Kalibrasi PH Meter
Kalibrasi PH Meter
lower costs
regulatory requirements
protect equipment
What is pH?
Definition pH
Principle of pH measurement
Electrode calibration (adjustment)
Temperature compensation
Definition of pH
pH value defined as:
logaritma negatif dari aktivitas ion hidrogen
pH = - log aH+
• logarithm x of a number n: n = 10x
examples: 100 = 10 * 10 = 102
10 = 10 101
1 = 100
0.1 = 1/10 10-1
0.01 = 1/100
10-2
• activity a = concentration c
pH scale is determined by dissociation (penguraian) of water, H2O:
–
H2O = H+ + OH
0 1 0.00000000000001
1 0.1 0.0000000000001
2 0.01 0.000000000001
acid 3 0.001 0.00000000001
4 0.0001 0.0000000001
5 0.00001 0.000000001
6 0.000001 0.00000001
neutral 7 0.0000001 0.0000001
8 0.00000001 0.000001
9 0.000000001 0.00001
10 0.0000000001 0.0001
alkaline 11 0.00000000001 0.001
12 0.000000000001 0.01
13 0.0000000000001 0.1
14 0.00000000000001 1
Food
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
HCl NaOH 4%
HCN 0.27% CaCO3 (sat)
0.37%
Acetic acid NH4OH 1.7%
H2SO4
0.6% NH4OH 0.017%
4.9 %
Potassium acetate 0.98%
Sodium bicarbonate 0.84%
Chemicals
What is pH?
Definition pH
Prinsip pengukuran pH
Electrode calibration (adjustment)
Temperature compensation
Principle of pH Measurement
glass reference
pH diukur dengan potensiometri electrode electrode
S7 Connector
Filling port
Reference electrolyte
Reference electrolyte
Ceramic junction
Round
Spherical
Cylindrical
Sharp
(puncture electrode)
for medium solid products
Micro
for small samples
Flat
InLab427
for surface
InLab423 InLab426
Different Diaphragm
Ceramic
ceramic frit,
chemical inert connection,
slow electrolyte flow
approx. 1 mL / 24 h
For clear solutions.
Ground-glass
ground-glass junction
with movable PTFE-ring.
Fast electrolyte flow
approx. 4 mL / 24 h
For suspensions and oils.
InLab420
InLab412
Reference
®
ARGENTHAL / Silver ion trap
Silver wire
solid state
electrolyte
open junction
Temp sensor
• Pengukuran pH didasarkan pada perbedaan potensial listrik
antara dua elektroda, satu elektroda berfungsi sebagai reference
dan yang lain berfungsi sebagai elektroda pengukur.
• Jika elektroda dicelupkan ke dalam larutan, maka elektroda akan
memberikan perubahan potensial listrik sesuai dengan
konsentrasi ion Hidronium dalam larutan, sedangkan potensial
listrik dalam reference adalah tetap.
What is pH?
Definition pH
Principle of pH measurement
Electrode calibration (adjustment)
Temperature compensation
E = Eo + 2.3 RT/F • log aH+ = Eo – 2.3 RT/F • pH
E Measured potential
Eo Standard potential
R Universal gas constant
T Temperature (Kelvin)
F Faraday constant
2,3 RT/F Nernst potential (slope)
aH+ H+ activity of solution
Before calibration
After calibration
pH
4 7 9
Zero-point = pH 6.9
- mV
2. Slope adjustment Buffer pH 4.00
+ mV
Before calibration
pH
4 7 9
After calibration
• Tergantung pada kualitas elektroda. Elektroda lama ... kalibrasi lebih sering.
oC mV/pH oC mV/pH
+ mV
0 54.2 50 64.1
10 56.2 60 66.1
pH
7
20 oC 20 58.2 80 70.1
25 oC
25 59.2 90 72.1
40 oC
- mV 30 60.1 100 74.0
40 62.1 110 76.0
Jangan pernah menyimpan elektroda dalam air suling atau kondisi kering.
Replace/refill reference electrolyte regularly.
E.g. every 2 weeks
▪Electrolyte level in electrode must be higher
than in sample solution.
If electrolyte level lower than sample solution, then
sample will contaminate reference system.
p
H
S
i
m
ul
a
t
o
r p
H
M
e
t
er
p
H
7
.
0
0
Rangakaian Kalibrasi dengan Buffer pH
p
H
M
et
e
r
p
H
7
.0
0
1. Memasang bagian output dari pH simulator ke bagian input pH meter.
2. Sesuaikan rentang ukur dan mode antara pH simulator dan pH meter.
3. Nyalakan pH simulator dan pH meter selama 20 menit.
4. Set pH simulator pada kedudukan pH 7 (E= 0 mV, suhu konpensasi 25 °C),
pada Rout Low. Saat yang bersamaan set pH meter pada pH7, suhu 25 °C,
dan slope pada 100 %. Kemudian set pH simulator pada kedudukan pH 7
(suhu 25 °C) pada Rout high. Untuk penunjukan ini, penunjukan pH meter
tidak menyimpang lebih dari ± 0.1 pH. Lihat manual alat tersebut.
5. Untuk kedudukan ini (pH simulator, pH 7, Rout high) set pH meter tepat
pada kedudukan pH 7, tunggu ± 30 menit. Setelah 30 menit, untuk pH
meter yang baik penunjukannya tetap pada pH 7.
6. Bila ada penyimpangan sedikit, atur kembali pada kedudukan pH 7.
Atur pH simulator pada pH 14 (E = -414.116 mV), catat penunjukan pada
pH meter, atur pH meter hingga penunjukan tepat pada pH 14 (dgn
pengaturan slope).
7. Atur pH simulator pada kedudukan 0 s/d. pH 14. Pada saat yang
bersamaan catat penunjukan pada pH meter. Nilai konversi pH/mV lihat
tabel atau gunakan persamaan :
R.T . ln10
E= [V ]
F
dengan :
E adalah tegangan DC yang dihasilkan untuk setiap kenaikan 1 pH.
R adalah konstanta universal gas (8.31441 J.mol-1.°K-1)
T adalah suhu konpensasi dalam Kelvin
F adalah konstanta Faraday (96484.56 C mol-1)
3. Untuk alat yang mempunyai multi range, beberapa range hanya dilakukan
Kalibrasi pada sekala penuh, salah satu range dilakukan Kalibrasi minimal
3 titik pada skala berangka dari 0 % sd. 100 % .
Model Matematis.
C = Std − Alat
dengan :
C = Koreksi
Std = Pembacaan Nilai Standar
Alat = Pembacaan Nilai Penunjukan Alat
• Ketidakpastian baku Standar
Ub1
ub1 =
k
1 100 2
b1 = ( )
2 R
1 100 2
b2 = ( )
2 R
2 2
𝑠𝑇 𝑆𝑒𝑟𝑡𝑖𝑓𝑖𝑘𝑎𝑡
𝑢 𝑇𝑒𝑚𝑝 = +
𝑛 𝑘
dengan : c1 = 1, c2 = 1, c3 = 1, c4 = pH/oC
U (95) = k .uC
Komponen Satuan Distribusi U Pembagi vi ui ci uici (uici)2 (uici)4/vi