TUGAS TRANSLATE JURNAL ILMIAH BERKAITAN DENGAN INSTRUMENT

ATOMIC ABSORPTION SPECTROFOTOMETRY (AAS)


“Validation of Mineral (Fe, Zn, and Cu) Analysis Methods in Carbohydrate, Protein and Fat-
Rich Samples Using Microwave Digestion Method”

“Analisis Penetapan Kadar Mineral (Fe,Zn dan Cu) pada sampel Karbohidrat, Protein dan Tinggi
Lemak Menggunakan Metode Destruksi Microwave”

Oleh :

Kelompok 1

1. Ni Komang Diah Ernawati 1810511010


2. Audrey Sophia Rachma Putri 1810511062
3. Pande Made Agus Suandayana Diputra 1810511064
4. I Kadek Dwi Andi Krishna Putra 1810511072

BALAI BESAR PENGAWAS OBAT DAN MAKANAN (BBPOM)

DI DENPASAR

TAHUN 2021
“Analisis Penetapan Kadar Mineral (Fe,Zn dan Cu) pada sampel Karbohidrat, Protein dan Tinggi
Lemak Menggunakan Metode Destruksi Microwave”

PENDAHULUAN

Makanan tersusun atas berbagai komponen antara lain air, karbohidrat, protein, vitamin,
lemak dan mineral yang mungkin dapat mempengaruhi sifat dari makanan. Secara khusus, mineral
esensial (Fe,Zn, dan Cu) berhubungan erat dengan beberapa efek merugikan pada makanan seperti
ketengikan,perubahan warna, rasa, dan aroma. Spektrofotometri Serapan Atom dengan atomizer
nyala api merupakan metode yang umum digunakan untuk mengukur kandungan mineral dalam
makanan. Proses destruksi dianggap sebagai salah satu langkah yang paling penting, desktruksi
dapat dilakukan dengan metode kering ataupun basah. Metode desktruksi kering kerap kali
menyebabkan hilangnya beberapa komponen mineral selama berada dalam tungku, selain itu
metode ini juga memerlukan waktu yang relative lebih lama. Sementara itu, metode destruksi
basah sistem terbuka berkontribusi terhadap pencemaran lingkungan dan mineral tertentu dapat
hilang karena penguapan. Di sisi lain, destruksi basah sistem tertutup menggunakan bantuan
microwave digestion memiliki beberapa keunggulan, seperti lebih cepat, mudah, menghasilkan
data pengukuran yang lebih akurat, serta menggunakan lebih sedikit reagen.

Pada penelitian kali ini, metode destruksi dengan bantuan microwave digestion ternyata
lebih efisien perihal waktu pemrosesan, jumlah pelarut, dan suhu dan dianggap sebagai metode
analisis yang lebih relevan dibandingkan dengan metode yang ditetapkan oleh Association of
Official Agricultural Chemist (AOAC) 2012. Berdasarkan kepada standar operasional prosedur
untuk microwave digestion, prosedur destruksi mungkin akan bervariasi tergantung pada jenis
komponen matriks penyusun pada makanan. Sebaliknya, prosedur metode Association of Officia l
Agricultural Chemist (AOAC) 2012 berlaku untuk semua jenis sampel, terlepas dari matriks
penyusun makanan tersebut. Hasil dari analisis antara metode AOAC dengan metode yang
dimodifikasi tidak berbeda nyata, menghasilkan rentang data analisis positif.

Metode destruksi mineral dalam penelitian ini berasal dari panduan perangkat microwave
digestion berbeda dengan metode standar Association of Official Agricultural Chemist (AOAC)
2012. Perbedaannya meliputi volume reagen, suhu, dan lama destruksi. Matriks makanan yang
berbeda memerlukan perlakuan yang spesifik selama proses destruksi. Untuk alasan ini, sampel
makanan dipilih sesuai dengan matriks segitiga makanan (AOAC 2012), menghasilkan tiga
kelompok makanan, yaitu makanan kaya protein (ikan kalengan), makanan kaya lemak (susu
bubuk full cream) dan makanan kaya karbohidrat (kacang polong kalengan). Pemilihan sampel
makanan kaleng bertujuan untuk mengurangi faktor-faktor yang dapat menyebabkan
ketidakstabilan dalam validasi metode. Metode yang dimodifikasi dapat diterima asalkan
memenuhi persyaratan setelah dibandingkan dengan metode standar. Tujuan dari penelitian ini
adalah (1) untuk memvalidasi metode kuantifikasi Fe, Zn dan Cu menggunakan instrument
spektrofotometri serapan atom nyala api (Flame-AAS) dengan bantuan microwave digestion untuk
proses destruksi, dan (2) untuk mendapatkan data tentang kandungan mineral Fe, Zn dan Cu pada
beberapa makanan yang berbeda matriks menggunakan metode yang divalidasi.

METODE PENELITIAN

Bahan

Bahan kimia yang digunakan adalah yang memenuhi standar analisis kimia seperti
aquademineral, HNO3(p) (Merck KgaA), H2O2 30% (MerckKgaA), larutan standar Zn 1000 ppm
(CertiPUR Merck KgaA, no 1.19806), larutan standar Fe 1000ppm (CertiPUR gabungan KgaA,
no.1. 19781), standar Cu larutan 1000 ppm (CertiPUR gabungan KgaA, no.1.1986).

Sampel Makanan (kacang polong kalengan, full cream susu bubuk, dan ikan kaleng)
dikumpulkan dari merek serupa, tanggal kedaluwarsa, dan batch produksi. Susu bubuk disimpan
dalam kemasan kedap udara disuhu kamar (metode quartering sampling), sedangkan kacang
polong kalengan dan ikan kalengan disimpan di dalam freezer (metode sampling homogenisasi)
(FDA 2014). Sampel yang dihomogenkan ditempatkan pada wadah polietilen dan kandungan
abunya dianalisis dalam dua rangkap. Homogentias sampel diperiksa menggunakan kriteria uji
beda (uji F) dengan ketentuan Fhitung < Ftabel. Uji kadar abau dapat dilakukan dengan menggunakan
metode standar AOAC 900.02A.

Alat

Instrument yang digunakan adalah Microwave Digestion system tertutup tipe SINEO
MDS-6G, Spektrofotometri Serapan Atom nyala api tipe Shimadzu AA 2000 dan lampu katoda
untuk Fe,Zn, dan Cu.
Prosedur

Prosedur Destruksi Sampel

Sampel seberat 0,5 g ditambahkan dengan HNO3 dan H2O2 30%. Untuk pengaturan metode
standar dari destruksi dan metode yang dimodifikasi disajikan pada tabel 2. Setelah melalui proses
destruksi, sampel akan dihomogenkan dengan air demineralisasi dalam labu tentukur 25 mL.
Larutan baku (terdiri dari asam nitrat dan hydrogen peroksida) dibuat sesuai dengan prosedur yang
sama seperti sampel.

Validasi Metode Analisis

Metode yang dikembangkan dievaluasi untuk akseptabilitas menurut indikator berikut,


yaitu, linearitas, LOD relatif dan absolut, LOQ, presisi, akurasi, dan reproduktifitas intra-lab.

1. Tes presisi.
Tes ini dilakukan dengan mengukur konsentrasi sampel sebanyak 7 kali ulangan untuk
pengukuran 3 jenis mineral (Fe, Cu, dan Zn).
2. Reproduksibilitas intra – lab.
Reproduksibilitas intra-lab ditentukan dengan mengukur konsentrasi sampel dalam
laboratorium yang sama, oleh operator yang sama dengan waktu yang berbeda. Setiap
pengujian dilakukan dengan tiga kali ulangan. Selanjutnya nilai RSD hasil analisis
dibandingkan dengan nilai RSD Horwitz. Reproduksibilitas intra-lab dianggap memenuhi
jika memiliki nilai RSD lebih kecil dari RSD nilai Horwitz. Hasil tes pada dua minggu
yang berbeda adalah diuji menggunakan uji-t.
3. Akurasi
Pengujian dilakukan dengan menggunakan sampel spike dengan larutan standar.
konsentrasi percobaan spiking adalah persentase konsentrasi sampel sample yaitu 80%
(konsentrasi rendah), 100% (sedang), dan 120% (tinggi). Spiking dilakukan di awal
persiapan sebelum pencernaan microwave. Akurasi ditentukan dengan menghitung
recovery percent.
4. Linearitas
Linieritas instrumen dilakukan dengan mengukur absorbansi larutan mineral standar (Fe,
Zn, dan Cu) pada berbagai konsentrasi. Linearitas uji mineral Fe dan Cu menggunakan
konsentrasi 0-0,5 mg/L sedangkan konsentrasi mineral Zn adalah 0-1,0 mg/L. Absorbansi
masing-masing larutan standar diukur dan dilakukan pengulangan tiga kali.
5. LOD (Limit of Detection) & LOQ (Limit of Quantitation)
Nilai LDI (Limit detection of instrument / batas deteksi instrument) dan LOQ dapat
diperoleh dari data analisis linieritas dengan menentukan nilai standar kesalahan kalibrasi
(sy/x). Secara teoritis LDI dapat diperoleh dengan rumus [3(sy/x)]/b. Konfirmasi dilakukan
sebanyak 7x secara berulang pengukuran solusi sesuai dengan teori Konsentrasi LDI untuk
3 jenis mineral (Fe, Zn dan Cu), LOQ teoritis diturunkan dari [10 (sy/x)]/b. Konfirmasi
nilai LOQ dilakukan dengan membuat larutan dengan konsentrasi LOQ teoretis nilai dan
diukur dengan tujuh ulangan untuk masing-masing mineral (Fe, Zn, dan Cu). LDM (Batasi
deteksi metode) nilai ditentukan dengan mengukur tujuh replikasi konsentrasi mineral
dalam sampel disiapkan sesuai dengan nilai LDI teoritis
6. Analisis Data
Data dinyatakan sebagai mean, diikuti dengan standar deviasi dan standar deviasi relative
(RSD). Selanjutnya, analisis statistik digunakan untuk tentukan nilai p (uji t independen)
di intra-lab reproduktifitas. Hasil yang didapat dari modifikasi metode dibandingkan
dengan metode standar
HASIL DAN PEMBAHASAN

Persiapan Sampel

Sebelum destruksi (dengan microwave digestion), ukuran sampel perlu dikurangi,


memungkinkan untuk menghasilkan luas permukaan yang lebih besar dan meningkatkan laju
reaksi. Sampel yang mengandung pelarut organik, seperti wine, perlu diuapkan hingga seperlima
dari volume awal karena pelarut dapat menyebabkan ledakan.

Pemilihan matriks makanan didasarkan pada komponen mayor seperti karbohidrat, protein,
dan lemak, diatur ke dalam segitiga matriks makanan (Gbr. 1). Itu analisis proksimat sesuai dengan
penempatan pada studi literatur (USDA 2006) [24], ditampilkan pada Tabel 3.

Selanjutnya, kadar air di semua sampel (Tabel 4) berada pada kisaran standar SNI dan
USDA. Kelembaban tingkat ikan kalengan ditemukan lebih tinggi dibandingkan dengan literatur
USDA 15089, yang mungkin dianggap berasal dari perbedaan jenis sampel. Penentuan dari kadar
air dilakukan dengan menggunakan metode oven vakum. Pengeringan oven vakum dilakukan pada
suhu dan tekanan yang rendah, untuk mempertahankan volatil komponen yang ada dalam sampel
yang diuji

Uji homogenitas dilakukan dengan mengukur kadar abu, memungkinkan untuk


memastikan bahwa sampel diuji untuk mengandung jumlah mineral yang sebanding dan tidak
mengandung matriks yang mengganggu (Tabel 5). Kandungan abu bisa digunakan sebagai
gambaran tentang mineral yang terkandung dalam sampel. Hasilnya menunjukkan bahwa nilai F
adalah 3,02, lebih rendah dari F tabel 5%, menunjukkan homogenitas yang dapat diterima.
Metode Validasi

Presisi

Hasil validasi metode yang disajikan pada Tabel 6 menunjukkan bahwa penentuan Fe, Zn
dan Cu disemua sampel telah presisi. Dimana ditunjukkan oleh % RSD < 2/3 CV Horwitz. Hasil
ini menunjukkan bahwa semua kesalahan acak dapat dikendalikan. Kesalahan acak termasuk
ketidakstabilan kinerja instrumen, perubahan suhu, perbedaan teknik, dan analisis

LOD & LOQ

Indicator batas deteksi yakni Absolut LOD (LOD Instrumen) dan Relatif LOD (metode
LOD). Absolut LOD mewakili konsentrasi terkecil dari analit yang dapat dideteksi oleh alat.
Sedangkan Relatif LOD adalah konsentrasi analit terkecil dalam sampel yang dapat dideteksi dan
hasilnya disajikan pada Tabel 7. Batas pengujian deteksi dilakukan dengan menggunakan kurva
kalibrasi dengan rumus 3[syx/b]. Nilai LOQ merupakan batas terendah dari kurva kalibrasi.

Hasil untuk Relative LOD dapat dilihat pada Tabel 8. Nilai Absolute LOD untuk ketiga
mineral yang diperiksa lebih kecil dari yang dilaporkan oleh penelitian sebelumnya, dimana pada
penelitian sebelumnya mencapai 0,07 mg/mL untuk Cu, 0,07 mg/mL untuk Fe, dan 0,11 mg/mL
untuk Zn. Yang berarti metode tersebut memiliki sensitivitas yang baik. Relative LOD sangat
penting dalam menangani kendala karena sinyal/tanggapan yang berbeda untuk setiap matriks
makanan.
Linearitas

Persamaan regresi untuk Fe,Cu dan Zn disajikan sebagai berikut: y = 0,084x-0,0002


dengan R2 = 0,9970; y = 0,1335x-0,0004 dengan R2 = 0,9969; y 0,4125x + 0,0086 dengan R2 =
0,9977, berturut-turut. Seperti ditunjukkan pada Tabel 9, hasil eksperimen menunjukkan bahwa
eksperimen memenuhi linearitas, dimana semakin tinggi konsentrasi mineral yang diuji sesuai
dengan kenaikan absorbansi.
Intercept untuk Fe, Zn, dan CU masing-masing mencapai -0,0002, 0,0086 dan -0,0004,
menunjukkan terjadinya kesalahan sistematik pada instrumen F-AAS. Selain itu, R2 nilai adalah >
0,990, yang sesuai dengan batas penerimaan, seperti yang digambarkan dalam Gambar. 2.
Intra-lab reproduktifitas

Tabel 10 menyajikan reproduksibilitas intra-lab, menunjukkan p-value> α 5% untuk semua


mineral pada ikan kaleng, kacang polong kalengan dan susu bubuk full cream, menunjukkan
bahwa nilainya tidak berbeda nyata.
Akurasi

Akurasi mewakili efisiensi metode berdasarkan pemulihan dari larutan standar yang
ditambahkan ke sampel (spiking). Hasil penelitian menunjukkan bahwa akurasi kadar Fe, Zn, dan
Cu bervariasi dari 80 sampai 100%, yang masih dalam kisaran batas akseptabilitas AOAC (75-
120%). Ini menunjukkan bahwa prosedur dalam metode yang dimodifikasi dapat diterima untuk
semua mineral. Hasil uji recovery dapat dilihat pada Tabel 11.

Analisis Mineral

Untuk mengoptimalkan hasil, beberapa kondisi selama fase destruksi dalam microwave
digestion perlu dilakukan dikontrol, termasuk jumlah asam nitrat dan hidrogen peroksida, suhu,
berat sampel, pra-digest, dan panjang destruksi. Pengoptimalan dapat dicapai dengan
menggunakan berbagai teknik, seperti desain faktorial Plackett-Burman. Namun, langkah ini tidak
termasuk, mengingat semua prosedur dalam hal ini pekerjaan didasarkan pada standar pencernaan
gelombang mikro instrumen untuk setiap matriks makanan. efisiensi dari proses penghancuran
menggunakan microwave destruksi adalah dipengaruhi oleh volume zat asam (HNO3 dan H2O2),
waktu, dan suhu.

Pengaturan daya di kedua metode (standar dan dimodifikasi) diadopsi dari buku manual,
sementara suhu penghancuran dilakukan sesuai dengan standar AOAC (2012). Berat sampel (0,5g
berat kering) juga didasarkan pada pengaturan AOAC. Proses destruksi mengakibatkan terjadinya
perubahan bentuk dari senyawa anorganik menjadi senyawa organic, dimana melibatkan konversi
unsur – unsur organic seperti C, H, O, N, S dan P menjadi CO2, H2O, NO2 sedangkan unsur –
unsur anorganik masih ada dalam bentuk aslinya. Destruksi menggunakan microwave digestion
memiliki beberapa keunggulan, termasuk waktu operasional yang lebih cepat, lebih sedikit
menggunakan reagen, akurasi yang lebih tinggi, mencegah kontaminasi dalam sampel dan
hilangnya komponen volatil.

Selama proses destruksi menggunakan microwave digestion, molekul – molekul terkena


radiasi elektromagnetik non-ionik, yang menginduksi momen rotasi dipol tanpa mengubah
struktur molekulnya. Efek gelombang pada microwave muncul secara magnetis dan secara
dielektrik. Bahan dielektrik terdiri dari molekul yang memperoleh momen dipol atau muatan netral
(bermuatan positif dan negative yang seimbang). Selain itu, gelombang dielektrik dalam
microwave disebabkan oleh kemampuan menyimpan muatan listrik pada berbagai potensi. Air
adalah molekul listrik, memiliki atom hidrogen (bermuatan positif) dan atom oksigen (bermuatan
negatif). Paparan medan listrik dengan bahan dielektrik menyebabkan perubahan arus karena
tolakan dan tarik – menarik dalam muatan positif atau muatan negative, dengan demikian
gelombang terjadi secara magnetis. Lebih banyak panas dihasilkan dalam sampel yang
mengandung lebih banyak air daripada yang sedikit air. Frekuensi dalam microwave digestion
beriksar antara 300 sampai 300.000 MHz dengan panjang gelombang antara 1 mm – 1m.

Sampel yang didestruksi dengan bantuan microwave digestion dicampur dengan asam kuat
dalam vessel yang tertutup. Dalam sistem, tekanan (1,1-2,2 MPa) dan suhu (30-190oC) meningkat
selama proses karena terjadi radiasi gelombang mikro, memungkinkan untuk mengubah wujud
padat menjadi cairan. Tekanan serta suhu yang tinggi (berkisar 200oC) dalam vessel
memungkinkan untuk meningkatkan titik didih dari asam nitrat. Dengan demikian, memperkuat
oksidasi dan mempercepat akselerasi reaksinya. HNO3 dan H2O2 digunakan sebagai oksidator
karena asam – asam tersebut mampu mengoksidasi mineral menjadi bentuk garamnya. Asa nitrat
pada suhu tinggi terdekomposisi menjadi oksida yang larut dalam air seperti NO 2 dan NO.
penggunaan asam campuran lebih baik daripada asam tunggal untuk proses destruksi mineral.
Proses destruksi tidak akan sempurna jika asam nitrat hanya satu – satunya senyawa yang
digunakan, karena akan menghasilkan “cloudy solution”. Penambahan H2O2 menyebabka
intesifitas keberadaan O 2 dalam vessel. Campuran pengoksidasi yang terdiri dari HNO3 dan H2O2
menghasilkan percent recovery yang lebih besar daripada campuran antara HNO3 dan HCl serta
HNO3 dan H2SO4.

Gradien atau sistem pemanasan bertingkat bertujuan mengintensifkan reaksi antara NO dan
O2 untuk membentuk NO2. NO2 direabsobsi dalam larutan asam dan meregenerasi HNO 3. Proses
rekasi di dalam vessel akan dijelaskan pada gambar 3.

Semakin lama proses pemanasan dalam microwave akan meningkatkan suhu dalam vessel,
bertanggung jawab untuk meredam proses kondensasi dan secara tidak langsung meningkatkan
kadar mineral dalam larutan. Proses destruksi yang sempurna ditandai dengan terbentuknya
larutaan bening, yang menunjukan reformasi organic senyawa. Larutan sampel kemudian siap
untuk dianalisis menggunakan Flame-AAS. Flame-AAS direkomendasikan untuk digunakan
dalam mengukur mineral karena sensitivitasnya yang baik dengan ambang batas deteksi kurang
lebih 1 ppm dan juga rendah biaya operasional.

Serat pangan larut air adalah berbasis karbohidrat polimer yang banyak ditemukan dalam
kacang – kacangan, buah – buahan dan sayuran. Serat yang dominan terkandung dalam kacang –
kacangan adalah stachyose (19.83 mg/g) dan verbscose (16.10 mg/g). selain itu kacang kaleng
juga mengandung gula sederhana seperti glukosa, fruktosa, sukrosa, dan galaktosa; oleh karena
itu, prroses destruksi dimulai dari suhu 130oC hingga 180oC. penambahan asam nitrat (6 ml) sesuai
dengan penelitian sebelumnya oleh Bakkaki et al. pada sampel sayur – sayuran, menghasilkan
tingkat akurasi tertinggi.

Susu bubuk full cream kaya akan lemak. Ikatan kovalen antara asam lemak sulit untuk
diputuskan, sementara lemak memiliki potensi untuk teroksidasi relative rendah, yang artinya
dibutuhkan destruksi dengan suhu hingga 200oC dengan jumlah asam pengoksidasi yang lebih
banyak (10ml). Laktosa (D-galaktosa dan D-glukosa) merupakan gula sederhana utama yang
terkandung dalam susu. Oleh karena itu, suhu awal destruksi dimulai pada suhu 130oC. Proses
destruksi seperti ini sebelumnya dilakukan pada penelitian Belete et al. dalam sampel susu sapi
menggunakan suhu destruksi 165oC-190OC setelah 10 menit fase pra-destruksi.

Ikan kaleng mewakili sampel makanan dengan kandungan protein. Kandungan protein
utama pada ikan yaitu aktin dan myosin, rentan terhadap denaturasi ketika terpapar microwave
(gambar 4). Meskipun myofibril protein mulai terdenaturasi pada suhu 85oC, suhu yang lebih
tinggi diperlukan untuk menginduksi denaturasi saat protein ada dalam ikan kaleng, mencapai
hingga suhu 130oC. molekul protein mengandung ikatan peptida dengan berat molekul yang besar,
dengan demikian jumlah asam pengoksidasi yang dibutuhkan sekitar 7ml.

Keberadaan mineral dapat mempengaruhi kualitas makanan. Tembaga (Cu) dan besi (Fe)
berperan dalam reaksi oksidasi-reduksi karena oksidasi berpotensi dapat menyebabkan perubahan
yang tidak diinginkan dalam makanan olahan yang terkait dengan oksidasi lemak yang mendorong
terjadinya penurunan nilai gizi pada makanan. Dalam proses oksidasi lemak/minyak, Fe dan Cu
mampu mempercepat proses inisiasi, propagasi, dan terminasi (bertindak sebagai katalis). Selain
itu, valensi Fe menentukan pigmen warna mioglobin dan hemoglobin dalam daging segar; ferric
(+3 keadaan valensi) menyebabkan warna coklat, sedangkan besi (+2 keadaan valensi)
menyebabkan warna merah. Reaksi Fe dengan belerang dalam makanan kaleng juga menginduksi
pembentukan kompleks besi sulfida (FeS), yang berupa suatu padatan hitam.
Pencoklatan enzimatis juga terkait dengan keberadaan Cu. Mineral ini berfungsi sebagai
kofaktor dari enzim polifenol oksidase (PPO), sebagai katalis untuk reaksi pencoklatan non-
enzimatik dan oksidasi asam askorbat. Zn dapat menstabilkan warna kacang hijau kalengan. ZnO
dapat digunakan sebagai agen pelapis pada makanan kaleng yang mampu mencegah terbentuknya
FeS selama proses pemanasan. Kadar Fe, Zn, dan Cu yang diperoleh melalui metode standar dan
modifikasi dapat dilihat pada Tabel 12.
Berdasarkan hasil statistik, ditemukan bahwa data yang diperoleh dari kedua metode tidak
berbeda secara signifikan, terbukti dengan nilai p > = 5% untuk semua sampel yang diuji.
Dari segi waktu dan biaya, metode yang dimodifikasi terlihat lebih disukai daripada metode
standar. Hidrogen peroksida dibutuhkan lebih sedikit dalam metode yang dikembangkan saat ini
(1 mL untuk setiap sampel) daripada metode standar (2 mL untuk setiap sampel). Perlu diingat
bahwa hidrogen peroksida tiga kali lebih mahal dibandingkan dengan asam nitrat. Selain itu,
metode yang dimodifikasi membutuhkan waktu proses yang lebih singkat (30 menit) daripada
metode standar (45 menit).
Kesesuaian pelarut alternatif yang digunakan juga diamati, dimana pada percobaan ini
hanya diuji untuk susu bubuk full cream. Uji kelarutan dilakukan untuk mengevaluasi pengaruh
pelarut terhadap hasil analisis. Dalam hal ini, 0,05M HNO 3 digunakan sebagai pengganti air
demineralisasi; dengan demikian kondisi sampel mirip dengan larutan standar. Selanjutnya
penambahan asam ke dalam larutan dapat berkontribusi memicu pengendapan dan mencegah
penyerapan pada dinding kaca. Kelarutan mineral terlihat memiliki hubungan positif dengan
lamanya paparan asam nitrat. Hasil penelitian menunjukan bahwa jenis pelarut secara signifikan
mempengaruhi hasil pengujian (p> = 0,05) (Tabel 13). Hal ini menunjukan bahwa penggantian
pelarut dengan 0,05 M HNO3 tidak dapat diterima.
Uji kinerja F-AAS bertujuan untuk memastikan bahwa instrumen layak digunakan dalam
proses analisis untuk menghasilkan data yang valid. Jumlah atom yang terdistribusi dalam nyala
berbeda, sehingga perlu untuk dilakukan penentuan ketinggian burner yang berbeda-beda untuk
setiap jenis mineral.

KESIMPULAN
Modifikasi metode pada proses destruksi untuk analisis kandungan Fe,Zn dan Cu pada
berbagai matriks makanan (kacang polong kalengan, susu bubuk full cream, dan ikan kalengan)
tidak berbeda secara signifikan dari AOAC (2012) dan telah memenuhi persyaratan penerimaan
untuk digunakan pada laboratorium analisis. Hasil analisis mineral yang diperoleh dari modifikasi
metode destruksi sistem dalam untuk mineral Fe,Zn, dan Cu. Untuk kacang polong kalengan 2,87;
7,02; 1,36 mg/kg. untuk susu full cream bubuk 24,60; 24,54; 0,86 mg/kg, dan untuuk ikan kalengan
23,44; 9.47; 0,99 mg/kg.
348 Indones. J. Chem., 2020, 20 (2), 348 - 359

Validation of Mineral (Fe, Zn, and Cu) Analysis Methods in Carbohydrate, Protein and
Fat-Rich Samples Using Microwave Digestion Method

Didah Nur Faridah1,*, Imas Solihat1,2, and Nancy Dewi Yuliana1


1
Department of Food Science and Technology, Faculty of Agricultural Technology, Bogor Agricultural University,
Jl. Tanjung, Kampus IPB Dramaga, Bogor 16680, Indonesia
2
AKA Bogor Polytechnic, Jl. Pangeran Sogiri No. 283, Tanah Baru, Bogor 16154, West Java, Indonesia

* Corresponding author: Abstract: Destruction process using both dry and wet conditions serves as a significant
step in mineral quantification. Wet destruction using microwave-assisted digestion refers
email: didah_nf@apps.ipb.ac.id
to the standard method of AOAC (2012); in this work, we modified the destruction
Received: December 31, 2018 procedures provided in the microwave digestion manual book, including temperature,
Accepted: May 23, 2019 reagent volume (HNO3 and H2O2), and length of destruction. The experiment works
DOI: 10.22146/ijc.42297 aimed (1) to validate the procedures in destruction process using microwave digestion; (2)
to quantify Fe, Zn and Cu in various food matrices (canned peas, canned fish, full cream
powdered milk) using a validated method. The method was evaluated according to
linearity, accuracy, precision, absolute and relative LOD, LOQ, and intra-reproducibility.
Measurement of Fe, Zn and Cu were considered to have a satisfying accuracy at a range
of 80–115%, with a good precision value (% RSD < 2/3 CV Horwitz), while regression
curves R2 > 0.995. The results showed that data collected from the modified method was
not significantly different compared to those from the AOAC method. The currently
developed method also fulfilled the acceptability requirements for laboratory analysis.

Keywords: minerals; F-AAS; method validation; microwave digestion

■ INTRODUCTION as being rapid, easy, and produce accurate assessment


[17-19], while also using less reagent.
The constituents of food, including water,
In our experiment, the destruction method using
carbohydrates, proteins, vitamins, fats, and minerals, may
microwave-assisted digestion is found to be more
affect the properties of the food. Specifically, essential
efficient in regards to the processing time, amount of
minerals (Fe, Zn, and Cu) [1-5] are closely related to some
solvent, and temperature, and regarded as a more
disadvantageous effects on food such as rancidity,
renewable technique in comparison with the method
changes in color, taste, and aroma. Flame-AAS is
prescribed by AOAC 2012. Based on the SOP for
commonly used for measuring mineral levels [6-10], in
microwave digestion, destruction procedures may vary
which the destruction process is considered as one of the
depending on the types of food matrices. In contrast, the
most important steps [11], often carried out in dry and
procedure method of AOAC 2012 is applied for all types
wet conditions [12-14]. Dry destruction accounts for the
of sample, regardless of food matrices. The results of the
loss of some mineral components in the furnace, while it
analysis between the AOAC method and the modified
is also a time-consuming process [15-16]. Meanwhile,
method were not significantly different, resulting in a
open system wet destruction contributes to
positive analysis gap.
contamination towards the environment, and certain
The method for mineral destruction in this study
minerals can be lost due to evaporation. On the other
which was derived from the manual of a microwave
hand, wet destruction with the closed system using
digestion device is different from the standard AOAC
microwave-assisted digestion has some advantages, such
method 999.10 (2012) [20]. The differences included

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 349

reagent volume, temperature, and length of destruction. The food samples (canned peas, full cream
Different food matrices should require a specific powdered milk, and canned fish) were collected from
treatment during the destruction process. For this reason, similar brands, expired dates, and production batches.
the food samples were selected according to the food The powdered milk was stored in airtight packaging at
matrix triangle (AOAC 2012), resulting in three main room temperature (quartering sampling method), while
groups of food, i.e. protein-rich food (canned fish), fat- canned peas and canned fish were stored in the freezer
rich food (full cream milk powder) and carbohydrate-rich (homogenization sampling method) (FDA 2014) [22].
food (canned peas). The selection of canned food samples Homogenized samples were placed in 10 polyethylene
aimed to reduce factors that can cause instability in the bags, and their ash content was analyzed in duplicate
method validation. The modified method was acceptable [23]. The homogeneity of these samples was checked
as long as it met the requirements after compared with the using different test criteria (F test) with the provision of
standard method [21]. Fcount < Ftable. Ash content test was achieved using the
The objective of this study was (1) to validate the standard method of AOAC 900.02A.
quantification method of Fe, Zn and Cu using F-AAS with
Instrumentation
microwave digestion, and (2) to obtain data on the
mineral content of Fe, Zn and Cu on several different food The instruments used were Close Microwave
matrices using the validated method. Digestion type SINEO MDS-6G, Flame-AAS type
Shimadzu AA 2000 and hollow cathode lamp of Fe, Zn,
■ EXPERIMENTAL SECTION and Cu. The F-AAS setting is presented in Table 1.
Materials Procedure
The chemicals used were analytical grade, including Sample digestion procedure
aquademineral, HNO3 (p) (Merck KgaA), H2O2 30% (Merck Sample (0.5 g dry basis was added with HNO3 (p)
KgaA), standard Zn solution of 1000 ppm (CertiPUR and H2O2 30%. The standard method setting of
Merck KgaA, no 1.19806), standard Fe solution of 1000 destruction and modification is presented in Table 2.
ppm (CertiPUR merge KgaA, no.1. 19781), standard Cu After completion of digestion, the sample was
solution of 1000 ppm (CertiPUR merge KgaA, no.1.1986). homogenized in demineralized water in a 25 mL flask.

Table 1. Setting of Flame-AAS


Wavelength Burner height Fuel flow Lamp current Slit bandwidth Sensitivity
Mineral Fuel
(nm) (mm) (L/min) (mA) (nm) Cu 2 ppm
Fe 248.3 Air-C2H2 9.0 2.2 4.0 0.2 0.0352
Zn 213.9 Air-C2H2 7.0 2.0 5.0 0.7
Cu 324.8 Air-C2H2 7.0 1.8 8.0 0.7

Table 2. The setting of microwave digestion


Sample Standard method Modified method
Full cream milk powder 130 °C/10 min/300 W
150 °C/5 min/600 W
180 °C/5 min/500 W
30 °C/3 min/300 W 200 °C/10 min/300 W
Canned peas 150 °C/5 min/600 W 130 °C /10 min/300 W
180 °C/22 min/500 W 150 °C/5 min/600 W
200 °C/15 min/300 W 180 °C/15 min/500 W
Canned fish 130 °C/10 min/300 W
150 °C/5 min/600 W
180 °C/15 min/500 W

Didah Nur Faridah et al.


350 Indones. J. Chem., 2020, 20 (2), 348 - 359

Control solution (comprising of nitric acid and hydrogen Cu), The theoretical LOQ is derived from [10 (sy/x)]/b.
peroxide) was made according to the same procedure as Confirmation of LOQ values was performed by making
the sample. a solution with a concentration of the theoretical LOQ
value and measured with seven replicates for each
Validation of analysis methods
The developed method was evaluated for its mineral (Fe, Zn, and Cu). The LDM (Limit detection of
acceptability according to the following indicators, i.e., method) value is determined by measuring seven
linearity, relative and absolute LOD, LOQ, precision, replications of the mineral concentration in the sample
accuracy, and intra-lab reproducibility. prepared according to the theoretical LDI values.
Precision test. This test is done by measuring the Data analysis
concentration of samples as much as 7 times replication for Data were expressed as mean, followed with
the measurement of 3 types of minerals (Fe, Cu, and Zn). deviation standard and relative standard deviation
Intra-lab reproducibility. Intra-lab reproducibility is (RSD). Furthermore, statistical analysis was used to
determined by measuring the concentration of samples in determine the p-value (independent t-test) in intra-lab
the same laboratory, by the same operator at different reproducibility. Results obtained from the modified
times. Each test was performed with three replications. method were compared to the standard method.
Furthermore, the RSD value of the analysis is compared
■ RESULTS AND DISCUSSION
with the RSD Horwitz value. Intra-lab reproducibility is
satisfactory if it has an RSD value smaller than the RSD Preparation of Sample
Horwitz value. Test results at two different weeks were Prior to destruction (by microwave digestion), the
tested using the t-test. sample size needs to be reduced, allowing to produce a
Accuracy. The test is performed using a sample spiked larger surface area and enhance the reaction rate.
with a standard solution. The experimental concentration Samples containing organic solvents, such as wine, must
of spiking is the percentage of the sample concentration be evaporated up to a fifth of the initial volume since the
that is 80% (low concentration), 100% (medium), and solvent can induce an explosion.
120% (high). Spiking is done at the beginning of the The selection of food matrix is based on major
preparation before the microwave digestion. Accuracy is components such as carbohydrate, protein, and fat,
determined by calculating percent recovery. arranged into the food matrix triangle (Fig. 1). The
Linearity. Instrument linearity was performed by proximate analysis was in accordance with the
measuring the absorbance of standard mineral solutions placements on literature studies (USDA 2006) [24],
(Fe, Zn, and Cu) at various concentrations. The linearity exhibited in Table 3.
test of Fe and Cu minerals used concentration 0–0.5 mg/L Furthermore, the water content in all samples (Table
while Zn mineral concentration was 0–1.0 mg/L. 4) was in the range of SNI and USDA standard. Moisture
Absorbance of each standard solution was measured and level in canned fish was found to be higher compared to
repetition was performed three times. literature USDA 15089, which might be ascribed to the
Limit of Detection (LOD) and Limit of Quantitation difference in the type of sample. The determination of
(LOQ). The values of LDI (Limit detection of instrument) water content was conducted using a vacuum oven
and LOQ can be obtained from linearity analysis data by method. Vacuum oven drying was conducted at low
determining the standard value of calibration error (sy/x). temperature and pressure, in order to retain volatile
Theoretical LDI is obtained from the formula [3 (sy/x)]/b. components present in the samples tested [25].
Confirmation was conducted by 7× repeated The homogeneity test is performed by measuring
measurement of the solution according to the theoretical ash content, enabling to ensure that the sample tested to
LDI concentration for 3 types of minerals (Fe, Zn and contain a comparable amount of mineral and does not

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 351

Fig 1. Food triangle

Table 3. Proximate composition based reference


Content (g/100g)
Sample
Water Protein Fat Carbohydrate Fiber Total sugar
Canned peas 85.13 3.01 0.48 10.60 3.3 3.20
Canned fish 66.86 20.86 10.45 0.54 0.1 0.43
Full cream powdered milk 2.47 26.32 26.71 38.42 0 38.42
Source: USDA 2016

Table 4. Water content


Water content (%)
Sample
Analysis result Reference
Full cream powdered milk 4.26 max 5 (SNI 2970-2015)
Canned peas 80.16 85.13 (USDA 11306)
Canned fish 76.54 66.86 (USDA 15089)

Table 5. Sample homogeneity test


Ash content (%)
Subject Full cream milk powder Canned peas Canned fish
1 (ai) (bi) 1 (ai) 2 (bi) 1 (ai) 2 (bi)
average 6.11 5.99 1.25 1.23 1.98 1.99
MSB 0.01 0.001 0.001
MSW 0.01 0.001 0.001
Fcount 0.96 0.87 1.38
Term of acceptance: Fcount < Ftable (3.02)

contain disturbing matrix (Table 5). Ash content can be indicated by % RSD < 2/3 CV Horwitz. This result
used as an illustration of the minerals contained in the displays that all random errors can be controlled. The
sample [26]. The results demonstrated that the F value random errors include instability of instrument
was 3.02, being lower than F table 5%, indicating the performance, changes in temperature, the difference in
acceptable homogeneity. techniques, and analysts.
Validation of the Methods LOD and LOQ
The indicators for detection limit are absolute LOD
Precision
Table 6 presents the satisfying precision for the (instrument LOD) and relative LOD (method LOD).
determination of Fe, Zn, and Cu in all samples, which was Absolute LOD represents the smallest concentration of

Didah Nur Faridah et al.


352 Indones. J. Chem., 2020, 20 (2), 348 - 359

Table 6. Precision test results


Mineral concentration (mg/kg)
Sample repeat Full cream powdered milk Canned peas Canned fish
Cu Zn Fe Cu Zn Fe Cu Zn Fe
average 0.87 24.34 24.72 1.35 7.01 2.82 0.95 9.42 23.33
SD 0.05 0.50 0.53 0.02 0.06 0.06 0.05 0.07 0.18
RSD (%) 6.79 2.07 2.14 1.63 1.63 0.89 4.89 0.76 0.77
2/3CV Horwitz 10.77 6.60 6.58 10.19 10.20 7.96 10.75 7.61 6.64
Term of acceptance: %RSD<2/3 CV Horwitz

analyte that can still be detected by the instrument, while reaching 0.07 mg/mL for Cu, 0.07 mg/mL for Fe, and
relative LOD is the smallest concentration of an analyte in 0.11 mg/mL for Zn. It means the methods have good
the sample that can still be detected and the results are sensitivity. The relative LOD is essential in dealing with
presented in Table 7. Limit detection testing was carried out constraints due to different signals/responses for each
by using a calibration curve with the formula 3[syx/b]. LOQ food matrix.
value constitutes the lowest limit of the calibration curve.
Linearity
The results for relative LOD can be seen in Table 8. The regression equation for Fe, Cu, and Zn was
The absolute LOD values for the three minerals examined presented as follows: y = 0.084x-0.0002 with R2 = 0.9970;
were smaller than those reported by a previous study [12],

Table 7. Absolute LOD and LOQ


Mineral Absolute LOD (mg/L) LOQ (mg/L)
Cu average 0.01 0.1
SD 0.0021 0.0032
RSD (%) 16.60 3.28
2/3CV Horwitz 5.14 15.18
Zn average 0.02 0.2
SD 0.0012 0.0047
RSD (%) 17.52 2.38
2/3CV Horwitz 22.70 13.64
Fe average 0.05 0.1
SD 0.0236 0.0081
RSD (%) 52.10 7.61
2/3CV Horwitz 17.00 22.40

Table 8. Relative LOD


Relative LOD (mg/L)
Mineral
Full cream powdered milk Canned peas Canned fish
Cu average 0.03 0.07 0.01
SD 0.0026 0.1578 0.0306
RSD (%) 8.77 212.18 307.43
2/3CV Horwitz 18.12 15.77 21.35
Zn average 0.1 0.04 0.06
SD 0.0853 0.0180 0.0042
RSD (%) 70.67 44.12 6.78
2/3CV Horwitz 14.66 17.26 16.23
Fe average 0.06 0.09 0.06
SD 0.0140 0.0173 0.0210
RSD (%) 23.65 18.88 35.31
2/3CV Horwitz 16.32 15.29 16.31

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 353

y = 0.1335x-0.0004 with R2 = 0.9969; y = 0.4125x + 0.0086 controlled, including the amount of nitric acid and
with R2 = 0.9977, respectively. As exhibited in Table 9, the hydrogen peroxide, temperature, sample weight, pre-
experimental result showed satisfying linearity, in which digestion, and destruction length. The optimization can
the higher concentration of tested mineral corresponded be achieved using various techniques, such as the
to the rise of absorbance. factorial design of Plackett-Burman. However, this step
The intercept for Fe, Zn, and CU reached -0.0002, was excluded, considering that all procedures in this
0.0086 and -0.0004, respectively, suggesting the occurrence work are based on the standard of microwave digestion
of systematic errors in F-AAS instrument. Furthermore, instrument for each food matrix. The efficiency of the
the R2 value was >0.990, which is in accordance with the destruction process using microwave digestion is
acceptance limit, as depicted in Fig. 2. influenced by the volume of acid substances (HNO3 and
H2O2), time, and temperature [28].
Intra-lab reproducibility
Table 10 presents the intra-lab reproducibility, Power setting in both methods (standard and
showing p-value > α 5% for all minerals in canned fish, modified) was adopted from the manual book, while
canned peas and full cream powdered milk, suggesting
Table 9. Linearity test results
that the value did not significantly differ [27].
Mineral Concentration (ppm) Average (abs)
Accuracy Fe 0.1 0.0092
Accuracy represents the method efficiency based on 0.2 0.0152
0.3 0.0245
recovery from standard solutions added to the sample
0.4 0.0332
(spiking). The results showed that the accuracy of Fe, Zn, 0.5 0.0425
and Cu levels varied from 80 to 100%, which is still in the Zn 0.2 0.0954
range of acceptability limit of AOAC (75–120%). This 0.4 0.1817
0.6 0.2730
indicates that the procedure in the modified method is
0.8 0.3421
acceptable for all minerals. Result of the recovery test can 1.0 0.4237
see in Table 11. Cu 0.1 0.0121
0.2 0.0257
Mineral Analysis 0.3 0.0422
To optimize the results, some conditions during the 0.4 0.0516
0.5 0.0665
destruction phase in microwave digestion need to be

Fig 2. Linear regression curve for quantification of Fe, Zn, and Cu

Didah Nur Faridah et al.


354 Indones. J. Chem., 2020, 20 (2), 348 - 359

Table 10. Intra-lab reproducibility


Content (mg/kg)
Repeat Cu Zn Fe
1 2 1 2 1 2
Full cream milk powder
average 0.94 0.87 24.34 24.48 24.72 25.01
SD 0.06 0.05 0.50 0.73 0.53 1.18
% RSD 6.78 5.78 2.07 2.99 2.14 4.70
2/3 CV Horwitz 16.15 16.33 9.89 9.88 9.87 9.86
p-value 0.110 0.712 0.545
Canned peas
average 1.36 1.34 7.01 6.88 2.82 2.80
SD 0.02 0.02 0.06 0.12 0.09 0.10
% RSD 1.51 1.52 0.89 1.73 3.45 3.59
2/3 CV Horwitz 15.28 15.31 11.94 11.97 13.69 13.70
p-value 0.120 0.089 0.668
Canned fish
average 0.95 0.95 9.42 9.48 23.33 23.24
SD 0.05 0.009 0.07 0.07 0.18 0.10
% RSD 4.89 1.02 0.76 0.77 0.77 0.43
2/3 CV Horwitz 16.12 16.12 11.41 11.40 9.96 9.96
p-value 0.821 0.210 0.355
Term of acceptance: p-value > α 95%

Table 11. Results of the recovery test


Recovery (%)
Spiking Full cream powdered milk powder Canned peas Canned fish
minerals 80% 100% 120% 80% 100% 120% 80% 100% 120%
Fe 94.52 85.19 85.23 92.16 89.78 82.05 92.28 85.53 84.14
Zn 101.21 91.23 100.76 91.66 81.79 86.93 90.45 97.28 99.43
Cu 84.15 94.64 83.18 87.48 85.46 82.05 92.68 82.11 89.52
Term of acceptance: 80–115%

destruction temperature was carried out according to the exposed to non-ionic electromagnetic radiation, which
standard of AOAC (2012). The sample weight (0.5 g, dry induces rotational dipole moments without changing
weight) was also based on the arrangement of AOAC. their molecular structure [29]. The effect of the wave in
Destruction process allows changing the form of microwave appears magnetically [30] and dielectrically
inorganic constituents to organic constituents, in which it [31]. Dielectric material comprised of molecules
involves the conversion of organic elements such as C, H, obtaining a dipole moment or a neutral charge (having
O, N, S and P into CO2, H2O, NO2, while inorganic a balanced positive and negative charge). Additionally,
elements still exist in their original form. Open the dielectric wave in the microwave is due to the
destruction using microwave digestion promotes several capability of storing electric charges at different
advantages, including faster time operation, fewer potentials. Water is an electrical molecule, possessing
reagents used, higher accuracy, thereby reducing hydrogen atoms (positively charged) and oxygen atoms
contamination in samples and loss of volatile elements. (negatively charged). The exposure of the electrical field
During the microwave-assisted process, molecules are to dielectric materials led to changes in the direction of

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 355

the current because of repulsion and attraction within The dominant fiber contents in peas are stachyose
positive charges or negative charges; thus, the wave occurs (19.83 mg/g) and verbascose (16.10 mg/g) [40]. In
magnetically. More heat was produced in samples addition, canned peas contain simple sugars, including
containing more water than in those with less water. The glucose, fructose, sucrose, and galactose; hence, the
frequency in the microwave ranged from 300 to 300.000 destruction process starts from 130 °C to 180 °C. Addition
MHz with a wavelength of 1 mm–1 m. of nitric acid (6 mL) is in accordance with a previous
Samples in microwave-assisted destruction are mixed work of Bakkali et al. on vegetable samples, yielding the
with strong acids in a closed vessel. In that system, pressure highest accuracy volume [6].
(1.1–2.2 Mpa) and temperature (30–190 °C) increased due Full cream powdered milk is rich in fat. The
to microwave radiation, allowing to convert the solid form covalent bond between fatty acids is difficult to break
into a liquid. High pressure and temperature (around down, while fat has a low relative potential of oxidation,
200 °C) [32] on the vessel is able to raise the boiling point which means that there is a need to perform destruction
of nitric acid. Thus, it strengthens oxidation and accelerates at temperature up to 200 °C with a greater amount of
the reaction. HNO3 and H2O2 are used as oxidizing agents oxidizing acid (10 mL). Lactose (D-galactose and D-
because they are capable to oxidize minerals into their salt glucose) constitutes a major simple sugar in milk [41].
form. Nitric acid at high temperature is decomposed into Therefore, the initial temperature of digestion begins at
water-soluble oxides such as NO and NO2. The use of 130 °C. Such destruction process was previously carried
mixed acids is better than a single acid for the mineral out by Belete et al. [42] in cow milk samples, using the
destruction process; the destruction process is incomplete destruction temperature of 165–190 °C following a 10 min
if nitric acid is the only used compound, resulting in a pre-destruction phase.
cloudy solution [33]. The addition of H2O2 leads to Canned fish represents a food sample with high
intensifying the presence of O2 in the vessel [34]. The protein content. The main protein content in fish, i.e.,
oxidizing mixture consisting of HNO3 and H2O2 produces actin and myosin, is susceptible to denaturation when
a greater % recovery than a mixture of HNO3 and HCl, exposed to microwaves (Fig. 4) [43]. Although these
and a mixture of HNO3 and H2SO4 [35]. myofibrillar proteins start to denature at about 85 °C, a
The gradient or multilevel heating system aims to higher temperature was needed to induce denaturation
intensify the reaction between NO and O2 to form NO2 when they exist in canned fish, reaching up to 130 °C.
[36]. The NO2 is reabsorbed in acidic solution, and it Protein molecules contain peptide bonds with a large
regenerates HNO3 [37]. The reaction process in the vessel molecular weight; thus, the amount of oxidizing acid
is described in Fig. 3 [38]. needed is 7 mL.
The longer heating process in the microwave would
lead to a higher temperature in the vessel, responsible for
attenuation of the condensation process and eventually
increase the mineral level in the solution. A complete
destruction process is marked by the formation of a clear
solution, which indicates the reformation of an organic
compound. Sample solutions are then ready to be
analyzed using F-AAS. F-AAS is recommended for use in
measuring minerals because of its good sensitivity with a
detection limit of more or less 1 ppm and also its low
operational cost.
Soluble dietary fiber is a carbohydrate-based Fig 3. The reaction process in the microwave digestion
polymer widely found in nuts, fruits, and vegetables [39]. vessel

Didah Nur Faridah et al.


356 Indones. J. Chem., 2020, 20 (2), 348 - 359

The presence of minerals can affect the quality of stabilize the color of canned green beans. ZnO can be
food. Copper (Cu) and iron (Fe) play a role in oxidation- used as a coating agent on canned foods capable of
reduction reactions because the oxidation potential can preventing the formation of FeS during the heating
cause unwanted changes in processed foods related to fat process. The levels of Fe, Zn, and Cu obtained through the
oxidation which promotes a decrease in the nutritional standard and modified methods can be seen in Table 12.
value of food [44]. In the process of oxidation of fat/oil, Fe Statistically, we found that data obtained from both
and Cu are able to accelerate the process of initiation, protocols did not differ significantly, as evidenced by p-
propagation, and termination (acting as a catalyst). In value > α = 5% for all samples examined.
addition, the valence of Fe dictates the color pigment of In terms of time and cost, the modified method
myoglobin and hemoglobin in fresh meat; ferric (+3 seemed to be more preferable than the standard method.
valence state) accounts for brown color, while ferrous (+2 The hydrogen peroxide was less needed in the currently
valence state) accounts for red color. The reaction of Fe developed method (1 mL for each sample) than the
with sulfur in canned food also induces the formation of standard method (2 mL for each sample). It is noteworthy
iron sulfide (FeS) complex, which is a black solid. that hydrogen peroxide is three times more expensive
Enzymatic browning is also linked to the presence of than nitric acid. Additionally, the modified method
Cu. This mineral serves as a cofactor of the enzyme required a shorter time of the process (30 min) than the
polyphenol oxidase (PPO), as a catalyst for non-enzymatic standard method (45 min).
browning reactions and oxidation of ascorbic acid. Zn can

Fig 4. The process of actin and myosin denaturation [42]

Table 12. The result of analysis mineral with standard and modification method
Standard method (mg/kg) Modified method (mg/kg)
Samples
Fe Zn Cu Fe Zn Cu
Full cream powdered milk 24.99 24.62 0.85 24.60 24.54 0.84
SD 0.80 0.13 0.01 0.20 0.19 0.02
% RSD 3.21 0.51 1.69 0.81 0.76 2.95
2/3 CV Horwitz 6.57 6.59 10.92 6.59 6.59 10.95
p-value 0.575 0.714 0.561
Canned peas 2.88 7.11 1.31 2.87 7.02 1.36
SD 0.06 0.03 0.02 0.04 0.04 0.01
% RSD 2.11 0.44 1.35 1.24 0.57 0.99
2/3 CV Horwitz 9.10 7.94 10.23 9.10 7.95 10.18
p-value 0.827 0.222 0.09
Canned fish 23.44 9.78 1.04 23.44 9.47 0.99
SD 0.23 0.24 0.03 0.23 0.05 0.03
% RSD 0.98 2.43 2.50 0.98 0.49 2.70
2/3 CV Horwitz 6.63 7.57 10.60 6.63 7.60 10.67
p-value 0.245 0.211 0.262

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 357

Table 13. Solvent alternative test


Mineral content (mg/kg)
Cu Zn Fe
Parameters
Demineralized HNO3 Demineralized HNO3 Demineralized HNO3
water 0.05M water 0.05M water 0.05M
average (mg/kg) 0.94 1.25 24.34 25.88 24.72 28.86
SD 0.06 0.08 0.50 1.34 0.53 0.43
% RSD 6.78 6.10 2.07 5.18 2.14 1.50
CV Horwitz 16.15 15.48 9.89 9.80 9.87 9.646
p-value 0.000 0.028 0.000

The suitability of solvent alternative was also for financial support and research facilities of this work.
observed, in which this experiment was only tested for full
■ REFERENCES
cream powdered milk. Toughness test was carried out to
evaluate the effects of the solvents on analytical results. In [1] Lazarte, C., Carlsson, N.G., Almgren, A., Sandberg,
this case, 0.05M HNO3 was used instead of demineralized A.S., and Granfeldt, Y., 2015, Phytate, zinc, iron,
water; thus the sample condition was similar to the and calcium content of common Bolivian food and
standard solution. Furthermore, the addition of acid to implications for mineral bioavailability, J. Food
the solution can contribute towards igniting and Compos. Anal., 39, 111–119.
preventing precipitation and absorption on the glass wall. [2] Gómez-Nieto, B., Gismera, M.J., Sevilla, M.T., and
The solubility of mineral seemed to possess a positive Procopio, J.R., 2017, Determination of essential
relation to the length of exposure to nitric acid [45]. The elements in beverages, herbal infusions and dietary
results showed that the type of solvent significantly altered supplements using a new straightforward
the test results (p < 0.05) (Table 13). This suggests that sequential approach based on flame absorption
replacement with 0.05 M HNO3 was unacceptable. spectrometry, Food Chem., 219, 69–75.
The F-AAS performance test aims to ensure that the [3] Zhu, F., Fan, W., Wang, X., Qu, L., and Yao, S.,
instrument is feasibly used in the analysis process to 2011, Health risk assessment of eight heavy metals
produce valid data. The number of atoms distributed in in nine varieties of edible vegetable oils consumed
the flame is different, so it is necessary to determine the in China, Food Chem. Toxicol., 49 (12), 3081–3085.
different burner height for each mineral. [4] Tuzen, M., 2009, Toxic and essential trace elemental
contents in fish species from the Black Sea, Turkey,
■ CONCLUSION
Food Chem. Toxicol., 47 (8), 1785–1790.
Modification methods in the process of destruction [5] Antoine, J.M.R., Fung, L.A.H., Grant, C.N., Dennis,
for analysis of Fe, Zn, and Cu in various food matrices H.T., and Lalor, G.C., 2012, Dietary intake of
(canned peas, full cream powdered milk, and canned fish) minerals and trace elements in rice on the Jamaican
were not significantly different from AOAC (2012) and has market, J. Food Compos. Anal., 26 (1-2), 111–121.
fulfilled the acceptability requirements for use in laboratory [6] Bakkali, K., Martos, N.R., Souhail, B., and
analysis. Mineral analysis result obtained by the modified Ballesteros, E., 2009, Characterization of trace
method in the process of destruction for Fe, Zn and Cu metals in vegetables by graphite furnace atomic
minerals for canned peas: 2.87; 7.02;1.36 mg/kg, full absorption spectrometry after closed vessel
cream powdered milk 24.60; 24.54; 0.86 mg/kg, canned microwave digestion, Food Chem., 116 (2), 590–594.
fish: 23.44; 9.47; 0.99 mg/kg. [7] Helaluddin, A.B.M., Khalid, R.S., Alaama, M., and
Abbas, S.A., 2016, Main analytical techniques used
■ ACKNOWLEDGMENTS
for elemental analysis in various matrices, Trop. J.
Authors would like to thank AKA Bogor Polytechnic Pharm. Res., 15 (2), 427–434.

Didah Nur Faridah et al.


358 Indones. J. Chem., 2020, 20 (2), 348 - 359

[8] Sharma, B., and Tyagi, S., 2013, Simplification of samples by ICP-OES, Food Chem. Toxicol., 49 (11),
metal ion analysis in fresh water samples by atomic 2800–2807.
absorption spectroscopy for laboratory students, J. [18] Santos, J., Oliva-Teles, M.T., Delerue-Matos, C.,
Lab. Chem. Educ., 1 (3), 54–58. and Oliveira, M.B.P.P., 2014, Multi-elemental
[9] Lajunen, L.H., and Perämäki, P., 2004, analysis of ready-to-eat ”baby leaf” vegetables using
Spectrochemical analysis by atomic absorption and microwave digestion and high-resolution
emission, 2nd Ed., The Royal Society of Chemistry, continuum source atomic absorption spectrometry,
Cambridge, UK, 6–8. Food Chem., 151, 311–316.
[10] Bader, N.R., and Zimmermann, B., 2012, Sample [19] Ranasinghe, P., Weerasinghe, S., and Kaumal,
preparation for atomic spectroscopic analysis: An M.N., 2016, Determination of heavy metals in
overview, Adv. Appl. Sci. Res., 3 (3), 1733–1737. tilapia using various digestion methods, IJSRIT, 3
[11] Korn, M.G.A., Morte, E.S., dos Santos, D.C.M.B., (6), 39–48.
Castro, J.T., Barbosa, J.T.P., Teixeira, A.P., Fernandez, [20] Association of Official Analytical Chemists
A.P., Welz, B., dos Santos, W.P.C., dos Santos, (AOAC), 2012, Official methods of analysis
E.B.G.N., Korn, M., 2008, Sample preparation for the agricultural chemicals, 10th Ed., Washington DC.,
determination of metals in food samples using (15), 237–242.
electroanalytical methods–A review, Appl. Spectrosc. [21] National Association of Testing Authorities
Rev., 43 (2), 67–92. (NATA), 2004, Guidelines for the validation and
[12] Silvestre, M.D., Lagarda, M.J., Farré, R., Martínez- verification of quantitative and qualitative test
Costa, C., and Brines, J., 2000, Copper, iron and zinc methods, Australia.
determination in human milk using FAAS with [22] Food and Drug Administration (FDA), 2014,
microwave digestion, Food Chem., 68 (1), 95–99. Elemental analysis manual for food and related
[13] Bragg, S.A., and Xue, Z.L., 2011, Optimization of dry product: Digestion and separation, Rockville
ashing of whole blood samples for trace metal Maryland, USA, 2–3.
analysis, Am. J. Anal. Chem., 2, 979–983. [23] Fiamegos, Y., Vahcic, M., Emteborg, H., Snell, J.,
[14] Idera, F., Omotola, O., Adedayo, A., and Paul, U.J., Raber, G., Cordeiro, F., Robouch, P., and de la Calle,
2015, Comparison of acid mixtures using conventional B., 2016, Determination of toxic trace element in
wet digestion methods for determination of heavy canned vegetables. The importance of sample
metals in fish tissues, JSRR, 8 (7), 1–9. preparation, TrAC, Trends Anal. Chem., 85, 57–66.
[15] Sneddon, J., Hardaway, C., Bobbadi, K.K., and [24] United States Department of Agriculture (USDA),
Reddy, A.K., 2006, Sample preparation of solid 2006, Canned Peas, Canned Fish, Full Cream
samples for metal determination by atomic Powdered Milk, USDA Publisher, United State.
spectroscopy-An overview and selected recent [25] De Melo, I.L.P., and de Almeida-Muradian, L.B.,
applications, Appl. Spectrosc. Rev., 41 (1), 1–14. 2011, Comparison of methodologies for moisture
[16] Elmastas, M., Can, M., Uzun, S., and Aboulenein, determination on dried bee pollen samples, Ciênc.
H.Y., 2005, Determination of copper zinc, cadmium, Tecnol. Aliment., 31 (1), 194–197.
and nickel in cows’, goats’, ewes’, and human milk [26] Mbatchou, V.C., and Dawda, S., 2013, The
samples using flame atomic absorption spectrometry nutritional composition of four rice varieties grown
(FAAS) microwave digestion, Anal. Lett., 38 (1), and used in different food preparations in Kassena-
157–165. Nankana district, Ghana, Int. J. Res. Chem.
[17] Altundag, H., and Tuzen, M., 2011, Comparison of Environ., 3 (1), 308–315.
dry, wet, and microwave digestion method for the [27] Solihat, I., Faridah, D.N., and Yuliana, N.D., 2018,
multi-element determination in some dried fruit Validation method of Flame-AAS with microwave

Didah Nur Faridah et al.


Indones. J. Chem., 2020, 20 (2), 348 - 359 359

digestion for mineral analysis in carbohydrate-rich 2015, Microwave-assisted diluted acid digestion for
samples, Molekul, 13 (2), 133–140. trace element analysis of edible soybean products,
[28] Momen, A.A., Zachariadis, G.A., Anthemidis, A.N., Food Chem., 175, 212–217.
and Stratis, J.A., 2007, Use of fractional design for [37] Ghriss, O., Amor, H.B., Jeday, M.R., and Thomas,
optimization of digestion procedures followed by D., 2019, Nitrogen oxides absorption into an
multi-element determination of essential and non- aqueous nitric acid solution containing hydrogen
essential elements in nuts using ICP-OES technique, peroxide tested using a cables-bundle contactor,
Talanta, 71 (1), 443–451. Atmos. Pollut. Res., 10 (1), 180–186.
[29] Belay, K., and Tadesse, A., 2014, Comparison of [38] Bizzi, C.A., Nóbrega, J.A., and Barin, J.S., 2014,
digestion methods for determination of Pb (II), Cr “Diluted Acids in Microwave-Assisted Wet
(VI) and Cd (II) contents in some Ethiopia spices using Digestion” in Microwave-Assisted Sample
atomic absorption spectroscopy, IJASR, 2 (3), 42–53. Preparation for Trace Element Determination, 1st
[30] Volkovskaya, I.I., Semenov, V.E., and Rybakov, K.I., Ed., Eds. de Moraes Flores, E.M., Elsevier, 179–204.
2017, Effective magnetic permeability of compacted [39] Chen, Y., Ye, R., Yin, L., and Zhang, N., 2014, Novel
metal powders at microwave frequencies, EPJ Web blasting extrusion processing improves the
Conf., 149–02008. physicochemical properties of soluble dietary fiber
[31] Rybakov, K.I., and Semenov, V.E., 2017, Effective from soybean residue and in vivo evaluation, J.
microwave dielectric properties of ensembles of Food Eng., 120, 1–8.
spherical metal particles. IEEE Trans. Microwave [40] Brummer, Y., Kaviani, M., and Tosh, S.M., 2015,
Theory Tech., 65 (5), 1479–1487. Structural and functional characteristic of dietary
[32] Ghanthimathi, S., Abdullah, A., Salmijah, S., Ujang, fibre in beans, lentils, peas and chickpeas, Food Res.
T., and Izzah, A.N., 2012, Comparison of microwave- Int., 67, 117–125.
assisted acid digestion methods for ICP-MS [41] Mourad, G., Bettache, G., and Samir, M., 2014,
determination of total arsenic in fish tissue, Sains Composition and nutritional value of raw milk,
Malays., 41 (12), 1557–1564. Issues Biol. Sci. Pharm. Res., 2 (10), 115–122.
[33] Yang, L., Yan, Q., Cao, Y., and Zhang, H., 2012, [42] Belete, T., Hussen, A., and Rao, V.M., 2014,
Determination of mineral elements of some coarse Determination of concentration of selected heavy
grains by microwave digestion with inductively metals in cow’s milk: Borena zone, Ethiopia, J.
coupled plasma atomic emission spectrometry, E-J. Health Sci., 4 (5), 105–112.
Chem., 9 (1), 93–98. [43] Liu, S., Yu, X., Fukuoka, M., and Sakai, N., 2014,
[34] Nóbrega, J.A., Pirola, C., Fialho, L.L., Rota, G., de Modeling of fish boiling under microwave
Campos Jordão, C.E.K.M.A., and Pollo, F., 2012, irradiation, Int. J. Food Eng., 140, 9–18.
Microwave-assisted digestion of organic samples: [44] Damodaran, S., Parkin, K.L., and Fennema, O.R.,
How simple can it become?, Talanta, 98, 272–276. 2007, Fennema’s Food Chemistry, 4th Ed., CRC
[35] Demirel, S., Tuzen, M., Saracoglu, S., and Soylak, M., Press, Boca Raton, Florida, US.
2008, Evaluation of various digestion procedures for [45] Mohammed, E., Mohammed, T., and Mohammed,
trace element contents of some food materials, J. A., 2017, Optimization of an acid digestion
Hazard. Mater., 152 (3), 1020–1026. procedure for the determination of Hg. As. Sb. Pb
[36] Barbosa, J.TP., Santos, C.M.M., Peralva, V.N., Flores, and Cd in fish muscle tissue, MethodsX, 4, 513–523.
E.M.M., Korn, M., Nóbrega, J.A., and Korn, M.G.A.,

Didah Nur Faridah et al.