OVERVIEW
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 2/26
SOPs
Microlab 300 & 300LX
DAFTAR ISI
1. Operasional Rutin
A. Menghidupkan Alat 4
B. Pengukuran – Metode Endpoint dan Twopoint 5
C. Pengukuran – Metode Kinetik 7
D. Mengisi Nilai Standar 8
E. Mematikan Alat 9
4. Perawatan Alat
A. Pergantian Test 13
B. Perawatan Setengah hari 13
C. Perawatan Harian 13
5. Perbaikan Alat
A. Penggantian dan Adjust Lampu Halogen 14
B. Penggantian Kertas Printer( Microlab 300) 17
C. Penggantian Pita Printer (Microlab 300) 17
8. Maintenance Tahunan 24
9. Kalibrasi Alat 25
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 3/26
SOPs
Microlab 300 & 300LX
1. Operasional Rutin
A. Menghidupkan Alat
1. Tekan tombol on/off pada stabilizer
2. Tekan tombol on/off pada UPS
3. Tekan tombol on/off alat Microlab 300/300LX yang terletak
pada body bagian belakang alat
4. Apabila muncul “tempatkan detergen 5% dan tekan sipper /
Place 5% detergent and press sipper”, siapkan larutan Labpro
Neutral 5 %, lalu letakkan dibawah sipper tube kemudian
hisapkan larutan tersebut dengan cara menekan sipper plat
dan tunggu sampai proses selesai (sampai dengan muncul
permintaan untuk memasukan aquabides)
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 4/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 5/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 6/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 7/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 8/26
SOPs
Microlab 300 & 300LX
E. Mematikan Alat
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 9/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 10/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 11/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 12/26
SOPs
Microlab 300 & 300LX
4. Perawatan Alat
A. Pergantian Tes
1. Letakkan larutan LabPro Neutral 5% di sipper tube
2. Tekan “sipper plat”
3. Biarkan proses “flushing” selama ± 2 menit
4. Letakkan aquabides dibawah sipper tube
5. Tekan “sipper plat”
6. Biarkan proses “flushing” selama ± 1 menit
C. Perawatan Harian
1. Letakkan LabPro Neutral 5% di sipper tube
2. Tekan “sipper plat”
3. Biarkan proses “flushing” selama ± 10 menit
4. Letakkan Aquabidest dibawah sipper tube
5. Tekan “sipper plat”
6. Biarkan proses “flushing” selama ± 2 menit
7. Matikan instrument, biarkan aquadest ada di dalam flowcell
CATATAN : Jangan pernah meninggalkan Sample atau campuran reagen
di dalam flowcell untuk jangka waktu yang lama.
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 13/26
SOPs
Microlab 300 & 300LX
5. Perbaikan Alat
A. Penggantian dan Adjust Lampu Halogen
Jika lampu mati atau rusak serta pembacaan absorbance tidak
normal maka perlu mengganti atau adjust lampu. Anda
membutuhkan obeng plus (+), tissue atau kain yang bebas lemak,
dan pin (bisa berupa obeng yang sangat kecil).
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 14/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 15/26
SOPs
Microlab 300 & 300LX
Lampu halogen 12V-20W dari International Light Technologies (eks Gilway) pemakaian
max 2000 jam (lampu pabrikan lain lama pemakaian ± 200 jam)
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 16/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 17/26
SOPs
Microlab 300 & 300LX
Metode IFCC
Immuno Modified Hexo-
IFCC Jaffe GOD-PAP
IFCC kinase
inhibition
Mode Kinetic Kinetic Kinetic Kinetic Endpoint Endpoint
Prosedur
kerja
4:1 4:1 1:1 4:1 Mono 4:1
Reagen 1 (μl) 800 800 500 800 800
1000
Reagen 2 (μl) 200 200 500 200 200
Sample (μl) 10 100 100 20 10 20
Inkubasi (min) 5 - - - 10 -
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 18/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 19/26
SOPs
Microlab 300 & 300LX
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 20/26
SOPs
Microlab 300 & 300LX
Parameter Calcium Chloride Cholesterol HDL- Dir LDL - Dir CK-NAC CKMB
General
Name Calcium Chlo Chol HDLL LDLL CKNAC CKMB
Units mg/dl mEq/dl mg/dl mg/dl mg/dl U/L U/L
R-Blank Yes Yes Yes No No No No
S-Blank No No No Yes Yes No No
Replication 1 1 1 1 1 1 1
Settings
Mode End-point End-point End-point End-point End-point Kinetic Kinetic
Wavelength 660 505 505 578 578 340 340
Volume 400 400 400 400 400 400 400
T-Faktor 1.000 1.000 1.000 1.000 1.000 1.000 1.000
Delay 60 3 3 3 3 120 180
Meas.-Time - - - - - 180 180
Non-Linear - - - - - 10 10
Bi-Chrom No No No No No - -
Wavel. 2 - - - - - - -
Limits
Low-Abs -0.100 -0.100 -0.100 -0.100 -0.100 -0.100 -0.100
High-Abs 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Ref-Low Entry Entry Entry Entry Entry Entry Entry
Ref-High Entry Entry Entry Entry Entry Entry Entry
Decimals 2 1 1 1 1 1 1
L-Abs RB 0.100 0.100 0.100 0.100 0.100 0.100 0.100
H-Abs RB 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Calibration
Method 1-Point 1-Point 1-Point 1-Point 1-Point Factor Factor
Name Elical Elical Std Cal-HDL Cal-LDL - -
Factor 1 1 1 1 1 4127 8254
Replication 1 1 1 1 1 - -
Deviation Entry Entry Entry Entry Entry - -
L-Abs 0.100 0.100 0.100 0.100 0.100 - -
H-Abs 2.300 2.300 2.300 2.300 2.300 - -
Standards
Value Entry Entry Entry Entry Entry - -
Abs
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 21/26
SOPs
Microlab 300 & 300LX
General
Name Crea Crea GGT Gluc Gluc HDLP Iron
Units mg/dl mg/dl U/L mg/dl mg/dl mg/dl µg/dl
R-Blank No No No Yes Yes Yes No
S-Blank No No No No No No Yes
Replication 1 1 1 1 1 1 1
Settings
End- End-
Mode Kinetic Kinetic Kinetic End-point End-point
point point
Wavelength 505 546 405 340 505 505 546
Volume 400 400 400 400 400 400 400
T-Faktor 1.000 1.000 1.000 1.000 1.000 1.000 1.000
Delay 10 60 60 3 3 3 3
Meas.-Time 120 120 30 - - - -
Non-Linear 10 20 10 - - - -
Bi-Chrom No No No No No No No
Wavel. 2 - - - - - - -
Limits
Low-Abs -0.100 -0.100 -0.100 -0.100 -0.100 -0.100 -0.100
High-Abs 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Ref-Low Entry Entry Entry Entry Entry Entry Entry
Ref-High Entry Entry Entry Entry Entry Entry Entry
Decimals 2 1 1 1 1 1 1
L-Abs RB 0.100 0.100 0.100 0.100 0.100 0.100 0.100
H-Abs RB 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Calibration
Method 1-Point 1-Point Faktor 1-Point 1-Point 1-Point 1-Point
Name Elical Elical - Std Std Std Std
Factor 1 1 1510 1 1 1 1
Replication 1 1 - 1 1 1 1
Deviation Entry Entry - Entry Entry Entry Entry
L-Abs 0.100 0.100 - 0.100 0.100 0.100 0.100
H-Abs 2.300 2.300 - 2.300 2.300 2.300 2.300
Standards
Value Entry Entry - Entry Entry Entry Entry
Abs
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 22/26
SOPs
Microlab 300 & 300LX
Magne- Phospo Total Triglyce- Urea
Parameter LDHL Uric Acid
sium rus Protein rides UV
General
Name LDH Magn Phos TP Trig Urea UA
Units U/L mg/dl mg/dl g/dl mg/dl mg/dl µg/dl
R-Blank No Yes Yes Yes Yes No Yes
S-Blank No No No No No No No
Replication 1 1 1 1 1 1 1
Settings
End- End-
Mode Kinetic End-Point End-Point Kinetic End-Point
point Point
Wavelength 340 505 340 546 505 340 546
Volume 400 400 400 400 400 400 400
T-Faktor 1 1.000 1.000 1.000 1.000 1.000 1.000
Delay 100 3 3 3 3 30 3
Meas.-Time 180 - - - - 100 -
Non-Linear 10 - - - - 20 -
Bi-Chrom - No No No No - No
Wavel. 2 - - - - - - -
Limits
Low-Abs -0.100 -0.100 -0.100 -0.100 -0.100 -0.100 -0.100
High-Abs 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Ref-Low Entry Entry Entry Entry Entry Entry Entry
Ref-High Entry Entry Entry Entry Entry Entry Entry
Decimals 1 1 1 1 1 1 1
L-Abs RB 0.100 0.100 0.100 0.100 0.100 0.100 0.100
H-Abs RB 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Calibration
Method 1-Point 1-Point 1-Point 1-Point 1-Point 1-Point 1-Point
Name Elical Elical Elical Std Std Std Std
Factor 1 1 1 1 1 1 1
Replication 1 1 1 1 1 1 1
Deviation Entry Entry Entry Entry Entry Entry Entry
L-Abs 0.100 0.100 0.100 0.100 0.100 0.100 0.100
H-Abs 2.300 2.300 2.300 2.300 2.300 2.300 2.300
Standards
Value Entry Entry Entry Entry Entry Entry Entry
Abs
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 23/26
SOPs
Microlab 300 & 300LX
8. Maintenance Tahunan
MRK Diagnostics menawarkan pelayanan kunjungan service melalui
telepon sebagaimana kontrak service yang sudah ada. Kontrak service
memberikan jaminan biaya kunjungan dan potongan / diskon untuk
sparepart dan alat-alat yang habis pakai (seperti: Filter, Lampu, dll).
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 24/26
SOPs
Microlab 300 & 300LX
9. Kalibrasi Alat
Teknisi dari MRK Diagnostics sebagian besar
dilatih langsung oleh teknisi Produsen dan sudah
memiliki pengalaman selama bertahun - tahun.
Pengalaman ini sangat penting saat kalibrasi
instrumen untuk menjamin mereka siap untuk
menghasilkan hasil yang benar
1. Kalibrasi
2. Validasi
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 25/26
SOPs
Microlab 300 & 300LX
Jika Anda ingin melakukan kalibrasi alat MicroLab 300 , Microlab 300LX
atau alat lain yang disediakan oleh MRK Diagnostics, jangan ragu untuk
menghubungi kami. Untuk dealer dan pelanggan kami siap untuk
mengkalibrasi sistem dalam kontrak KSO.
Satu lagi tips : Simpan di tempat yang aman kemasan asli Anda! Khusus
untuk pelanggan jarak jauh, ini sangat membantu ketika mengirimkan
instrumen kantor pusat kami untuk dikalibrasi. Hal ini jauh lebih mudah
daripada mengirim teknisi ke Aceh, Papua atau NTT ...
© MRK Diagnostics, 2008-2016, V 2.2 Service Hotline: 0856 950 88000 Page 26/26