OPL ( One Point Lesson )

PROSEDUR KALIBRASI No Dokumen


Topik / Tema VORTEX FLOW METER YOKOGAWA Department /Seksi E&I PM1 & PM2
Bagian mesin Vortex yokogawa

OPL INI DI GUNAKAN SEBAGAI PANDUAN SAAT KALIBRASI VORTEX FLOWMETER YOKOGAWA
DENGAN SOURCE DARI FLUKE 702
1. TOOLS KALIBRASI YANG DI PAKAI
- FLUKE 702 UNTUK SOURCE FREKWENSI
- MT-220 UNTUK MENGUKUR CURRENT OUTPUT

2. KONEK CONVERTER VORTEX KE MT-220


- CHECK SPAN FREKWENSI DI PARAMETER K36

2. PERSIAPAN SETTING FLUKE 702


- ON KAN POWER FLUKE
- TEKAN TOMBOL "SETUP"
- TEKAN " ENTER"
- PILIH " " PILIH " ENABLE 24VDC"
- TEKAN " ENTER"
- TEKAN " DONE" SOFT KEY PADA MONITOR FLUKE
- TEKAN "MEAS/SOURCE" SAMPAI KELUAR TAMPILAN SOURCE
- PILIH " V~ Hz "
- TEKAN "ENTER"
- MASUKAN NILAI FREKUENSI LALU TEKAN " ENTER"

FLUKE 702
OPL ( One Point Lesson )
PROSEDUR KALIBRASI No Dokumen
Topik / Tema VORTEX FLOW METER YOKOGAWA Department /Seksi E&I PM1 & PM2
Bagian mesin Vortex yokogawa

3. CHECK KALIBRASI
- RANGKAI KABEL DARI SOURCE FLUKE 702 KONEK DI TERMINAL A ( + ) DAN C ( - )
- MASUKAN NILAI FREQUENSI YANG MAU DI SOURCE 0 - 100% RANGE
- CHECK SETIAP PERUBAHAN OUTPUT PADA MANOMETER MT-220
- CHECK SETIAP PERUBAHAN PADA DISPLAY CONVERTER
- CATAT PADA FORM KALIBRASI DAN HITUNG ERROR
- APABILA ERROR LEBIH DARI 0,3% LAKUKAN KALIBRASI

WIRING KABEL

CURRENT OUTPUT

DISPLAY FLOW

FREKWENSI SOURCE