KAJIAN KESELAMATAN RADIASI

RADIOLOGI DIAGNOSTIK
002/PPR/GMP/2022

PT. Global Medik Persada


Jl. Raya Kb. Jeruk No.26, RT.9/RW.2, Kb. Jeruk, Kec.
Kb. Jeruk, Kota Jakarta Barat, Daerah Khusus Ibukota
Jakarta 11530
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 1 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

LEMBAR PENGESAHAN

Tanda
Nama Jabatan Tanggal
Tangan

Ramadhani Dwi 18 Agustus


Staff UKES
Nugraha 2022
Disiapka 27826221
n
Oleh
Slamet Raharjo 18 Agustus
PPR Medik 1
615032017 2022

Disahkan
Suprianto Direktur 23 Agustus
Oleh
02012017 Keuangan 2022

Disahkan
Daniel Șoeprianto 24 Agustus
Oleh Direktur Utama
01072014 2022

Nomor 002/PPR/GMP/2022
Tanggal 24 Agustus 2022
KAJIAN KESELAMATAN RADIASI
RADIOLOGI DIAGNOSTIK Revisi 01

Halaman 1 s/d 24
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 2 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

DAFTAR ISI

HALAMAN JUDUL.................................................................................................................................1
LEMBAR PENGESAHAN.......................................................................................................................1
DAFTAR ISI............................................................................................................................................2
BAB I PENDAHULUAN..........................................................................................................................3
1.1 Latar Belakang.................................................................................................................3
1.2 Tujuan..............................................................................................................................3
1.3 Ruang Lingkup.................................................................................................................4
1.4 Definisi..............................................................................................................................4
1.5 Referensi..........................................................................................................................6
BAB II DESKRIPSI PESAWAT SINAR-X DAN PROSEDUR PENGGUNAANNYA...............................7
2.1 Data Pesawat Sinar-X......................................................................................................7
2.2 Prosedur Pengoperasian Pesawat Sinar-X......................................................................8
BAB III DENAH RUANGAN PENYINARAN DAN HASIL PENGUKURAN PAPARAN RADIASI.........15
3.1 Denah Ruangan Penyinaran Pesawat Sinar-X..............................................................15
3.2 Hasil Pengukuran Paparan Radiasi...............................................................................18
3.3 Analisa Hasil Pengukuran Paparan Radiasi..................................................................18
BAB IV PAPARAN POTENSIAL DAN HASIL PENGUKURAN KEBOCORAN TABUNG....................19
4.1 Paparan Potensial..........................................................................................................19
4.1.1 Identifikasi Masalah.........................................................................................19
4.1.2 Resiko..............................................................................................................19
4.1.3 Mitigasi............................................................................................................20
4.1.4 Tindaklanjut.....................................................................................................20
4.2 Hasil pengukuran Kebocoran Tabung............................................................................21
BAB V KENDALI MUTU PESAWAT SINAR-X.....................................................................................22
5.1 Uji Fungsi Pesawat Sinar-X...........................................................................................22
5.2 Uji Kesesuaian Pesawat Sinar-X...................................................................................22
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 3 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

BAB I
PENDAHULUAN

1.1 Latar Belakang


Setiap pengguna Pesawat Sinar-X radiologi diagnostik dan/atau
intervensional sebagaimana diatur dalam Peraturan Bapeten Nomor 03 Tahun
2021 tentang Standar Kegiatan Usaha dan Standar Produk pada
Penyelenggaraan Perizinan Berusaha Berbasis Resiko Sektor Ketenaganukliran
wajib memenuhi persyaratan administratif dengan menyusun Dokumen Laporan
Kajian Keselamatan Radiasi. Dokumen Laporan Kajian Keselamatan Radiasi
tersebut antara lain memuat :

1. Pengukuran pemantauan paparan radiasi di daerah kerja;

2. Identifikasi terjadinya paparan potensial; dan

3. Kendali mutu Pesawat Sinar-X.


PT. Global Medik Persada telah menyusun Laporan Kajian Keselamatan
Radiasi untuk memastikan bahwa penggunaan Pesawat Sinar-X di PT. Global
Medik Persada dalam kondisi selamat bagi Pekerja, Masyarakat, dan Lingkungan.
Ruangan penyinaran sudah dirancang sedemikian rupa sehingga paparan radiasi
di ruang operator dan di ruangan sekitar tidak melebihi ketentuan keselamatan
sesuai dengan NBD dan Pembatas Dosis yang berlaku. Wadah tabung sinar-X
dalam kondisi baik dengan memastikan nilai kebocoran tabung tidak melebihi
ketentuan. Seluruh sistem Pesawat Sinar-X telah dilakukan uji fungsi oleh
instalatir dan dilakukan uji kesesuaian secara berkala oleh lembaga uji yang
berkompeten.

1.2 Tujuan
Tujuan pembuatan Dokumen Laporan Kajian Keselamatan Radiasi ini adalah:
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 4 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

1. Memberikan gambaran yang jelas mengenai Pesawat Sinar-X beserta prosedur


penggunaannya.

2. Menjelaskan kondisi ruangan penyinaran dan hasil pengukuran paparan radiasi


di sekitar ruangan penyinaran

3. Memastikan bahwa tidak terjadi paparan potensial akibat adanya kebocoran


wadah tabung.

4. Memastikan bahwa kondisi Pesawat Sinar-X selalu dalam kondisi baik dan
aman untuk digunakan dalam menegakkan diagnosa terhadap pasien.

5. Memastikan bahwa semua personil telah bekerja dengan memenuhi semua


Prinsip Keselamatan Radiasi.
1.3 Ruang Lingkup
Laporan Kajian Keselamatan Radaisi ini berisi pembahasan tentang:

1. Pendahuluan

2. Deskripsi Pesawat Sinar-X dan Prosedur Penggunaannya

3. Denah Ruangan Penyinaran dan Hasil Pengukuran Paparan Radiasi

4. Paparan Potensial dan Hasil Pengukuran Kebocoran Tabung

5. Uji Fungsi dan Uji Kesesuaian Pesawat Sinar-X


1.4 Definisi
Dalam Laporan Kajian Keselamatan Radiasi dalam penggunaan Pesawat
Sinar-X Radiologi Diagnostik dan/atau Intervensional di PT. Global Medik
Persada ini yang dimaksud dengan:

1. Keselamatan Radiasi Pengion yang selanjutnya disebut Keselamatan Radiasi


adalah tindakan yang dilakukan untuk melindungi pekerja, anggota
masyarakat, dan lingkungan hidup dari bahaya radiasi.

2. Pesawat Sinar-X adalah sumber radiasi yang terdiri dari generator tegangan
tinggi, panel kendali, tabung sinar-X, Kolimator, dan peralatan pendukung
lainnya.
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 5 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

3. Radiologi Diagnostik adalah teknik Radiologi untuk mendiagnosis suatu


penyakit atau kelainan morfologi dalam tubuh pasien dengan menggunakan
pesawat sinar-X.

4. Radiologi Intervensional adalah teknik Radiologi dengan menggunakan


pesawat sinar-X untuk pemandu citra secara langsung (real-time imageguided)
dalam mendiagnosis dan melakukan tindakan terapi dengan memasang kawat
penuntun, stent, dan komponen terkait di dalam tubuh pasien.

5. Uji Kesesuaian Pesawat Sinar-X adalah uji untuk memastikan Pesawat Sinar-X
dalam kondisi andal.

6. Kondisi Penyinaran adalah kombinasi pengaturan tegangan, kuat arus, dan


waktu penyinaran, atau kombinasi pengaturan tegangan dan perkalian kuat
arus dengan waktu yang digunakan untuk melakukan penyinaran radiasi. citra
diagnostik real time.

7. Kebocoran Radiasi adalah radiasi yang keluar dari tabung Pesawat Sinar-X
selain berkas utama.

8. Uji Fungsi atau Uji Keberterimaan adalah serangkaian kegiatan yang dilakukan
terhadap Pesawat Sinar-X untuk memastikan bahwa bagian-bagian Pesawat
Sinar-X baik elektrik maupun mekanik bekerja dengan baik.

9. Pengukuran Paparan Radiasi adalah serangkaian kegiatan yang dilakukan


untuk mengukur paparan radiasi di sekitar ruangan penyinaran pada saat
Pesawat Sinar-X dioperasikan.

10. Nilai Batas Dosis adalah dosis terbesar yang diizinkan oleh Kepala Badan
yang dapat diterima oleh Pekerja Radiasi dan anggota masyarakat dalam
jangka waktu tertentu tanpa menimbulkan efek genetik dan somatik yang
berarti akibat pemanfaatan tenaga nuklir.

11. Pembatas Dosis adalah batas atas dosis Pekerja Radiasi dan anggota
masyarakat yang tidak boleh melampaui Nilai Batas Dosis yang digunakan
pada optimisasi proteksi dan Keselamatan radiasi untuk setiap pemanfaatan
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 6 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

tenaga nuklir.

1.5 Referensi
Prosedur Operasi Instalasi Pesawat Sinar-X Radiografi Mobile yang berlaku
di PT. Global Medik Persada ini disusun berdasarkan ketentuan-ketentuan yang
tercantum di dalam:
1. Peraturan Kepala Bapeten Nomor 04 Tahun 2013 tentang Proteksi Radiasi
dalam Pemanfaatan Tenaga Nuklir.

2. Peraturan Kepala Bapeten Nomor 02 Tahun 2018 tentang Uji Kesesuaian


Pesawat Sinar-X Radiologi Diagnostik dan Intervensional.

3. Peraturan Bapeten Nomor 04 Tahun 2020 tentang Keselamatan Radiasi dalam


Penggunaan Pesawat Sinar-X Radiologi Diagnostik dan Intervensional.

4. Peraturan Bapeten Nomor 03 Tahun 2021 tentang Standar Kegiatan Usaha


dan Standar Produk pada Penyelenggaraan Perizinan Berusaha Berbasis
Resiko Sektor Ketenaganukliran.

5. Peraturan Bapeten Nomor 01 Tahun 2022 tentang Penatalaksanaan


Perizinan Berusaha Berbasis Resiko Sektor Ketenaganukliran
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 7 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

BAB II

DESKRIPSI PESAWAT SINAR-X DAN PROSEDUR PENGGUNAANNYA

2.1 Data Pesawat Sinar-X


Pesawat Sinar-X yang dimiliki oleh PT. Global Medik Persada Raya adalah
Pesawat Sinar-X Radiografi Mobile dengan spesifikasi sebagai berikut:

a. Brand Unit : Merk GE


Model Unit : Mobile X-Ray
Tipe Unit : Optima XR200amx
Brand Tabung : Canon Electron Tubes & Devices CO.. LTD.
Model Tabung : E7894X
Nomor seri Tabung : -
Tegangan Maksimum : 150 kV
Kuat Arus Maksimum : 300 mA

b. Brand Unit : Merk GE


Model Unit : Mobile X-Ray
Tipe Unit : Optima XR220amx
Brand Tabung : Canon Electron Tubes & Devices CO.. LTD.
Model Tabung : E7894X
Nomor seri Tabung : -
Tegangan Maksimum : 150 kV
Kuat Arus Maksimum : 300mA

c. Brand Unit : Merk GE


Model Unit : Mobile X-Ray
Tipe Unit : Optima XR240amx
Brand Tabung : Canon Electron Tubes & Devices CO.. LTD
Model Tabung : E7894X
Nomor seri Tabung : -
Tegangan Maksimum : 150 kV
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 8 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Kuat Arus Maksimum : 630 mA

d. Brand Unit : Merk GE


Model Unit : Mobile X-Ray
Tipe Unit : Traveller
Brand Tabung : IMD Generators s.r.l
Model Tabung : MHF2030
Nomor seri Tabung : -
Tegangan Maksimum : 125 kV
Kuat Arus Maksimum : 450 mA

e. Brand Unit : Merk GE


Model Unit : Stationary X-Ray
Tipe Unit : XR6000
Brand Tabung : GEBE Private Limited
Model Tabung : 533186
Nomor seri Tabung : -
Tegangan Maksimum : 150 kV
Kuat Arus Maksimum : 630 mA

f. Brand Unit : Merk GE


Model Unit : Stationary X-Ray
Tipe Unit : Brivo XR575,
Brand Tabung : GEBE Private Limited
Model Tabung : 533186
Nomor seri Tabung : -
Tegangan Maksimum : 150 kV
Kuat Arus Maksimum : 630 mA

2.2 Prosedur Pengoperasian Pesawat Sinar-X


Prosedur pengoperasian Pesawat Sinar-X Radiografi Mobile dan Stationer merk GE
untuk tipe Optima XR200amx, Optima XR220amx, XR240amx, XR6000, Traveller dan
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 9 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

XR575 terlampir.

2.3 Tabel Jenis Penyinaran dan Kondisi Penyinaran


Tabel Jenis Penyinaran dan Kondisi Penyinaran yang dapat diberikan
kepada pasien yang diperiksa menggunakan Pesawat Sinar-X Radiografi
Stationer merk GE tipe
XR575 adalah sebagai berikut:
TABEL JENIS PENYINARAN DAN KONDISI PENYINARAN PESAWAT SINAR-X

Abdomen kV mAs Focal Spot Grid


Small Adult 70 8 Large In
Medium Adult 75 12.5 Large In
Large Adult 85 25 Large In
Small Ped 60 1.6 Small Out
Medium Ped 65 3.2 Small Out
Large Ped 65 5 Small Out

Chest kV mAs Focal Spot Grid


Small Adult 100 2 Large In
Medium Adult 115 2.5 Large In
Large Adult 120 4 Large In
Small Ped 65 0.4 Small Out
Medium Ped 75 1 Small Out
Large Ped 85 1.6 Small Out

Chest/Ribs kV mAs Focal Spot Grid


Small Adult 70 5 Small In
Medium Adult 75 8 Small In
Large Adult 80 10 Small In
Small Ped 55 1.6 Small Out
Medium Ped 60 2.5 Small Out
Large Ped 65 3.2 Small Out

Shoulder kV mAs Focal Spot Grid


Small Adult 65 4 Small In
Medium Adult 70 5 Small In
Large Adult 80 8 Small In
Small Ped 55 0.63 Small Out
Medium Ped 60 1 Small Out
Large Ped 65 2.5 Small Out
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 10 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Clavicula kV mAs Focal Spot Grid


Small Adult 70 5 Small In
Medium Adult 70 6.3 Small In
Large Adult 75 8 Small In
Small Ped 55 2 Small Out
Medium Ped 60 2.5 Small Out
Large Ped 65 3.2 Small Out

Scapula kV mAs Focal Spot Grid


Small Adult 75 8 Small In
Medium Adult 75 10 Small In
Large Adult 85 12.5 Small In
Small Ped 55 2 Small Out
Medium Ped 60 3.2 Small Out
Large Ped 65 4 Small Out

Humerus kV mAs Focal Spot Grid


Small Adult 65 4 Small In
Medium Adult 70 5 Small In
Large Adult 80 8 Small In
Small Ped 55 2 Small Out
Medium Ped 60 2.5 Small Out
Large Ped 65 3.2 Small Out

Elbow kV mAs Focal Spot Grid


Small Adult 60 2 Small In
Medium Adult 60 3.2 Small In
Large Adult 60 4 Small In
Small Ped 50 1 Small Out
Medium Ped 50 1.6 Small Out
Large Ped 55 2 Small Out

Forearm kV mAs Focal Spot Grid


Small Adult 60 2.5 Small In
Medium Adult 65 2.5 Small In
Large Adult 65 4 Small In
Small Ped 50 1.6 Small Out
Medium Ped 50 2 Small Out
Large Ped 55 2 Small Out

Wrist kV mAs Focal Spot Grid


Small Adult 55 2 Small In
Medium Adult 55 2.5 Small In
Large Adult 60 2.5 Small In
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 11 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Small Ped 50 1 Small Out


Medium Ped 50 1.25 Small Out
Large Ped 55 1.6 Small Out

Manus kV mAs Focal Spot Grid


Small Adult 50 1.5 Small In
Medium Adult 55 1.5 Small In
Large Adult 55 2 Small In
Small Ped 50 1 Small Out
Medium Ped 50 1.25 Small Out
Large Ped 55 1.6 Small Out

Pelvis kV mA Focal Spot Grid


s
Small Adult 70 5 Small In
Medium Adult 85 12.5 Small In
Large Adult 85 20 Small In
Small Ped 60 1.25 Small Out
Medium Ped 65 3.2 Small Out
Large Ped 70 4 Small Out

Hip kV mA Focal Spot Grid


s
Small Adult 70 5 Small In
Medium Adult 85 12.5 Small In
Large Adult 85 20 Small In
Small Ped 60 1.25 Small Out
Medium Ped 65 3.2 Small Out
Large Ped 70 4 Small Out

Femur kV mAs Focal Spot Grid


Small Adult 70 8 Small In
Medium Adult 75 12.5 Small In
Large Adult 75 16 Small In
Small Ped 55 2.5 Small Out
Medium Ped 65 3.2 Small Out
Large Ped 65 4 Small Out

Knee kV mAs Focal Spot Grid


Small Adult 65 3.2 Small In
Medium Adult 70 4 Small In
Large Adult 75 6.3 Small In
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 12 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Small Ped 50 1.6 Small Out


Medium Ped 60 2 Small Out
Large Ped 60 2.5 Small Out

Cruris kV mAs Focal Spot Grid


Small Adult 60 4 Small In
Medium Adult 70 5 Small In
Large Adult 75 6.3 Small In
Small Ped 50 1.6 Small Out
Medium Ped 55 2 Small Out
Large Ped 60 2.5 Small Out

Ankle kV mAs Focal Spot Grid


Small Adult 60 4 Small In
Medium Adult 65 5 Small In
Large Adult 65 6.3 Small In
Small Ped 50 1 Small Out
Medium Ped 55 1 Small Out
Large Ped 60 1.6 Small Out

Pedis kV mAs Focal Spot Grid


Small Adult 60 1.6 Small In
Medium Adult 60 2 Small In
Large Adult 65 2.5 Small In
Small Ped 50 1.6 Small Out
Medium Ped 55 2 Small Out
Large Ped 60 2 Small Out

C Spine AP kV mAs Focal Spot Grid


Small Adult 65 6.3 Small In
Medium Adult 75 8 Small In
Large Adult 75 10 Small In
Small Ped 50 0.63 Small Out
Medium Ped 50 1.25 Small Out
Large Ped 60 2.5 Small Out

C Spine Lat kV mAs Focal Spot Grid


Small Adult 65 6.3 Small In
Medium Adult 75 8 Small In
Large Adult 75 10 Small In
Small Ped 50 0.63 Small Out
Medium Ped 50 1.25 Small Out
Large Ped 60 2.5 Small Out

T Spine AP kV mAs Focal Spot Grid


Small Adult 75 8 Small In
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 13 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Medium Adult 80 10 Small In


Large Adult 85 20 Small In
Small Ped 55 2 Small Out
Medium Ped 65 3.2 Small Out
Large Ped 70 5 Small Out

T Spine Lat kV mAs Focal Spot Grid


Small Adult 75 16 Small In
Medium Adult 80 20 Small In
Large Adult 85 32 Small In
Small Ped 60 2 Small Out
Medium Ped 65 6.3 Small Out
Large Ped 70 10 Small Out

L Spine AP kV mAs Focal Spot Grid


Small Adult 70 12.5 Small In
Medium Adult 85 16 Small In
Large Adult 85 20 Small In
Small Ped 60 2 Small Out
Medium Ped 65 5 Small Out
Large Ped 70 8 Small Out

L Spine Lat kV mAs Focal Spot Grid


Small Adult 80 20 Small In
Medium Adult 90 32 Small In
Large Adult 95 40 Small In
Small Ped 60 4 Small Out
Medium Ped 65 10 Small Out
Large Ped 70 12.5 Small Out

Skull/Cranium kV mAs Focal Spot Grid


Small Adult 60 6.3 Small In
Medium Adult 65 10 Small In
Large Adult 85 16 Small In
Small Ped 50 1.6 Small Out
Medium Ped 55 2 Small Out
Large Ped 60 3.2 Small Out
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 14 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Skull/Cranium kV mAs Focal Spot Grid


Lat
Small Adult 60 5 Small In
Medium Adult 65 6.3 Small In
Large Adult 75 10 Small In
Small Ped 55 1.6 Small Out
Medium Ped 55 2.5 Small Out
Large Ped 60 3.2 Small Out
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 15 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

BAB III

DENAH RUANGAN PENYINARAN DAN HASIL PENGUKURAN PAPARAN RADIASI

3.1 Denah Ruangan Penyinaran Pesawat Sinar-X

Merujuk pada ketentuan-ketentuan Peraturan Bapeten Nomor 04 Tahun


2020, dan Peraturan Menteri Kesehatan Nomor 24 Tahun 2020, maka PT. Global
Medik Persada telah menyediakan Ruangan Penggunaan/Pengoperasian
Pesawat Sinar-X Radiografi Mobile dengan spesifikasi sebagai berikut:

A. Tampak Atas
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 16 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Data ruang radiologi I Keterangan

Nama ruangan :

Ukuran ruang :

Nomor izin pemanfaatan :


Data pesawat
Merk pesawat sinar-X : GE
Tipe/model pesawat sinar-X : Mobile X-Ray/Optima XR200
No. Seri pesawat sinar-X :-
Tahun pembuatan :-
Tahun pemasangan :-

Data tabung
Merk tabung : Canon Electron Tubes & Devices CO.. LTD
Tipe tabung : E7894X
No. Seri tabung :-
Beda tegangan maksimum (kV) : 150 kV
Arus (mA) maksimum : 300 mA
Arus waktu (mAs) maksimum : 630 mAs

Data ruang radiologi I Keterangan

Nama ruangan :

Ukuran ruang :

Nomor izin pemanfaatan :


Data pesawat
Merk pesawat sinar-X : GE
Tipe/model pesawat sinar-X : Mobile X-Ray/Optima XR220
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 17 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

No. Seri pesawat sinar-X :-


Tahun pembuatan :-
Tahun pemasangan :-

Data tabung
Merk tabung : Canon Electron Tubes & Devices CO.. LTD
Tipe tabung : E7894X
No. Seri tabung :-
Beda tegangan maksimum kV) :150 kV
Arus (mA) maksimum : 300 mA
Arus waktu (mAs) maksimum : 630 mAs

Data ruang radiologi I Keterangan

Nama ruangan :

Ukuran ruang :

Nomor izin pemanfaatan :


Data pesawat
Merk pesawat sinar-X : GE
Tipe/model pesawat sinar-X : Stationary X-Ray/Brivo XR575
No. Seri pesawat sinar-X :-
Tahun pembuatan :-
Tahun pemasangan :-

Data tabung
Merk tabung : GEBE
Tipe tabung : 5331186
No. Seri tabung :-
Beda tegangan maksimum kV) : 150 kV
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 18 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

Arus (mA) maksimum : 630 mA


Arus waktu (mAs) maksimum : 625 mAs

3.2 Hasil Pengukuran Paparan Radiasi

Untuk memastikan bahwa ruangan penyinaran Pesawat Sinar-X telah


memenuhi prinsip-prinsip keselamatan radiasi, maka dilakukan pengukuran
paparan radiasi di sekitar ruangan penyinaran oleh instalatir dengan hasil sebagai
berikut:

Pengukuran
Tebal Jenis
Lokasi disekitar ruang radiologi + Pb paparan
dinding material
(µSv/h)

Kanan : Bata

Kiri : Bata

Atas : Lantai Cor

Bawah : Lantai Keramik

Belakang : Bata

Depan : Bata

Di depan Pintu Kayu berlapis Pb

Ruang Operator Kaca Intip Pb

3.3 Analisa Hasil Pengukuran Paparan Radiasi


Berdasarkan ketentuan Peraturan Kepala Bapeten Nomor 04 Tahun 2013,
PT. Global Medik Persada menetapkan Nilai Pembatas Dosis (Dose Constraint),
yaitu:

1. Pembatas Dosis untuk Pekerja Radiasi sebesar 6 mSv/tahun

2. Pembatas Dosis untuk Masyarakat sebesar 0,3 mSv/tahun


PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 19 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

BAB IV

PAPARAN POTENSIAL DAN HASIL PENGUKURAN KEBOCORAN TABUNG

4.1 Paparan Potensial


Paparan Potensial adalah paparan yang tidak diharapkan atau diperkirakan
tetapi mempunyai kemungkinan terjadi akibat kecelakaan Sumber atau karena
suatu kejadian atau rangkaian kejadian yang mungkin terjadi termasuk kegagalan
peralatan atau kesalahan operasional. Berikut ini adalah identifikasi masalah yang
dapat menimbulkan paparan potensial di PT. Global Medik Persada.
4.1.1 Identifikasi Masalah
Masalah yang mungkin dapat memberikan kontribusi terhadap paparan
potensial di PT. Global Medik Persada. adalah:

1. Ketidaksesuaian lapangan berkas kolimasi

2. Kebocoran wadah tabung

3. Kebocoran radiasi pada tabir radiasi Pb mobile

4. Kebocoran radiasi pada Apron, pelindung gonad, pelindung tiroid, dsb.

5. Penggunaan faktor penyinaran yang tidak sesuai

4.1.2 Resiko
Resiko yang mungkin muncul akibat adanya masalah yang dapat
memberikan kontribusi terhadap paparan potensial di PT. Global Medik
Persada adalah:

1. Penerimaan dosis berlebih terhadap pasien

2. Penerimaan dosis berlebih terhadap pasien

3. Penerimaan dosis berlebih terhadap Pekerja dan Masyarakat

4. Penerimaan dosis berlebih terhadap Pekerja

5. Penerimaan dosis berlebih terhadap pasien


PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 20 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

4.1.3 Mitigasi
Mitigasi terhadap resiko yang mungkin muncul akibat adanya masalah yang
dapat memberikan kontribusi terhadap paparan potensial di PT. Global
Medik Persada adalah:

1. Melakukan pemeriksaan harian (daily check) setiap pagi pada saat


pemanasan (warmup) Pesawat Sinar-X.

2. Melakukan Uji Kebocoran wadah tabung pada saat awal instalasi/atau


pindah ruangan

3. Melakukan Uji Kebocoran tabir radiasi Pb mobile minimal 1 tahun


sekali.

4. Melakukan Uji Kebocoran Apron, pelindung gonad, pelindung tiroid,


dsb. minimal 1 tahun sekali.

5. Membuat prosedur penyinaran dan menempatkannya di ruang


operator.

4.1.4 Tindaklanjut
Tindaklanjut terhadap rencana mitigasi untuk resiko yang mungkin muncul
akibat adanya masalah yang dapat memberikan kontribusi terhadap
paparan potensial di PT. Global Medik Persada adalah:

1. Melakukan Uji Kolimasi minimal 2 bulan sekali

2. Melakukan Uji Kebocoran wadah tabung pada saat awal instalasi/atau


pindah ruangan

3. Memperbaiki tabir radiasi Pb mobile yang mengalami kebocoran.

4. Mengadakan Apron, pelindung gonad, pelindung tiroid, yang baru


dan/atau menggudangkan Apron, pelindung gonad, pelindung tiroid,
yang tidak layak.
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 21 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

5. Memastikan setiap penyinaran dilakukan sesuai dengan Prosedur


Penyinaran (SOP) masing-masing Pesawat Sinar-X.

4.2 Hasil pengukuran Kebocoran Tabung


Kebocoran wadah tabung insersi Pesawat Sinar-X merupakan parameter
yang secara langsung dapat mempengaruhi dosis radiasi pasien dan menentukan
kelayakan operasi Pesawat Sinar-X. Untuk itu kebocoran wadah tabung
merupakan parameter penting yang dilakukan pengujian sebelum Pesawat Sinar-
X digunakan untuk pelayanan penyinaran/pemeriksaan terhadap pasien.

Kriteria keberterimaan atau batas toleransi kebocoran tabung harus lebih kecil dari 1
mGy/jam pada jarak 100 cm.

Hasil Pengukuran Kebocoran Tabung

Kondisi Pengukuran : (………… kVp) ; (………… mA) ; (………… mAs) / ( s)

Posisi Hasil Pengukuran


Titik Ukur (μSv/jam) Nilai Tertinggi

A. Kanan

B. Kiri Hasil pengukuran kebocoran


tabung tertinggi adalah di
C. Depan posisi ……………
D. Belakang
yaitu.......................(μSv/jam)
E. Atas atau.......................(mSv/jam)

F. Bawah
PROGRAM PROTEKSI DAN KESELAMATAN No. Dok :
RADIASI UNTUK TUJUAN PENGALIHAN 002/PPR/GMP/2022
PEMBANGKIT RADIASI PENGION BIDANG MEDIK Tanggal : 18/Aug/2022
“PT. GLOBAL MEDIK PERSADA”
Jalan Raya Kebon Jeruk No.26 Revisi : 00
Rt.09/RW.02, Kebon Jeruk, Jakarta Barat
Telp/Fax :(021) 50899710/29319448, Halaman 22 dari 24
Email :info@GLOMEDA.co.id
www.GLOMEDA.co.id

BAB V

KENDALI MUTU PESAWAT SINAR-X

5.1 Uji Fungsi Pesawat Sinar-X


Uji Fungsi atau Uji Keberterimaan adalah serangkaian kegiatan yang
dilakukan terhadap Pesawat Sinar-X untuk memastikan bahwa bagian-bagian
Pesawat Sinar-X baik elektrik maupun mekanik bekerja dengan baik. Uji Fungsi
atau Uji Keberterimaan dilakukan oleh Vendor atau Instalatir yang melakukan
pemasangan Pesawat Sinar-X. Jika Uji Fungsi atau Uji Keberterimaan sudah
dilakukan oleh Vendor atau Instalatir, maka akan diterbitkan Berita Acara Hasil Uji
Fungsi yang ditandatangani oleh pihak Instalatir dan diketahui oleh manajemen
PT. Global Medik Persada.
TERLAMPIR

5.2 Uji Kesesuaian Pesawat Sinar-X


Uji Kesesuaian adalah serangkaian kegiatan pengujian untuk memastikan
pesawat sinar-X dalam kondisi andal. Untuk memastikan keandalan Pesawat
Sinar-X uji kesesuaian di lakukan oleh Lembaga Uji Kesesuaian yang telah
mendapatkan penunjukkan dari Kepala Bapeten. Berdasarkan hasil uji kesesuaian
selanjutnya diterbitkan Sertifikat Hasil Uji Kesesuaian oleh Tenaga Ahli.
PT. Global Medika Persada merupakan Lembaga Uji Kesesuaian yang dapat
melakukan Uji Kesesuaian untuk Pesawat Sinar-X lingkup Radiografi Umum
dengan Nomor KEPUTUSAN KEPALA BADAN PENGAWAS TENAGA NUKLIR :
018.01.02.1.210622