Anda di halaman 1dari 22

FETAL DO P P L E R

KELOMPOK II A
ANGGOTA KELOMPOK

• BINTANG TRI MITRA JAYA (2102003)


• MHD ASYRAF SAPUTRA (2102023)
• MUHAMMAD TEGAR ARIANSYAH (2102032)
• RAHMANDA (2102037)
PEMBAHASAN
• TEORI DASAR FETAL DOPPLER • PERBEDAAN ALAT
• PRINSIP DASAR FETAL DOPPLER • SOP FETAL DOPPLER
• BAGIAN BAGIAN FETAL DOPPLER • PEMELIHARAAN FETAL DOPPLER
• BLOCK DIAGRAM • KALIBRASI FETAL DOPPLER
• WIRING DIAGRAM • KEMUNGKINAN KERUSAKAN FETAL DOPPLER
• JENIS JENIS FETAL DOPPLER
TEORI DASAR
• FETAL DOPPLER MERUPAKAN ALAT YANG DIGUNAKAN UNTUK MENDENGARKAN DETAK JANTUNG JANIN
SELAMA MASIH ADA DIDALAM KANDUNGAN. YANG MENGGUNAKAN PRINSIP PANTULAN GELOMBANG
ELEKTROMAGNETIK, DOPPLER SANGAT BERGUNA UNTUK MENGETAHUI KONDISI KESEHATAN JANIN
DALAM KANDUNGAN IBU.

• DOPPLER MENGGUNAKAN 2 SENSOR YAITU :


ULTRASOUND MENGGUNAKAN TRANSMITTER DAN RECEIVER, KEUNTUNGANNYA LEBIH PEKA DAN AKURAT,
TETAPI HARGANYA LEBIH MAHAL.
MIKROSOUND TIDAK MENGGUNAKAN TRANSMITTER DAN RECEIVER.HANYA MENERIMA, TIDAK
MEMANCARKAN,SEHINGGA KURANG PEKA.
PRINSIP DASAR

• PADA PEMINDAI ULTRASONIK SERANGKAIAN DITRANSMISIKAN UNTUK PERGERAKAN DARAH. GEMA DARI
JARINGAN STASIONER SAMA DARI PULSA KE PULSA. GEMA DARI SCATTERER YANG BERGERAK
MENUNJUKAN SEDIKIT PERBEDAAN PADA WAKTU UNTUK SINYAL DI KEMBALIKAN KE RECIVER
PERBEDAAN INI DAPAT DIUKUR SEBAGAI PERBEDAAN WAKTU LANGSUNG ATAU BIASANYA DALAM
PEGESERAN FRASA DARI MANA FREKUENSI DOPPLER. DIPEROLEH MEREKA KEMUDIAN DIPROSES UNTUK
MENGHSILKAN TAMPILAN ALIR WARNA ATAU SONOGRAM DOPPLER.
BAGIAN BAGIAN ALAT
• PROBE (TRANDUSER)
• MAIN UNIT:
A.TOMBOL POWER
B.BATERAI (1,2V) MENGGUNAKAN 2BATERAI
C.MODE: REAL TIME, RATA-RATA, BACK LIGHT
D.POTENSIO (UNTUK VOLUME)
E.SPEAKER
F. DISPLAY
BLOCK DIAGRAM
• DOPPLER MENGGUNAKAN FREKUENSI SEBESAR 2,25 MHZ YANG DIGUNAKAN UNTUK MENDETEKSI DETAK
JANTUNG JANIN USIA 16 MINGGU, FREKUENSI DIBANGKITKAN OLEH OSCILATOR KEMUDIAN
DIPANCARKAN OLEH TRANSMITTER KE MEDIA PENGUKURAN DAN HASIL PENGUKURAN DITERIMA
KEMBALI OLEH RECIEVER, LALU SINYAL MASUK KE PRE-AMP UNTUK DIKUATKAN KEMUDIAN DISARING
MELALUI FILTER DAN DIKUATKAN OLEH AMPLIFIER (PENGUAT AKHIR).

• KEMUDIAN OUTPUT DARI AMPLIFIER MASUK KE ADC (ANALOG TO DIGITAL CONVERTER) DIRUBAH
MENJADI DATA DIGITAL. KEMUDIAN DITAMPILKAN JUMLAH DETAKAN JANTUNG JANIN YANG TERUKUR
MELALUI DISPLAY DAN SPEAKER
WIRING DIAGRAM
JENIS JENIS
• FETAL DOPPLER LCD LOTUS
FETAL DOPPLER LCD LOTUS MENDETEKSI DETAK JANTUNG JANIN / FETUS HEART RATE (FHR), PALING CEPAT
PD UMUR 12 MINGGU KEHAMILAN. DIGUNAKAN DI RUMAH SAKIT, KLINIK ATAU RUMAH TANGGA BIASA.
-FITUR:
• MENAMPILKAN ANGKA DETAK JANTUNG (FHR) PADA LAYAR BIRU.
• MERELAY SUARA DETAK JANTUNG JANIN, DILENGKAPI DENGAN VOLUME CONTROL SUARA.
• DILENGKAPI DENGAN CHARGER.
• MENGGUNAKAN RECHARGEABLE BATTERY SEHINGGA TIDAK PERLU MEMBELI BATTERY LEAD ACID.
• FETAL DOPPLER BESTMAN BF500++
FETAL DOPPLER BESTMAN BF500++ MERUPAKAN ALAT UNTUK MENDENGAR DETAK JANTUNG JANIN
MELALUI SPEAKER YANG DILENGKAPI DENGAN LAYAR LCD UNTUK MENAMPILKAN ANGKA DETAK JANTUNG
JANIN/ MENIT DAN LAMPU BACKLIGHT BERWARNA BIRU UNTUK MEMUDAHKAN MEMBACA ANGKA PADA
LAYAR LCD.
PADA UMUMNYA DETAK JANTUNG JANIN SUDAH DAPAT TERDETEKSI PADA USIA 14 MINGGU. FETAL DOPPLER
BESTMAN BF500++ DILENGKAPI DENGAN BATERAI YANG DAPAT DI-CHARGE KEMBALI, SERTA MEMILIKI
FITUR ALARM JIKA DETAK JANTUNG JANIN DI BAWAH / DI ATAS NORMAL. FETAL DOPPLER BESTMAN
BF500+ AKAN OTOMATIS SWITCH-OFF SETELAH 1 MENIT PEMAKAIAN.
• HI-BEBE BT-200 LCD FETAL DOPPLER
HI BEBE BT-200 LCD FETAL DOPPLER ADALAH CARA YANG IDEAL UNTUK MEMUNGKINKAN ANDA UNTUK
MENDENGARKAN DETAK JANTUNG BAYI ANDA DALAM KENYAMANAN RUMAH ANDA, DAN DENGAN LAYAR LCD
ANDA DENGAN MUDAH DAPAT MELIHAT DAN MEREKAM DETAK JANTUNG BAYI ANDA. PENGGUNAAN FETAL
DOPPLER DI RUMAH DAPAT DILAKUKAN OLEH IBU HAMIL DI ANTARA KUNJUNGAN KE DOKTER ATAU RUMAH
SAKIT. HI BEBE BT-200 LCD FETAL DOPPLER DENGAN DESAIN YANG RINGKAS MEMILIKI FITUR BERIKUT:
FITUR FITUR:

• ANDA MUNGKIN DAPAT MENDENGAR BAYI ANDA DARI SEJAK DINI (USIA KEHAMILAN 10 MINGGU).
• MENAMPILKAN DETAK JANTUNG JANIN / FETAL HEART RATE (FHR)
• SINYAL MENGENAI KUALITAS TAMPILAN
• CARRYING CASE (TAS) UNTUK PELINDUNG ALAT FETAL DOPPLER
• AMAN DAN MUDAH DIGUNAKAN
• COCOK UNTUK PENGGUNAAN KLINIS DAN DI RUMAH
• TERMASUK BATERAI 2 X 1.5V (AA)
• TERMASUK USG GEL
PERBEDAAN ALAT
SOP PENGOPERASIAN FETAL DOPPLER
• PETUGAS MENEKAN TOMBOL POWER DOPPLER SAMPAI DALAM KEADAAN ON
• PETUGAS MENYIAPKAN JELLY DAN TISSSUE
• PETUGAS MENGOLESKAN JELLY DI DINDING PERUT IBU PADA POSISI JANIN YANG
• TEPAT UNTUK DETEKSI JANTUNG JANIN
• PETUGAS MENEMPELKAN DOPPLER PADA DINDING PERUT IBU YANG TELAH DIBERI
• JELLY DAN MENGHITUNG DETAK JANTUNG JANIN (NORMAL 120-160X/MENIT)
• PETUGAS MENEKAN TOMBOL POWER SAMPAI DALAM KEADAAN OFF
• PETUGAS MENGHAPUS JELLY DENGAN TISSUE
• PETUGAS MENYIMPAN DOPPLER PADA TEMPATNYA
• DOPPLER DISIMPAN DALAM TEMPATNYA
PEMELIHARAAN

A. BERSIHKAN DOPLER DARI SISA JELLY DENGAN TISSUE SAMPAI BERSIH


B. CEK INDIKATOR BATERAI
C. CHARGE BILA BATERAY LEMAH SAMPAL PENUH
D. RAPIKAN DAN SIMPAN ALAT
KALIBRASI ALAT
A. PERSIAPAN
1. LAKUKAN PENDATAAN ALAT..
• CATAT MERK, TYPE, SERIAL NUMBER, TAHUN PEMAKAIAN DARI UUT
• CATAT MERK, TYPE, SERIAL NUMBER DARI KALIBRATOR
• DATA WAKTU PELAKSANAN KALIBRASI
2. CATAT KONDISI AWAL LINGKUNGAN, YAITU SUHU DAN KELEMBAPAN MENGGUNAKAN
THERMOHYGROMETER.
3. LAKUKAN PENGECEKAN FISIK PADA ALAT
B. UJI PERFORMA
1. PASANG PROBE PADA FETAL DOPPLER DAN PASANG SENSOR UNTUK FETAL DOPPLER KALIBRASI
2. SAMBUNGKAN MAIN POWER CABLE KE STOP KONTAK 3. TEKAN TOMBOL POWER PADA UUT DAN KALIBRATOR
3. TUNGGU PROSES INISIALISASI
4. SETTING PADA KALIBRATOR DENGAN MODE FETAL
5. BERI GEL PADA PROBE SEBELUM MENEMPELKANNYA PADA SENSOR KALIBRATOR
6. TEMPELKAN PROBE UUT KE KALIBRATOR TEPAT PADA SENSORNYA
7. LAKUKAN PENGUKURAN DENGAN RENTANG 30 BPM-240 BPM
8. LAKUKAN PENGUKURAN SEBANYAK 3X
9. CATAT HASIL PENGUKURAN
10. SETELAH SELESAI, BERSIHKAN DAN KEMBALIKAN ALAT SEPERTI POSISI SEMULA
11. LAKUKAN PENGECEKAN SUHU DAN KELEMBAPAN RUANGAN SETELAH MELAKUKAN KALIBRASI
KEMUNGKINAN KERUSAKAN
1. CHARGER BATERAI
SEBAGAI PENGISI SUMBER TEGANGAN PADA BATERAI, CHARGER MEMPUNYAI PERANAN PENTING DIMANA
APABILA CHARGER TIDAK MENGELUARKAN TEGANGAN MAKA BATERAI PADA DOPPLER TIDAK AKAN TERISI
DAN ITU BERARTI RANGKAIAN TIDAK MENDAPAT SUPLLY TEGANGAN SEHINGGA TIDAK DAPAT BEKERJA.
UNTUK ITU LAKUKAN PENGECEKAN CHARGER DENGAN MENGGUNAKAN ALAT UJI TEGANGAN ATAU
MULTIMETER. JIKA JACK CHARGER TIDAK MENGELUARKAN TEGANGAN KEMUNGKINAN KABEL PENGHANTAR
ATAU TRAFO PUTUS.
2. BATERAI
SEBAGAI PENYIMPAN DAN PENYEDIA TEGANGAN PADA RANGKAIAN JUGA MEMPUNYAI PERANAN PENTING
PADA SEBUAH PERALATAN ELEKTRONIK, APABILA TERJADI KERUSAKAN PASTINYA RANGKAIAN TIDAK
DAPAT MENYALA, KERUSAKAN PADA BATERAI DOPPLER DITANDAI DENGAN APABILA CHARGER
DIHUBUNGKAN, PERALATAN DAPAT MENYALA DAN AKAN MATI KEMBALI APABILA CHARGER DILEPAS, INI
MENANDAKAN BAHWA BATERAI SUDAH MENGALAMI PENURUNAN DAYA SIMPAN (DROP) ATAU BAHKAN
SUDAH TIDAK DAPAT MENYIMPAN DAYA LAGI.
3. KONEKTOR
SEBAGAI PENGHUBUNG RANGKAIAN ATAU TEGANGAN, SEPERTI PADA KERUSAKAN DIATAS DIMANA
KERUSAKAN DOPPLER TIDAK DAPAT MENYALA ATAU TIDAK DAPAT MENGELUARKAN SUARA DETAK JANTUNG
DISEBABKAN DARI KONEKTOR BATERAI YANG SUDAH AUS SEHINGGA TIDAK DAPAT MENGHANTARKAN
TEGANGAN DARI BATERAI. CARA MENGATASI DAN MEMPERBAIKINYA ADALAH MEMBERSIHKAN ATAU
MENGGANTI KONEKTOR YANG RUSAK.