6.60 6.0
6.40 6.0
6.30 6.0
6.10 7.0
5.90 6.0
6.04 0.41 6.1 0.3
5.70 6.0
5.60 6.0
5.60 6.0
5.60 6.0
6.60 6.0
Certificate No. :
Order No. :
Place of Calibration :
Setting Penunjukan Penunjukan Fluktuasi
Alat Alat Standar Maksimum
%RH %RH %RH %RH
20 20 20.1 0.5
30 30 30.2 0.7
40 40 40.2 0.7
50 50 50.2 0.8
60 60 60.2 0.7
70 70 70.2 0.8
80 80 80.2 0.8
Technical Manager
Hasil Kalibrasi :
Calibration result :
Penunjukan Koreksi
Penunjukan Ketidakpastian
alat (ºC) (ºC)
standar (ºC) Uncertainty
Equipment Correction
Standard reads (ºC) (k=2)
reads (ºC) (ºC)
-24.3 -23.81 0.49 0.43
n 1
U₁ = 0 / √3 = 0.000
U₂ = (½ x 0,1) / √3 = 0.029
Ue = k x uc = 2 x 0,214 = 0.428
Page 3
Master File (pembulatan 3 angka)
QA Supervisor QA Manager
Page 4
Master File (pembulatan 2 angka)
Hasil Kalibrasi :
Calibration result :
Penunjukan Koreksi
Penunjukan Ketidakpastian
alat (ºC) (ºC)
standar (ºC) Uncertainty
Equipment Correction
Standard reads (ºC) (k=2)
reads (ºC) (ºC)
-24.3 -23.81 0.49 1.23
n 1
U₁ = 1 / √3 = 0.58
U₂ = (½ x 0,1) / √3 = 0.03
Ue = k x uc = 2 x 0,62 = 1.23
Page 5
Master File (pembulatan 2 angka)
QA Supervisor QA Manager
Page 6