Instrumentasi
FF-UNAIR 2016
Mochammad Yuwono
Fakultas Farmasi Universitas Airlangga
yuwono05@yahoo.com
HPLC Instrumentation
1. Mobile Phase Supply System
2. Pump
3. Sample Injector
4. Column
5. Detector
6. Data System
Gradient
Controller
•
Pump Column
Detector
Injector
Mobile
Phases
Instrumentation
HPLC Instrumentation
Instrumentation
HPLC Instrumentation
Degasser
70 mbar
A A
B B
C C
Pump
A A B C
A B C
B
C
Fase gerak
diletakkan dalam
botol-botol
reservoir.
Fase Gerak
Piston
Cam
Seal
Inlet
Check Valve
PENYEBAB GANGGUAN POMPA
Seal aus, menyebabkan kebocoran fasa
gerak ke bagian belakang pompa
Kontaminasi pada Check valve akibat partikulat
atau kotoran
Gelembung udara yang berpengaruh
pembukaan dan penutupan check valve
Pengatasan:
• Ganti Seal secara berkala
• Cuci sistem sebelum dimatikan
• Lakukan Degassing Fase gerak
HPLC, Mengapa perlu tekanan tinggi?
Pada HPLC digunakan ukuran partikel packing
kolom (silika) yang sangat kecil, umumnya 3-10um.
Faktor C pada hukum van Deemter sebanding
dengan dp2
Ukuran partikel makin kecil luas permukaan
makin besar transfer massa dari fase gerak ke
fase diam atau sebaliknya makin besar kolom
semakin efisien dan resolusi peak semakin baik.
Penggunaan ukuran partikel kecil diperlukan
tekanan tinggi
Pada kecepatan alir 0.5 sampai 5 ml/menit, panjang
kolom 10-30 cm dibutuhkan tekanan 70-400 atm
(1000 – 6000 psi)
Van Deemter Plot (smaller particles have smaller A & C terms)
Particle Size and Column Performance
(300 psi)
(1300 psi)
(10,000 psi)
(34,000 psi)
HPLC vs UHPLC/UPLC
Load
Inject
Sample Injection
Load and Inject -Position
INJEKTOR
Autosampler:
Sampel dimasukkan dengan
syringe otomatis yg
mengambil sampel dari vial
pada rak bermotor
penggerak
HPLC Columns
Kolom:
Bagian
terpenting
dalam
pemisahan
HPLC Column Categories
Column Hardware: Standard or Cartridge,
stainless, titanium
Chromatographic Modes: Normal-phase
(NPC), reversed-phase (RPC), ion-exchange
(IEC), size exclusion (SEC)
Dimensions: Preparative, semi-prep,
analytical, fast LC, micro, nano
Support Types: Silica, polymer, zirconia,
hybrid
Column Dimension
Length x ID
Art.-Nr. xxxx HPLC-column 250x3 mm
Lichrospher 100 RP-18 5µm
Flow Particle size in µm
Modification of silicagel
Flow direction
Chromatography Stationary Phases
Where R = C18H37
hydrocarbon chain
(octadecylsilyl deriv.
silica or “C18”)
33
Kolom HPLC
Bentuk Partikel
Spherical atau irregular
Partikel spherical
mengurangi tekanan dan kolom
lebih awet
Ukuran Partikel
Umumnya 3-10µm
Makin kecil ukuran partikel,
makin tinggi efisiensi
KARAKTERISTIK KOLOM HPLC
Carbon Load
Amount of bonded phase attached to base material,
expressed as %C
Higher carbon loads generally offer greater resolution
and longer run times.
Low carbon loads shorten run times and many show a
different selectivity.
Kolom HPLC
Ukuran Pori-pori
Bervariasi antara 60-
10,000Å
Makin besar pori-pori
analit molekul besar
dapat tertahan lebih lama.
Pilih 150Å untuk sampel
BM 2000.
Kolom Monolitik
irregular
spherical
monolithic
40
High Pure Silica
Similar to NP-HPLC
but with aqueous mobile phase
42
UV-Vis Detector
Photodiode array Detector (PDA = DAD)
Refractive Index Detector (RID)
Fluorescence Detector (FLD)
Evaporative Light Scattering Detector (ELSD)
Electrochemical Detector (ECD)
Conductivity Detector
Radiometric Detector
Hyphenated Systems
LC-MS; LC-MS/MS; LC-NMR
HPLC Detectors
UV-Vis
Good detection limits (0.1-1 ng; linear range 105)
Can be used with gradient elution
Requires chromophore
Refractive Index
Poor detection limits (100-1000 ng; linear range 102)
Isocratic only
Nearly universal detection
Evaporative Light Scattering
Good detection limits (0.1-1 ng; linear range 105);
Non-linear calibration required
Can be used with gradient elution
Nearly universal detection
UV-Vis Detector
Paling banyak digunakan
Tersusun dari Lampu deuterium, monokromator dan
flow cell
Monokromator terdiri dari grating atau prisma yang
dapat bergerak dan dapat memilih panjang
gelombang spesifik yang masuk melalui celah (slit)
190 – 600nm
Ditambah lampu tungsten untuk meningkatkan
sensitivitas pada daerah visibel.
Macam UV-Vis:
Fixed wave length
Variable wave length
Multiple variable wave length
Photo diode-array
UV-Vis Detector
Lamp Cut-off filter
Holmium oxide filter
Slit
Sample diode
Mirror 1
Grating
Flow cell
Mirror 2
Reference diode
Photodiode array Detector (PDA)
Disebut juga diode
array detector (DAD)
Dapat menampilkan
spektra UV/Vis dari
puncak yang terelusi
Mampu melakukan
identifikasi puncak
PDA
Dengan menggunakan software
dapat menampilkan kromatogram dan
spektra serapan UV/Vis sampel
Dapat mengenali panjang gelombang
maksimum, peak matching
(menghitung match factor atau MF),
library search dan evaluasi kemurnian
peak (peak purity)
Dapat menampilkan 3-D spektra
Spektra 3D
HPLC-DAD
Spektra UV/Vis analit dan baku pembanding
overlay
Evaluasi korelasinya (r, MF)
53
Memilih Panjang Gelombang Detektor UV/PDA
Pemilihan
Panjang
Gelombang
(UV)
?
Refractive index
Mengukur perubahan indeks refraksi antara sel
sampel yang mengandung analit yang terelusi
dengan sel pembanding.
Sensitivitas lebih rendah (0,01-0,1 ug)
Dapat mendeteksi hampir semua senyawa dan
digunakan umumnya untuk analit yang memiliki
gugus kromofor lemah seperti gula, trigliserida,
asam-asam organik dll.
Detektor standar untuk GPC
Tidak dapat digunakan untuk gradien
Refractive Index Detector
Evaporative Light Scattering Detector