Anda di halaman 1dari 10

Tgl Unit Auditor Nama Peralatan Medis

18 Mei 21 NICU Dyah Candra DP-MNT-NCU-0816


18 Mei 21 NICU Dyah Candra DP-MNT-NCU-0815
18 Mei 21 NICU Dyah Candra DP-MNT-ICU-0954
9 Juni 2021 ICU Dyah Candra DP-VEN-ICU 0991
9 Juni 2021 ICU Dyah Candra DP-MNT-ICU-0960
18 Juni 2021 ICU Dyah Candra DP-VEN-NCU-0832
18 Juni 2021 ICU Dyah Candra DP-VEN-NCU-0831
18 Juni 2021 ICU Dyah Candra DP-MNT-ICU-0953
18 Juni 2021 ICU Dyah Candra 5001-1700009602
2 JULI 2021 ICU Dyah Candra DP-MNT-ICU-0954
9 JULI 2021 ICU Dyah Candra DP-VEN-ICU 0991
16 JULI 2021 ICU Dyah Candra DP-MNT-ICU-0960
23 JULI 2021 ICU Dyah Candra DP-VEN-NCU-0832
30 JULI 2021 ICU Dyah Candra DP-VEN-NCU-0831
6 AGUSTUS 2021 ICU Dyah Candra DP-VEN-ICU 0991
6 AGUSTUS 2021 ICU Dyah Candra DP-SYR-ICU-3254
13 AGUSTUS 2021 ICU Dyah Candra DP-SYR-ICU-3247
13 AGUSTUS 2021 ICU Dyah Candra 280245
13 AGUSTUS 2021 ICU Dyah Candra 280265
20 AGUSTUS 2021 ICU Dyah Candra DP-MNT-ICU-0953
20 AGUSTUS 2021 ICU Dyah Candra DP-VEN-NCU-0832
26 AGUSTUS 2021 ICU Dyah Candra DP-INC-NCU-0799
26 AGUSTUS 2021 ICU Dyah Candra DP-IRS-NCU-0804
Terjadi malfungsi
Setting alarm
Jenis Peralatan Medis sesuai Alarm dapat alarm ketika
berbunyi peralatan medis
paramater
sedang digunakan
Bedside Monitor Ya Ya Ya
Bedside Monitor Ya Ya Ya
Bedside Monitor Tidak Ya Ya
Central Ventilator Ya Ya Tidak
Bedside Monitor Ya Ya Ya
Central Ventilator Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Bedside Monitor Ya Ya Tidak
Bedside Monitor Ya Ya Tidak
Bedside Monitor Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Bedside Monitor Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Syringe Pump Ya Ya Tidak
Syringe Pump Ya Ya Tidak
IV Pump Ya Ya Tidak
IV Pump Ya Ya Tidak
Bedside Monitor Ya Ya Tidak
Central Ventilator Ya Ya Tidak
Incubator Ya Ya Tidak
Dokumentasi
Terdapat laporan
malfungsi alarm insiden malfungsi Lainnya
pada berkas rekam
alarm (bila terjadi)
medik
NA Ya
NA Ya
NA Ya
NA NA
NA Ya
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
NA NA
ITEM NUM DENUM CAPAIAN TARGET
Setting alarm sesuai paramater 21 22 95% 80%
Alarm dapat berbunyi 22 22 100% 80%
Terjadi malfungsi alarm ketika peralatan medis sedang
digunakan 18 22 82% 0%
Dokumentasi malfungsi alarm pada berkas rekam medik 0 4 0% 80%
Terdapat laporan insiden malfungsi alarm (bila terjadi) 4 4 100% 80%

Anda mungkin juga menyukai