritas 70%