Anda di halaman 1dari 3

Diagram Fishbone

Kalibrasi pH Presisi metode

kalibrasi

temperatur
pH
Lar. pH 10
Lar. pH 4
Lar. pH 7

Perhitungan
I. Data Pendukung
𝜇 𝐾𝑎𝑙𝑖𝑏𝑟𝑎𝑠𝑖 𝑝𝐻 4 0.02
𝜇 𝐾𝑎𝑙𝑖𝑏𝑟𝑎𝑠𝑖 𝑝𝐻 7 0.02
𝜇 𝐾𝑎𝑙𝑖𝑏𝑟𝑎𝑠𝑖 𝑝𝐻 10 0.02
Efek temp lab oC 25
Pengukuran sampel 1 2 Rata Rata
7.37 7.37 7.37

II. Ketidakpastian pH
1. Ketidakpastian kalibrasi
a. 𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 4
Kuantitas Nilai Terukur 𝝁
pH 4 4.01 ± 0.02 0.01
0.02
= 0.01
2
𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 4 0.01
b. 𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 7
Kuantitas Nilai Terukur 𝝁
pH 7 7.00 ± 0.02 0.01
0.02
= 0.01
2
𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 7 0.01

c. 𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 10
Kuantitas Nilai Terukur 𝝁
pH 10 10.01 ± 0.02 0.01
0.02
= 0.01
2
𝜇 𝐵𝑢𝑓𝑓𝑒𝑟 𝑝𝐻 10 0.01

d. 𝜇 𝑉𝑎𝑟𝑖𝑎𝑠𝑖 𝑡𝑒𝑚𝑝𝑒𝑟𝑎𝑡𝑢𝑟 𝑙𝑎𝑏𝑜𝑟𝑎𝑡𝑜𝑟𝑖𝑢𝑚


Variasi (oC) 𝝁
± 2 0.0002

±2o C x 2.1 x 10−4 ml


= 0.0002
√3
𝜇 𝑉𝑎𝑟𝑖𝑎𝑠𝑖 𝑡𝑒𝑚𝑝𝑒𝑟𝑎𝑡𝑢𝑟 𝑙𝑎𝑏𝑜𝑟𝑎𝑡𝑜𝑟𝑖𝑢𝑚 0.0002

III. Presisi Metode


No pH
1 7.37
2 7.36
3 7.36
4 7.37
5 7.36
6 737
7 7.37
rata2 7.3671429
SD 0.0048795
rsd 0.0662333

IV. Ketidakpastian Gabungan


No Data Nilai 𝝁 𝑩𝒂𝒌𝒖 𝝁 𝒓𝒆𝒍𝒂𝒕𝒊𝒇
1 Faktor kalibrasi buffer
pH 4 4.01 0.01 0.0025
pH 7 7.00 0.01 0.0014
pH 10 10.01 0.01 0.0009
Variasi Temperatur 0.0002 0.0002

2 Presisi Metode 0.00066 0.00066

μ Kalibrasi pH
= √(μ Kal pH 4)2 + (μ Kal pH 7)2 + (μ Kal pH 10)2 + (μ Efek Temp)2 + RSD2
pH Terukur

μ Kalibrasi pH
= √0.00252 + 0.00142 + 0.00092 + 0.00022 + 0.000662
7.37

μ Kalibrasi pH
= √0.0000094956
7.37

μ Kalibrasi pH = 0.0000067144 x 7.37

μ Kalibrasi pH = 0.0000494852
ketidakpastian diperluas dengan tingkat kepercayaan 95% = 0.00009

V. Pelaporan
pH = 7.37 ± 0.00009

Anda mungkin juga menyukai