Memastikan limbah klinis dalam sirkuit ekstra corporeal dibuang dengan aman dan
Pengertian sesuai dengan sistem mesin 4008
Understanding Ensuring clinical waste in the extra corporeal circuit is disposed of safely and in
accordance with the 4008 engine system
Kebijakan
SK…………………….
Policy
A. Peralatan
Equipment
1. 4008 Dialysis Machine
2. Dialysis Fluid port cap
3. Recirculatory
4. Clamp
5. APD
Prosedur
6. Tempat sampah klinis
Procedure
Clinical Waste bin
B. Pelaksanaan
Impelementation
1. Kumpulkan peralatan yang dibutuhkan
2. Simpan salah satu dari Dialysis Fluid Port Caps (dengan penutup plastik)
Keep one of the Dialysis Fluid Port Caps (with plastic cover)
ID-MA-OP-007
PENGOSONGAN SIRKUIT EKSTRA CORPOREAL 4008
SETELAH PELEPASAN PASIEN
Emptying 4008 Extra Corporeal Circuit Following Patient
Disconnection
No. Dokumen No. Revisi Halaman
Document Number Revision Number Page
00 2/6
ID-MA-OP-007
PENGOSONGAN SIRKUIT EKSTRA CORPOREAL 4008
SETELAH PELEPASAN PASIEN
Emptying 4008 Extra Corporeal Circuit Following Patient
Disconnection
No. Dokumen No. Revisi Halaman
Document Number Revision Number Page
00 3/6
ID-MA-OP-007
PENGOSONGAN SIRKUIT EKSTRA CORPOREAL 4008
SETELAH PELEPASAN PASIEN
Emptying 4008 Extra Corporeal Circuit Following Patient
Disconnection
No. Dokumen No. Revisi Halaman
Document Number Revision Number Page
00 4/6
8. Pastikan pembuluh darah vena dan arteri kanula / CVC di klem sebelum
dilepaskan
Ensure venous and arterial blood lines cannulas/cvc clamps are clamped
before disconnection
13. Buka tutup interlock shunt dan kembalikan inlet fluid coupling kembali ke posisi
semula
ID-MA-OP-007
PENGOSONGAN SIRKUIT EKSTRA CORPOREAL 4008
SETELAH PELEPASAN PASIEN
Emptying 4008 Extra Corporeal Circuit Following Patient
Disconnection
No. Dokumen No. Revisi Halaman
Document Number Revision Number Page
00 5/6
Open the shunt interlock flap and return the dialysis fluid inlet coupling back to
position
ID-MA-OP-007
PENGOSONGAN SIRKUIT EKSTRA CORPOREAL 4008
SETELAH PELEPASAN PASIEN
Emptying 4008 Extra Corporeal Circuit Following Patient
Disconnection
No. Dokumen No. Revisi Halaman
Document Number Revision Number Page
00 6/6
ID-MA-OP-007