SPESIFIKASI DPLyte

Sample Whole Blood, serum, Plasma or Urine

Sample Size 95 uL

Linearity Range B/P/S/Q Na : 40 – 200 mmol/L


K : 1.7 - 15 mmol/L
Cl : 50 - 200 mmol/L

Urine Na : 3 - 300 mmol/L


K : 5 - 120 mmol/L
Cl : 15 – 300 mmol/L
Correlation coefficients were all greater

Analysis Time 60 sec without printout ( 60 tests per hour )


80 sec with printout ( 45 tests per hour )

Quality Control QC measurement results within the target value range less than

Washing System Analyzer will take approximately one minute to flush all line with wat

Data Storage 1000 patients result


500 QC Results : Level 1
500 QC Results : Level 2
500 QC Results : Level 3

Calibration 2 points or 3-points calibration every 4 hours


1 point calibration after each sample

Reproducibility Within run ( n=3 )


Blood, Ser um, Plasma
Na : CV ≤ 1% ( 120 – 160 mmol/L )
K : CV ≤ 2% ( 2.5 - 6 mmol/L )
Cl : CV ≤ 2% ( 85 - 130 mmol/L )

Urine
Na : CV ≤ 5% ( 100 – 250 mmol/L )
K : CV ≤ 5% ( 10 - 60 mmol/L )
Cl : CV ≤ 5% ( 100 - 250 mmol/L )
Ambient Conditions 15-32o C, room temperature ( 60 – 90 o F )
< 85 % humidity

Reagent Control RFID Reagent Kit

Interference Albumin and Bicarbonate can also cause a negative bias in chlorid
Other ion such as Thiocyanate and Bromide can also cause a po
in Chloride ion

User Input Keypad External Keyboard


External barcode Scanner (optional)

Languages Bahasa Indonesia, English, Spanish, French, German, italian, Portugu

Size & Weight 13.2 x 12.4 x 12 inches or 335 x 315 x 295 mm


14 lbs or 6 kgs
et value range less than 2 SD

e to flush all line with water


e a negative bias in chloride ion.
mide can also cause a positive bias

h, German, italian, Portuguese