Sertifikat Kalibrasi
Calibration Certificate
Certificate number: S.21-0066
S.21-0000.R1
Order number: I-21-00-000
E-21-02-001
Deskripsi Objek yang Dikalibrasi/Diukur
Description of object being calibrated or measured
Identitas Pemilik
Owner's Identification
Pengesahan
Authorization
Selisih Titik Tengah Pasangan Nomor 6/ Difference of Central Length of Pair Number 6
Nomor Panjang Nilai Rata-Rata Pengukuran/ Selisih Rata-Rata/ Standar Deviasi
Pasangan/ Nominal/ Measured Mean Value Difference Pengukuran/
Number of Nominal (µm) of mean Measured Standard
Pair Length (µm) Deviation
(mm) (µm)
Balok Ukur B Sisi Balok Ukur B Sisi Posisi d – Posisi u/ Posisi d/ Posisi u/
Datar di Bagian Datar di Bagian Position d –Position u Position d Position u
Bawah (Posisi d)/ Atas (Posisi u)/
A B Gauge block B Gauge block B
flat side down flat side up
(position d) (Position u)
(B – A) (B – A) (B – A) (B – A)
6 6 6 -0,16 - - 0,24 -
No. sertifikat / Cert. Number: S.21-0066
No. Order / Order Number: E-21-02-001
Halaman 3 dari 3 halaman
Page 3 of 3 pages
Penyimpangan fo dan fu dari Titik Tengah/ Deviations fo and fu from the Central Length
Panjang Nilai dari Sertifikat/ Nilai Rata- Rata Selisih/ Standar Deviasi
Nominal/ Value from Certificate Pengukuran/ Difference Pengukuran/
Nominal C Measured Mean Value C–M Measured Standard
Length (µm) M (µm) Deviation
(mm) (µm) (µm)
fo fu fo fu fo fu
Catatan/ Notes :
Untuk pasangan 1 mm dengan 1,01 mm dan gauge block bridge tidak terbaca oleh gauge block comparator/
Combination of pair 1 mm with 1.01 mm and gauge block bridge will not be measured by the gauge block comparator.
Nilai pengulangan untuk masing-masing pengukuran tidak stabil/ The indicated repeatability value for each
measurement unstable.
Standar Kalibrasi/ Reference Standard : Gauge Block Grade K, MAHR, S.N. 07012
Prosedur Kalibrasi/ Calibration Procedure : I.ML.2.09
Hasil pengukuran yang dilaporkan tertelusur ke SI melalui Tubitak UME, Turki/ The reported measurement result
is traceable to the SI through Tubitak UME, Turkey.
Ketidakpastian pengukuran dinyatakan pada tingkat kepercayaan tidak kurang dari 95 % dengan faktor
cakupan k = 2 adalah 1,27 µm/ Uncertainty of measurement is expressed at a confidence level of not less than 95 % with
coverage factor = 2 is 1.27 µm.